Displaying System Information - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents
Warnings and errors issued by the Analyzer collect in the Warnings/Errors list after their initial
display. The W icon displays in the Menu screen header when a warning has been issued and the E
icon displays when an error has been issued; if both a warning and an error has been issued, then
the E icon will display, as errors are more serious alerts than warnings. (See Chapter 8,
"Troubleshooting" for additional information on warnings and errors.)
To display the Warnings/Errors list, follow these steps:
1.
Select the Maintenance tab.
2.
Press the Warnings/Errors button.
3.
Use the scroll buttons to browse through the warnings and errors. For each warning/error,
you have three options:
• Remove — Remove the warning/error from the list.
• Export — Send the warning/error to the serial (RS-232) or USB port.
• Print — Send the warning/error to the printer.
The Analyzer displays the date, error number, error description, a brief suggestion for corrective
action, time of last occurrence, and the number of times this warning or error has occurred. A
complete listing of warnings and errors is given in "Warnings and Error Messages" on page 196.

Displaying System Information

If you need to contact GE Analytical Instruments about your Analyzer, you can find most essential
information on a single screen.
1.
Select the Maintenance tab.
2.
Press the System Info button. The following unit information is displayed:
• Location name
• Firmware version number
• Analyzer serial number
• GE Analytical Instruments phone numbers and Web site information
GE Analytical Instruments ©2009
114 of 236
DLM 90488-01 Rev. A