Removal procedure (cont'd)
5.) Remove the battery.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-37
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-9-4
Section 4-3-4
Section 4-3-5
Section 4-5
Section 10-5-2
Section 10-5-2-2
Section 10-5-5
8-2-36-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8-90
5
Figure 8-124 Pull out the battery from SYSCON PWA
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
SYSCONML Assy Function Validation
System B/M-Mode Checks
System CFM and PWD Checks
Peripheral Checks
Functional Checks (See Also Chapter 4)
Peripheral/Option Checks
Physical Inspection
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Procedure
5394141, Rev 1+, Section 8-2-36. Equipment
passed all required tests and is ready for use. "
LOGIQ™ P5 S
M
EVICE
ANUAL
Debrief Script
"Service Manual, Direction