Ugin PROGRAMIX 50 Instructions For Use Manual

Ugin PROGRAMIX 50 Instructions For Use Manual

Burnout furnace

Quick Links

INSTRUCTIONS FOR USE
Burnout furnace
PROGRAMIX 50 & PROGRAMIX 100
Rév. 031121
Table of Contents
loading

Summary of Contents for Ugin PROGRAMIX 50

  • Page 1 INSTRUCTIONS FOR USE Burnout furnace PROGRAMIX 50 & PROGRAMIX 100 Rév. 031121...
  • Page 2: Table Of Contents

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 PROGRAMIX 50 & PROGRAMIX 100 EQUIPMENT DESCRIPTION ........................3 ELECTRICAL DIAGRAM ..........................4 INSTRUCTIONS FOR USE .......................... 5 SAFETY INSTRUCTIONS ........................5 1.1. USE ............................5 1.2. SAFETY INSTRUCTIONS ......................5 1.3.
  • Page 3: Equipment Description

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 A. EQUIPMENT DESCRIPTION Rear panel N° DESCRIPTION Chimney outlet Thermal fuse Power socket Fumes extractor connection socket Location of triac Thermocouple cover UGIN DENTAIRE...
  • Page 4: Electrical Diagram

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 B. ELECTRICAL DIAGRAM UGIN DENTAIRE...
  • Page 5: Instructions For Use

    Programix 50 and Programix 100 have to use in the conditions and for the uses described in this manual. Any other use that will be made of it (such as firing other materials, heating food products, etc.) is considered inappropriate.
  • Page 6: Waste Disposal

    Content of the declaration EC: ▪ Name and address of the manufacturer: UGIN DENTAIRE. 25 rue de la Tuilerie. 38170 SEYSSINET-PARISET – FRANCE ▪ UGIN DENTAIRE declares, on its own exclusive responsibility, that the product: - Burnout furnace PROGRAMIX 50 or PROGRAMIX 100 - Serial number: see the equipment’s EC declaration of conformity...
  • Page 7: Technical Information

    3.1. TECHNICAL DATA Programix 50 and Programix 100 burnout furnaces are intended use to heating the coated cylinders. These furnaces have to be operated only by dental professionals and in the conditions described in this manual (see chapter 1.1 and 3.2).
  • Page 8: Conditions Of Use, Transport And Storage

    4. INSTALLATION AND START-UP 4.1. UNPACK - Unpack PROGRAMIX 50 / 100 and check that it is in perfect order (furnace and accessories). - Any defects can be reported to the carrier. - Remember to get the carrier to sign the delivery note.
  • Page 9: Installation

    UGIN DENTAIRE recommends installing a fumes extractor on the PROGRAMIX 50 / 100 in order to achieve optimal gas evacuation. It will be installed in place of the chimney outlet. It is recommended to use the UGIN DENTAIRE fumes extractor.
  • Page 10: Programming

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 5. PROGRAMMING 5.1. Description of the control panel N° DESCRIPTION Program progress table Program number display (0 to 29) Temperature / time display Indicator of the unit displayed on 3 (temperature / time)
  • Page 11: Operating Parameters

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 5.2. OPERATING PARAMETERS - To program the operating parameters, press key - When the program number blinking (display No.2), enter the data with keys - To switch to another parameter, use the key.
  • Page 12: Program Number Display

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 ▪ 3 STAGE Rate No. 3 (RP 3): from 1 to 15°C/min – FULL - STOP Temperature rise to the 3 stage. stage temperature: from 0 to 1200°C Temperature setting for the 3...
  • Page 13: Start/Stop Touch Control

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 In programming position: press the key increases the value of the parameter displayed on the screen (No. 3), press the key decreases that value. 5.10. START/STOP TOUCH CONTROL After checking the number of the program selected, press the key START/STOP control to start the heating cycle.
  • Page 14: Use

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 It is programmed after the final time-delay parameter has been set by pressing the CYCLE key and FORWARD/REVERSE keys. The fumes extractor starts operating at the beginning of the heating cycle and stops when the programmed temperature is reached.
  • Page 15: Calibration

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 7. CALIBRATION The furnace has been subjected to a temperature test using a temperature calibrator during his factory. However, you can change the set value the initial setting satisfactory. You can test the temperature using a pure silver wire.
  • Page 16: Protection

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 8. PROTECTION ▪ PROGRAM PROTECTION The programmer is equipped with a lithium battery which allows programs to be stored in memory. It is recommended to change the battery approximately every 5 years.
  • Page 17: Maintenance And Trouble-Shooting

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 9. MAINTENANCE AND TROUBLE-SHOOTING Before to carry out the maintenance work and/or cleaning: - disconnect from the main power supply (switch OFF and unplug to mains socket), - wait the cooling at room temperature after use (risk of burns and/or damage).
  • Page 18: Trouble-Shooting

    - Remove the back panel and check resistance at element) is cut the muffle with a multimeter: - Programix 50: 24 ohms - Programix 100: 15 ohms - Replace the muffle as indicated in chapter 9.4. 2. The furnace does...
  • Page 19: Thermocouple Replacement

    - Let the furnace heat up at a steady 200°C to check if is working properly. It is then ready for use again. 9.4. MUFFLE REPLACEMENT IMPORTANT – Personal Protective Equipment PROGRAMIX 50 / 100 contains insulating material composed of high temperature insulation wool (alkaline earth silicate), a substance not classified as dangerous according to Regulation 1272/2008/EC (CLP).
  • Page 20: Table Of Programs

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 10. TABLE OF PROGRAMS 10.1. TABLE OF SINGLE PROGRAMS (No. 0 to No.19) Start. Temp. Hold Temp. Hold Temp. Hold Final Description temp time time time temp. UGIN DENTAIRE...
  • Page 21: Table Of Double Programs (No. 20 To No. 29)

    Instructions for use (rev 031121) - En PROGRAMIX 50 - 100 10.2. TABLE OF DOUBLE PROGRAMS (No. 20 to No. 29) Start. Hold Hold Hold Final Temp. Temp. Temp. Description temp time time time temp. UGIN DENTAIRE...

This manual is also suitable for:

Programix 100

Table of Contents