Procedures Setup; Setting Vital Capacity Defaults; Setting Cycling Controls And Settings - GE auisys cs2 User's Reference Manual

Digital anesthesia carestation
Table of Contents
Aisys CS²

Procedures setup

Setting vital capacity defaults

Setting cycling controls and settings

12-16
Use the Procedures menu to set the procedures available to the
user. From this menu select the default settings for Vital Capacity
and Cycling procedures.
1.
From Super User mode, select Procedures.
2.
Select Vital Capacity.
3.
Set Show Vital Capacity to Yes to show this menu item on the
user's Procedures menu. Set to No to hide this menu item on
the user's Procedures menu.
4.
Select Pressure Hold and set to the desired value.
5.
Select Hold Time and set to the desired value.
6.
Set Show PEEP on Exit to Yes to show this menu item on the
user's Vital Capacity menu. Set to No to hide this menu item on
the user's Vital Capacity menu.
7.
Select PEEP on Exit and set to the desired value.
8.
Select Back to return to the Procedures menu.
1.
From Super User mode, select Procedures.
2.
Select Cycling.
3.
Set Show Cycling to Yes to show this menu item on the user's
Procedures menu. Set to No to hide this menu item on the
user's Procedures menu.
4.
Select the procedure to modify from the drop-down list.
5.
Select Actions to set up the cycling procedure. There are a
maximum of seven steps and a minimum of one step for a
cycling procedure.
Select Adjust Settings to make changes to the parameters for
a step.
Select Insert Step to add a step to the cycling procedure.
Select Delete Step to remove a step from the cycling procedure.
6.
Select Close to exit the menu.