Installing A Cassette - GE auisys cs2 User's Reference Manual

Digital anesthesia carestation
Table of Contents
Aisys CS²

Installing a cassette

13-8
1.
Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2.
If using Aladin2, unlock the cassette handle before installing it in
the active bay.
3.
Insert the cassette into the active bay until a click is heard,
ensuring the cassette is in the correct position.
4.
Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5.
The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Note
Store the cassettes in the cassette storage bay when they are not in
use.