Instrument Care And Maintenance; Cleaning Instructions; Backlights; Operating Temperature - Honeywell X Series User Manual

Hide thumbs Also See for X Series:
Table of Contents

Section 11: Instrument Care and Maintenance

Instrument Care and Manintenance
The X Series range of recorders contains no parts serviceable by the user, except for a
fuse on the DC power supply, see
In the event of a recorder failure contact your nearest Service Department (or an author-
ised agent) to arrange for the return of the unit for repair.

Cleaning Instructions

The recorder's touch screen can be temporarily disabled for cleaning, see
on page 141.
Cleaning the unit should be done with a soft lint cloth and warm soapy water or screen
cleaner recommended for use on a PC. Solvents and prolonged exposure to detergents
can cause damage to the front panel. It is recommended that any cloth used for cleaning
is damp but NOT wet, to avoid water collecting in the unit.
For Nema 4X /IP66 rated recorders (option) refer to the Nema 4X standards.

Backlights

1.
2.
3.
4.

Operating Temperature

Operating temperatures are described in
operation at temperatures over 50 °C will accelerate degradation of the display backlight.
If the unit has been moved from a cold environment into a warm one, ensure that the unit
has reached a minimum temperature of 12 °C or is left to stand for 1 hour at room temper-
ature before applying power, to ensure no condensation remains in the unit.
43-TV-25-30 Iss.6 GLO Aug 07 UK
MTTF (Mean Time To Fail) is defined as the time at which 50% of a batch of backlights
remain in excess of half their original brightness i.e. a display has a greater than 50%
chance of being half its original brightness when used at 100% after:
• 55,000 hrs for the Minitrend QX. Maximum luminosity 400 cd/m2, (86,000hr if used at
80%).
• 43,000 hrs for the Multitrend SX. Maximum luminosity 400 cd/m2, (67,000hr if used at
80%).
• 40,000 hrs for the eZtrend QXe. Maximum luminosity 450 cd/m2, (62,500hr if used at
For this recorder the backlight cannot be changed separately so the display
80%).
has to be replaced.
Reducing the brightness can significantly improve the life of the backlight.
At extended temperatures these times are reduced.
The backlight is a consumable item, and will NOT last forever.
A maintenance schedule of exchanging backlights should be every 5 years to maintain
the optimal disk clarity.
"Fuses" on page
13.
"Specification Tables" on page
"Clean Screen"
259. Prolonged
253
Table of Contents
loading

This manual is also suitable for:

Multitrend sxMinitrend qxEztrend qxe

Table of Contents