GE AxiChrom 50 Operating Instructions Manual page 37

Columns
Table of Contents
Removing a column from the
AxiChrom pivot stand
The following procedure is used to remove the column from the pivot stand.
1 Make sure that the pivot stand is placed on an even foundation.
2 Make sure that the pivot stand arm is locked with the pin at the back of the vertical
3 Unscrew and remove the screws [701-2] at the top bayonet [804] and bottom flange
4 Lift the column vertically to remove the column from the pins at the upper and lower
Fitting the column onto the
AxiChrom pivot stand
Assemble the column onto the pivot stand as follows:
1 Make sure that the pivot stand is placed on an even foundation.
2 Make sure the stand arm is locked with the pin at the back of the vertical stand bar.
3 Lift the column vertically and begin by guiding the two holes at the bottom flange
4 Secure the column top at the stand arm with the screws [701-2].
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
stand bar.
CAUTION
The pivot stand locking pin is secured to prevent the column
from turning over by accident.
[805].
end of the stand arm.
CAUTION
The pivot stand locking pin is secured to prevent the column
from turning over by accident.
onto the pins at the lower end of the stand arm. Then guide the two holes at the top
bayonet onto the pins at the upper end.
4 Getting started
4.6 Routine procedures
37