Flow Specifications - GE auisys cs2 User's Reference Manual

Digital anesthesia carestation
Table of Contents
Aisys CS²

Flow specifications

11-12
Alternate O2
Minimum flow
Maximum flow
Indicator
Indicator accuracy
Fresh gas
Flow range
Total flow accuracy
O2 flow accuracy
Balance gas flow accuracy
O2 concentration range
O2 concentration accuracy
Electronic mixer response
time
Compensation
Hypoxic guard
Note: Leakage from one gas inlet to another gas inlet is less than 10 ml per
hour.
Note: The stated flow accuracy may not be met for total flows between 200
and 400 ml/min. The total flow will maintain a minimum of 21% O2.
500 ml/min to 700 ml/min
10 l/min to 13 l/min
Flow tube
+/- 5% full scale
0 and 200 ml/min to 15 l/min
Minimum total flow O2 and balance gas is
200 ml/min
+/- 10% or +/- 40 ml/min of setting
(whichever is larger)
+/- 5% or +/- 20 ml/min of setting (whichever
is larger)
+/- 5% or +/- 20 ml/min of setting (whichever
is larger) Air/N2O
21%, 25% to 100%
+/- 2.5% of setting for total flow of greater
than or equal to 1 l/min and less than or
equal to 15 l/min
+/- 5% of setting for total flow of greater than
or equal to 0.4 l/min and less than 1 l/min
+/- 6.5% of setting for total flow of greater
than or equal to 0.2 l/min and less than 0.4
l/min
500 ms (10% to 90% flow step)
Temperature and atmospheric pressure
compensated to standard conditions of 20°C
and 101.3 kPa (14.7 psi)
Electronic
2067226-001