EVISION
8-2-43

Air Filter Set

8-2-43-1
Tools
No need tool
8-2-43-2
Preparations
Shut down the system and switch off the main breaker.
8-2-43-3
Removal procedure
1.) Pull out the 2 Air filters from the bottom left and right side of the system (1-2).
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-44
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-8-3
Section 10-5-5
8-2-43-4
Mounting Procedure
Install the new parts in the reverse order of removal.
5
(1)
Figure 8-133 Air filters
Functional Tests
Functional Test / Diagnostic Test
Air filter set Function Validation Procedure
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Physical Inspection
LOGIQ™ P5 S
EVICE
(2)
Debrief Script
"Service Manual, Direction
5394141, Rev 1+, Section 8-2-43. Equipment
passed all required tests and is ready for use. "