GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Quick Links

GE Healthcare
Innova 2121-IQ/3131-IQ Biplane
Cardiovascular Imaging System
Pre-Installation Manual
OPERATING DOCUMENTATION
5177951-4-1EN
Revision 1
Table of Contents
loading

Summary of Contents for GE Innova 2121-IQ Biplane

  • Page 1 GE Healthcare Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual OPERATING DOCUMENTATION 5177951-4-1EN Revision 1...
  • Page 2 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1...
  • Page 3 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information LANGUAGE Important Information...
  • Page 4 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 5 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 6 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 7 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 8 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 9 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 10 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 11 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 12 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 13 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Important Information...
  • Page 14 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. Important Information...
  • Page 15 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Revision History Part/Rev Date Reason for Change Pages 5177951-4-1E 25 March, 2010 Initial release of 5177951-4-1EN. N rev 1 Revision History...
  • Page 16 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. Revision History...
  • Page 17: Table Of Contents

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table of Contents Chapter 1 INTRODUCTION........................25 1 Object and Scope of this manual....................25 2 Responsibilities of the Purchaser/Customer..................26 3 Pre-Installation Process.........................27 Chapter 2 INNOVA SYSTEM.........................29 1 Description of the System......................29 1.1 Innova LC/LP positioner.......................29...
  • Page 18 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.7 Fluoro UPS (option)......................43 2.8 3kVA Cabinets UPS......................45 2.9 X–Ray Tube Chiller.......................46 2.10 Digital Detector Conditioner....................46 2.11 Innova User Area........................47 2.12 IVUS Option........................48 3 System Compatibility........................50 3.1 System Compatibility Cross–Reference –...
  • Page 19 3.2.1 Door Height........................66 3.2.2 Door Width........................66 3.2.3 Elevator Depth......................66 3.3 Floor.............................66 3.3.1 General Vascular GE Healthcare Policy..............66 3.3.2 Floor requirements when using provided floor anchors..........67 3.3.3 Pan Type Floor Construction Requirement..............67 3.3.4 Hole dimension and preferred location in concrete floor..........69 3.3.5 Water Pipe Requirements....................71...
  • Page 20 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3.4.2.2 Rail Mounting......................78 3.4.2.3 Bolt Specifications....................78 3.4.2.4 Rails selection.....................78 3.5 Walls.............................78 3.5.1 General requirement....................78 3.5.2 Seismic Areas......................78 4 Functional Requirements.......................79 4.1 Electrical Requirements......................79 4.1.1 Optional Isolation transformer specifications...............80 4.2 Room Speaker........................81...
  • Page 21 Direction 5177951-4-1EN, Revision 1 2.1 Floor Loading and Recommended Mounting Methods............124 2.2 Positioner and Table Floor Mounting..................126 2.3 Innova LC Positioner and Omega/Tilt Table Floor Preparation Kits (GE Healthcare supplied)........................135 2.3.1 2285048 - Base Plate Assembly................136 2.3.2 2360538 – Tilt Table Add–On Kit................137 2.3.3 2285050 –...
  • Page 22 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 6.3 Run Interconnection cables through the Conduit...............164 6.4 Electrical Requirements......................165 6.5 Ethernet Network Requirement..................166 7 X-Ray ON Lamp Distribution.......................167 8 Ready For X-Ray Lamp.......................168 9 Room Light Distribution.......................169 10 Room Speaker...........................170...
  • Page 23 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.4 Innova LP4 Positioner Stationary Rails Packaging............192 1.2.5 Innova C1 Frontal, C1 Lateral, and C2 LC/LP Cabinets ...........193 1.2.6 Omega Shipment.......................194 1.2.7 Power Distribution Box (PDB) CE................195 1.2.8 Power Distribution Box (PDB) UL................195...
  • Page 24 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.13 Main Disconnect Panel (PDB - UL)..................232 2.14 Omega IV Table.......................234 2.15 Omega V Long Table with Baseplate................238 2.16 Omega V Long Table.......................242 2.17 Overhead Flat Panel Suspension – Ceiling mounted............246 2.18 8 Monitor Flat panel Suspension - Ceiling Mounted............248...
  • Page 25: Chapter 1 Introduction

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 1 Introduction 1 Object and Scope of this manual This document is intended as a guide and information resource to properly plan and prepare a site for the installation of an Innova system.
  • Page 26: Responsibilities Of The Purchaser/Customer

    Be sure to inform the customer that the cost of any alteration or modification not specified in the sales contract are liable to the customer. The following GE-supplied equipment must be installed by the Hospital's Contractors, per room drawings: Main Disconnect Panel ) •...
  • Page 27: Pre-Installation Process

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3 Pre-Installation Process ROOM LAYOUTS , ELECTRICAL CONNECTIONS , and ADDITIONAL Complete the checklists in PLANNING AIDS of this manual. They represent an important part of the pre–installation process.
  • Page 28 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 3 Pre-Installation Process...
  • Page 29: Chapter 2 Innova System

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 2 Innova System 1 Description of the System 1.1 Innova LC/LP positioner A complete Innova LC/LP positioner comprises: • LC positioner including an X–ray tube housing with a collimator and a Revolution Digital Detector •...
  • Page 30: X-Ray Generator

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 2-1: 1.3 X–Ray generator The Innova System uses two Innova High Voltage Generator System with component parts as follows according to the power requested (100 kW): •...
  • Page 31: Innova Imaging System

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.5 Innova Imaging System An Innova LC/LP System is managed and controlled by a System including: • Two C1 cabinets, inside which are located the RTAC, the DL, (DL in frontal C1 cabinet only), the modem, the HUB, the KVM, •...
  • Page 32: Aw Workstation Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.8 AW workstation option AW workstation option is composed of a workstation, 1 or 2 monitors 19’’ flat panel in the Control Room and 1 monitor (flat panel) fixed on suspension (option) and 1 video switcher. Check Innova...
  • Page 33 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTICE General safety instructions • Move the UPS in an upright position in its original package to the final destination room. To lift the cabinets, use a forklift or lifting belts with spreader bars.
  • Page 34 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTICE Installation safety instructions: • Contractor responsibility: ○ Electrical contractor is responsible for providing and connecting the cables and configuring the PDB in by-pass mode. ○...
  • Page 35: Touch Screen

    If the preinstallation preparation is an IB update, go to IST Job Card Mechanical Installation . Volcano is a manufacturer of IntraVascular UltraSound systems, which are used in the cath lab. GE has a strategic agreement with Volcano. GE will be selling and servicing some of the Volcano products. 1.13.3 Product Description The Volcano IVUS s5i and s5i/GE systems are dedicated equipments for IntraVascular UltraSound imaging, designed as add on equipment to Vascular Imaging systems.
  • Page 36 The s5i/GE rev 1 is used with phased array catheter provided by Volcano (Eagle Eye Gold catheter). The s5i/GE rev 2 includes a new PIM for rotational catheter and another one for FFR capabilities. The s5i/GE rev 2 release is based on •...
  • Page 37: Innova Lc/Lp Product Identification

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Innova LC/LP Product Identification 2.1 Innova LC/LP Basic Product The Innova LC/LP positioner system includes C2 LC/LP cabinet, two positioners (frontal and lateral) and TSUI.
  • Page 38 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 ○ Pivot (Item 9), ○ C–arc (Item 10), ○ Motorized elevator (Item 11) for the Revolution Digital Detector, ○ X–ray tube + tube elevator (Item 12), ○...
  • Page 39: C1 Frontal Cabinet And C1 Lateral Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.2 C1 Frontal Cabinet and C1 Lateral Cabinet Illustration 2-3: C1 frontal and lateral cabinets Chapter 2 Innova System...
  • Page 40: C2 Lc/Lp Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.3 C2 LC/LP Cabinet Illustration 2-4: C2 LC/LP Cabinet 2 Innova LC/LP Product Identification...
  • Page 41: Omega Patient Table

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.4 Omega Patient Table Illustration 2-5: Omega tables 2.5 Tilt Patient Table Illustration 2-6: Tilt table Chapter 2 Innova System...
  • Page 42: User Interface

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.6 User Interface Illustration 2-7: User interface devices 2 Innova LC/LP Product Identification...
  • Page 43: Fluoro Ups (Option)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.7 Fluoro UPS (option) Illustration 2-8: UPS - UL Chapter 2 Innova System...
  • Page 44 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 2-9: UPS - CE 2 Innova LC/LP Product Identification...
  • Page 45: 3Kva Cabinets Ups

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.8 3kVA Cabinets UPS Illustration 2-10: 3 kVA UPS - model 9125 Illustration 2-11: 3 kVA UPS - model 9130 Chapter 2 Innova System...
  • Page 46: X-Ray Tube Chiller

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.9 X–Ray Tube Chiller Illustration 2-12: Coolix 4000 chillers (Frontal and Lateral) 2.10 Digital Detector Conditioner Illustration 2-13: Thermo-con chillers (Frontal and Lateral) 2 Innova LC/LP Product Identification...
  • Page 47: Innova User Area

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.11 Innova User Area Illustration 2-14: Innova User Area devices Chapter 2 Innova System...
  • Page 48: Ivus Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.12 IVUS Option Illustration 2-15: IVUS Rev 1 components 2 Innova LC/LP Product Identification...
  • Page 49 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 2-16: IVUS Rev 2 components Chapter 2 Innova System...
  • Page 50: System Compatibility

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3 System Compatibility 3.1 System Compatibility Cross–Reference – Innova LC/LP Positioner/Table Sub-System Product Name Pre-Installation Document Number Notes LC Combo Positioner (Innova 3131-IQ) This document LC Cardio Positioner (Innova 2121-IQ)
  • Page 51: System Compatibility Cross-Reference - Monitor Support & Suspension Sub-System

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Product Name Pre-Installation Document Number Notes C1 Frontal Cabinet This document C1 Lateral Cabinet This document Innova 2121-IQ, 21 cm Revolution Digital This document detector Innova 3131-IQ, 31 cm Revolution Digital...
  • Page 52 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 3 System Compatibility...
  • Page 53: Chapter 3 Physical Requirements Of Room

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 3 Physical Requirements of room 1 Presentation of the 3 Rooms 1.1 Examination room • Innova LC Positioner • Innova LP4 Positioner • Omega Patient Table •...
  • Page 54: Environmental Requirements/Limitations

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Environmental Requirements/Limitations 2.1 Room Climate NOTICE Avoid extremes in temperatures Innova system room climate requirements – relative humidity and temperature (individual products or components are classed by their installation area)
  • Page 55: Relative Humidity And Temperature

    Storage 10 to 40 ºC (50 to 104 °F); 10 to 90 % RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 °F) and 10 to 80% RH. The Detector chiller HEAT DRY 1 is shipped within GE Healthcare packaging. NOTE: (3a): In use temperature limits specify the range where the system shall work.
  • Page 56: Equipment Heat Output (Dissipation)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTICE In some cases condensation occurs and water drops from outlets and pipes of the air conditioner in the technical room. Therefore, it is critical to install the cabinets where there is no risk of flood from the air conditioner.
  • Page 57 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 3-3: HEAT OUTPUT Maximum Use (maxi‐ Moderate Use (8 cas‐ Typical Use (13 cas‐ Stand by mum peak power dur‐ es / a 10 hour day)
  • Page 58: Fluoro Ups Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 HEAT OUTPUT Maximum Use (maxi‐ Moderate Use (8 cas‐ Typical Use (13 cas‐ Stand by mum peak power dur‐ es / a 10 hour day) es / a 10 hour day)
  • Page 59: Ivus Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 3-4 indicates the typical heat dissipation at nominal load and specify cooling air flow in two situations: • 1000 m (3280 ft) altitude, cooling air between 25°C (77°F) and 30°C(86°F) •...
  • Page 60: Electromagnetic Emission

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 • Immunity Compliance level & recommendations to maintain equipment clinical utility (see Table 3-8,Table 3-9 Table 3-10). NOTE: This system complies with above–mentioned EMC standard when used with supplied cables up to maximum lengths referenced in the MIS MAPS or system cables interconnect diagrams.
  • Page 61: Electromagnetic Immunity Iec 60601-1-2

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Immunity Test IEC 60601–1–2 Test Compliance Level Electromagnetic Environment Level Mains power quality is that of a typical commercial and/or hospital Surge 1 kV line(s) to 1 kV line(s) to lines(s) environment.
  • Page 62: Recommended Separation Distances For Portable And Mobile Rf

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Immunity Test IEC 60601-1-2 Test Level Compliance Level Electromagnetic Environment Note: P is the power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the rec‐...
  • Page 63: Limitations Management

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.3.4 Limitations Management Adhering to the distance separation recommended in Table 3-10, between 150 KHz & 2.5 GHz, will reduce disturbances recorded at the image level, but may not eliminate all disturbances.
  • Page 64: Installations Requirements & Environment Control

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.3.6 Installations Requirements & Environment Control NOTICE In order to minimize interference risks, the following requirements shall apply: • Cables shielding & grounding: All interconnect cables to peripheral devices must be shielded and properly grounded.
  • Page 65: Radiation Protection

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.4 Radiation Protection Because x–ray equipment produces radiation, special precautions may be needed or special site modifications may be required. The General Electric Company does not make recommendations regarding radiation protection.
  • Page 66: Structural Requirements

    GE Healthcare’s Customer is responsible for the structural analysis and mounting of the base plates. If GE Healthcare is forced to mount the base plate, the Local Customer Team must hire a structural engineer to design and approve the mounting method and provide GE Healthcare with an engineering report.
  • Page 67: Floor Requirements When Using Provided Floor Anchors

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-2: The preferred installation method for the Innova LC/LP Positioner or the Omega tables is through- bolting. The through-bolting method can be used in all seismic zones. If through-bolting cannot be used, use provided floor anchors instead.
  • Page 68 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Single Plane and Biplane Innova NOTE: For specific floor preparation procedures, refer to Systems Pre–Installation Kit Installation Procedures . Illustration 3-4: Through–Bolt Supplied (Pan Type Floor Construction)
  • Page 69: Hole Dimension And Preferred Location In Concrete Floor

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: Prepare the floor such that the Tilt Table baseplate will be flush with the floor surface. For alternative table bolts or seismic area, refer to template drawing shown in Illustration...
  • Page 70 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-6: Hole location in concrete floor 3 Structural Requirements...
  • Page 71: Water Pipe Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 3-12: D distance of Illustration 3-6 ANGIO / CARDIO CARDIO / NEURO Omega IV Compact 1395 mm (54.9 in) Omega V Long 1278 mm (50.3 in) 1395 mm (54.9 in)
  • Page 72 GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-8: Water Conduit location Innova LC Positioner and NOTE: For further information on kit 2286398 (S18101SK), refer to Omega/Tilt Table Floor Preparation Kits (GE Healthcare supplied) in Chapter 5, Mounting Requirements. 3 Structural Requirements...
  • Page 73 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-9: Cable Curvature Illustration 3-10: Cable Curvature NOTE: In case of on grade cabling, because of the minimum curvature constraint of 150 mm (6”), the cable will have to come from the side between Innova LC/LP Positioner and patient table.
  • Page 74: Seismic Areas

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3.3.6 Seismic areas Chiller: The seismic kit for both Coolix 4000 is included with the chiller when shipped. In Seismic areas all cabinets must be anchored to the floor. See...
  • Page 75 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-11: Location of Stationary Rails on Ceiling – Recommended max. configuration NOTE: It is recommended to install the LP4 Stationary Rails at the maximum configuration (Illustration 3-11).
  • Page 76 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-12: Location of Stationary Rails on Ceiling – Absolute min. configuration 3 Structural Requirements...
  • Page 77 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 3-13: LP4 Rail Mounting Specifications Chapter 3 Physical Requirements of room...
  • Page 78: Monitor Suspension Rails

    (6228 N). 3.4.2.4 Rails selection Monitor suspension rails in different lengths can be selected. Please refer to the GTC or contact the GE representative. 3.5 Walls 3.5.1 General requirement The C1 Cabinets, the C2 Cabinet, and the PDB Cabinet must be securely fastened to the wall to prevent them from tipping.
  • Page 79: Functional Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 4 Functional Requirements 4.1 Electrical Requirements Innova System requires a specific power line, with three phase (see Illustration in Chapter 6, Power Distribution). Table 3-13: ELECTRICAL REQUIREMENTS: system without Fluoro UPS Model (max Line peak pow‐...
  • Page 80: Optional Isolation Transformer Specifications

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 3-14: ELECTRICAL REQUIREMENTS: system with Fluoro UPS UL Model (max Line peak pow‐ Frequency Mains voltage Peak currents Standby cur‐ Mains wiring Added trans‐ output power)
  • Page 81: Room Speaker

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Line V I peak Illustration 3-14: TNS scheme 4.2 Room Speaker The speaker must be installed out of the patient vicinity in order to not disturb the environment, to give access to the FE for maintenance operations and to medical staff to adjust the tone signal level and be able to see the power LED indicator.
  • Page 82 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 places: Illumination requirements for indoor workplaces corresponding to assembly of medium size electrical components, e.g. control panel) for the electrical industry). 4 Functional Requirements...
  • Page 83: Emergency

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 5 Emergency During an examination, any operator can encounter two main cases of failures. 5.1 Main power supply cut In this case, refer to Chapter 6, Physical Runs.
  • Page 84: Insite

    A request form has been defined. For more information, please refer to Chapter 8, IP Addressing Process or contact your GE Healthcare OLC representative. NOTICE The C1 frontal cabinet comes equipped with a Firewall unit. The hospital network must be capable of connecting to this firewall.
  • Page 85: Chapter 4 Room Layouts

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 4 Room Layouts 1 Room Layout Considerations 1.1 Service Access Allow appropriate space for service access of equipment. Consult component pre–installation directions for clearance information.
  • Page 86: Patient Environment Equipment

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3kVA UPS FOR CE CONFIGURATION WITHOUT FLUORO UPS), including UPS and Fluoro UPS outputs. To remove input power from THE INPUT CABLE OF THE 3kVA UPS (FOR CE CONFIGURATION WITHOUT FLUORO UPS), turn OFF the PDB main breaker Protect the Emergency Stop from accidental actuation.
  • Page 87: Preference Cabinet Locations

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Zone 2: Variable lighting without specular reflection on display and monitor screens. Indirect lighting but preferable from ceiling. Automatically cut by the Innova system. Zone 3: Variable lighting without artifact on display and monitor screens. manually cut.
  • Page 88: Layout Constraints

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 4-2: NOTE: The order of the cabinets above is C1 frontal, C1 lateral and C2 cabinets. This is just an example, cabinets can follow different orders.
  • Page 89 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Location Constraint Name Parameter Specification Comment Technical room Detector conditioner loca‐ In tech room Max. 3 tion m (9 ft, 10 in) below LP4 rails Technical room...
  • Page 90 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Location Constraint Name Parameter Specification Comment Exam room minimum rail length past Length 1.3 m (4 ft, 7 in) LP4 parking at patient foot end not...
  • Page 91: Room Layout Drawings

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Room Layout Drawings WARNING LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY. THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER, DETECTOR CHILLER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF PULSED–AIR.
  • Page 92 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 IVUS Volcano ready kit NOTE: The phone and/or Network drop outlet must be located less than 1 meter (3 feet) from the C1 Frontal Cabinet (20).
  • Page 93 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 4-3: Room Layout for Innova System Chapter 4 Room Layouts...
  • Page 94: Ivus Option Room Layout

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3 IVUS Option Room Layout Illustration 4-4: IVUS Rev 1 3 IVUS Option Room Layout...
  • Page 95 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 4-5: IVUS Rev 2 Chapter 4 Room Layouts...
  • Page 96 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 3 IVUS Option Room Layout...
  • Page 97: Chapter 5 Physical Characteristics

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 5 Physical Characteristics 1 Dimension Drawings Refer to this section for the dimensional drawings of the components of the Innova 2121-IQ and 3131-IQ systems. This section also contains the LC, LP4 and patient table sweep volume curves.
  • Page 98 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-1: Innova LC Positioner Dimensions - Side View Illustration 5-2: Innova LC Positioner Dimensions - Top View 1 Dimension Drawings...
  • Page 99 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-3: Innova LC Positioner Dimensions - Front View Chapter 5 Physical Characteristics...
  • Page 100 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-4: Innova LP4 Positioner Dimensions - Side View 1 Dimension Drawings...
  • Page 101 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-5: Innova LP4 Positioner Dimensions - Top View Chapter 5 Physical Characteristics...
  • Page 102 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-6: Innova LP4 Positioner Dimensions - Front View 1 Dimension Drawings...
  • Page 103 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-7: Omega IV Table Dimensions Chapter 5 Physical Characteristics...
  • Page 104 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: The 500 mm (19.7 in) minimum clearance between the table foot end and nearest object must take Into account any table devices installed on the table end rail. If there are any devices installed on the table foot end, the width of these devices must be added to the existing 500 mm (19.7 in) to maintain absolute minimum distance of 500 mm (19.7 in).
  • Page 105 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-8: Omega V Table Dimensions Chapter 5 Physical Characteristics...
  • Page 106 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: The 500 mm (19.7 in) minimum clearance between the table foot end and nearest object must take Into account any table devices installed on the table end rail. If there are any devices installed on the table foot end, the width of these devices must be added to the existing 500 mm (19.7 in) to maintain absolute minimum distance of 500 mm (19.7 in).
  • Page 107 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-9: Omega Table side clearence (CPR access) Chapter 5 Physical Characteristics...
  • Page 108 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-10: Gas box outlets Omega table 1 Dimension Drawings...
  • Page 109 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-11: Innova LC/LP Positioner And Omega Patient Table Relative Positions - Side View NOTE: (1) For LC isocenter to Table distance, refer to Table 5-2.
  • Page 110 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-12: Innova LC/LP Positioner and Omega Patient Table Relative Positions - Top View Table 5-2: Patient Table - LC isocenter distances ANGIO / CARDIO CARDIO / NEURO Omega IV 1395 mm (54.9 in)
  • Page 111 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-13: LP4 Positioner Cable Drape Length Maximum cable drape length is 6 m / 236 in (with vinyl zipper cable cover of 6.2 m / 244 in) from LP4 to ceiling exit point.
  • Page 112 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-14: C2 LC/LP Cabinet Dimensions 1 Dimension Drawings...
  • Page 113 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-15: C1 Frontal Cabinet Dimensions Chapter 5 Physical Characteristics...
  • Page 114 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-16: C1 Lateral Cabinet Dimensions 1 Dimension Drawings...
  • Page 115 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-17: Coolix 4000 Chiller Dimensions and Floor Space Diagram Exact dimensions of the Coolix 4000 chiller depend on model type. Required floor space depends on ambient room temperatures. When in doubt, allow for maximum floor space.
  • Page 116 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-18: Thermo-Con Detector Conditioner (and Mounting brackets) Dimensions 1 Dimension Drawings...
  • Page 117 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-19: 3 kVA Cabinets UPS - model 9125 Chapter 5 Physical Characteristics...
  • Page 118 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-20: 3 kVA Cabinets UPS - model 9130 1 Dimension Drawings...
  • Page 119 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-21: Fluoro UPS Cabinets (Optional) - UPS UL The left, right or back side of the UPS cabinet can be positioned against the wall. The front side of the UPS cabinet must be accessible for maintenance operation.
  • Page 120 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-22: Fluoro UPS Cabinets (Optional) - UPS CE The UPS cabinet can be positioned against the wall but, in order to improve the ventilation and to make easier the maintenance operations for UPS and battery, we recommend a minimum distance of 200 mm (8”) from the wall.
  • Page 121 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-23: DL Keypad Dimensions Chapter 5 Physical Characteristics...
  • Page 122 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-24: DL Image Monitor Dimensions 1 Dimension Drawings...
  • Page 123 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-25: VCIM Chapter 5 Physical Characteristics...
  • Page 124: Mounting Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Mounting Requirements 2.1 Floor Loading and Recommended Mounting Methods Table 5-3. To obtain floor loading and recommended mounting methods for components not specified in Table...
  • Page 125 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 DIMENSIONS MM (INCHES) LOAD BEARING PRODUCT OR WEIGHT/ OC‐ WEIGHT AREA MM (IN‐ MOUNTING METHOD COMPONENT CUPIED AREA WIDTH DEPTH HEIGHT KG (LBS) CHES) UPS 3 kVA - model 346 (13.6)
  • Page 126: Positioner And Table Floor Mounting

    2.2 Positioner and Table Floor Mounting The distances between the Innova LC Positioner and the Tables are critical for a proper clinical usage. For this reason, GE Healthcare provides two floor mounting templates to ensure these components are properly placed in relation to one another.
  • Page 127 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-26: Innova LC Positioner Floor Mounting Methods (1/2) Chapter 5 Physical Characteristics...
  • Page 128 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-27: Innova LC Positioner Floor Mounting Methods (2/2) 2 Mounting Requirements...
  • Page 129 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-28: Cable Conduit For On-Grade Floor Anchor Kit Chapter 5 Physical Characteristics...
  • Page 130 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-29: Inner Base Plate For Above Grade Floor Anchor Kit 2 Mounting Requirements...
  • Page 131 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-30: Fixing Bolt Overview NOTE: For more details on Omega/Tilt Baseplate, refer to Illustration 5-31. Chapter 5 Physical Characteristics...
  • Page 132 Pull out efforts and recommendations about chemical anchors not provided by GE. The following table provides the recommended chemical anchors for Omega/LC baseplates and for the floor plate ordered locally that they could be used instead of bolts provided by GE. Table 5-5: Chemical anchors Pull out efforts and recommendations...
  • Page 133 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-31: Gantry and table mounting holes Chapter 5 Physical Characteristics...
  • Page 134 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 5-6: D distance of Illustration 5-31 ANGIO / CARDIO CARDIO / NEURO Omega IV Compact 1395 mm (54.9 in) Omega V Long 1278 mm (50.3 in) 1395 mm (54.9 in)
  • Page 135: Innova Lc Positioner And Omega/Tilt Table Floor Preparation Kits

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.3 Innova LC Positioner and Omega/Tilt Table Floor Preparation Kits (GE Healthcare supplied) Illustration 5-32: Chapter 5 Physical Characteristics...
  • Page 136: 2285048 - Base Plate Assembly

    GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-33: Anchor kit choice All GE supplied vascular system floor preparation are contained in catalog. There are some additional gantry/table mounting kits based on each mounting method: through bolts or floor anchors. •...
  • Page 137: 2360538 - Tilt Table Add-On Kit

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Item Name Part # Description Quan. Notes Hex Head 5166535 Hex Head Screw 1/2”-13UNC L1 2 pc Screw 1/2” 12.9 Black Grease 2295599 Lithium grease lubricant 100g...
  • Page 138: 2285050 - Assembly And Separation Select Kit

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Item Name Part # Description Quan. Notes Nut, Hex 99141607 Hex Nut M20 STL galvanized, two 12 pc for each bolt Dowel 2290937 Wood Dowel; 24 mm diameter...
  • Page 139: 2286398 - On Grade, Floor Anchor (S18101Sk)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Item Name Part # Description Quan. Notes Nut, Hex 99141607 Hex Nut M20 STL galvanized two 32 pc same as above for each bolt Grout assy...
  • Page 140: 2285053 - Through Bolts With Insert (S18101Sf)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Item Name Part # Description Quan. Notes Grout assy 2285055 - 10 kg Powdered Mortar Ardex 1 kit Used in constructing LC Po‐ K-15 sitioner grout dam...
  • Page 141: 2285632 - Floor Plate / Base Plate Assembly (S18101Sg)

    4 only are used for Omega Pass NOTE: The kit 2285632 Floor plate / Baseplate assembly is needed when the floor plate is installed. This part isn’t GE part. It’s ordered locally. 2.3.4 2285056 - A/C/N Templates Table 5-15: Item...
  • Page 142 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-34: Baseplate plan 2359194 - Angio configuration 2 Mounting Requirements...
  • Page 143 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 5-35: Baseplate plan 2359194 - Cardio configuration Chapter 5 Physical Characteristics...
  • Page 144: Injector Mounting Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.4 Injector Mounting Requirements CAUTION Table accessory rail load consideration: The maximum load per table accessory rail is 40 kg at 150 mm (60 N.m). Therefore: •...
  • Page 145: Chapter 6 Electrical Connections

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 6 Electrical Connections 1 Power Distribution NOTE: All short AC power cable less than 3 meters, between peripherals and respective power strip, are not shown. All have a single phase line, a neutral line and a ground wire.
  • Page 146 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-2: Power distribution - UL Cable CE UPS configuration UL UPS configuration Max length Recommended (1) Type Max gauge Recommended (1) Type R rectifier 5 x 10 mm²...
  • Page 147: Power Distribution System

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Power Distribution System NOTICE It is recommended to separate power supply cables from the other cables. 2.1 Power Distribution Box - CE Illustration 6-3: PDB CE - 50 Hz...
  • Page 148 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-4: PDB Schematic CE - 1/2 2 Power Distribution System...
  • Page 149 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-5: PDB Schematic CE - 2/2 Chapter 6 Electrical Connections...
  • Page 150: Power Distribution Box - Ul

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.2 Power Distribution Box - UL Illustration 6-6: PDB UL - 60 Hz (US only) 2 Power Distribution System...
  • Page 151 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-7: PDB Schematic UL - 1/2 Chapter 6 Electrical Connections...
  • Page 152 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-8: PDB Schematic UL - 2/2 2 Power Distribution System...
  • Page 153: Grounding

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3 Grounding A vascular lab is a critical care area and requires a special grounding system for patient safety. An equipotential grounding system is recommended for meeting patient safety requirements.
  • Page 154 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-9: Grounding distribution 3 Grounding...
  • Page 155 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 6-1: CABLE FROM WIRE GAUGE 28189 C1 Frontal Cabinet LCD Monitor Suspension AWG6 28190 C1 Frontal Cabinet LCD Monitor Suspension AWG6 2308096 C1 Frontal Cabinet...
  • Page 156 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 CABLE FROM WIRE GAUGE hospital C2 Cabinet 6 mm² AWG8 hospital Fluoro UPS 10 mm² AWG6 hospital Fluoro UPS 10 mm² AWG6 hospital Frontal XRT Chiller 6 mm²...
  • Page 157: Power And Grounding Recommendations

    0.108 0.114 0.12 NOTE: These 3 phases cables are not furnished by GE Healthcare. Provided by installer. • These cables must be kept separated as much as possible from room system cables. • The shield of any shielded cable coming from the distribution cannot replace the ground wire.
  • Page 158 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 6-3: System FEEDER from hospital Power supply voltage 480 V Panel (PDB) to C1 run in m (ft) 8 (26) 16 (52) 24 (79) Feeder run length:...
  • Page 159: Ivus Rev 1 Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 5 IVUS Rev 1 Option 5.1 Wiring Diagram The connection cables between the IVUS Control Room and Procedures Room components run in a dedicated under floor conduit (see Section 5.2).
  • Page 160: Run Interconnection Cables Through The Conduit

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 • Establish a 76 mm (3”) conduit below floor from the IVUS CPU location in the Control Room, to the foot end area of the Patient Table in the Procedures Room: ○...
  • Page 161: Electrical Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote controls expected. • Ensure sufficient cable length on the Procedures Room end of the cables according to the...
  • Page 162 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 DICOM Image Storage: Saving patient cases to DVD: The archived images are stored in DICOM format with Volcano s5i system acting as a File Set Creator (FSC), following the guidelines in the 2004 DICOM 3.0 specification.
  • Page 163: Ivus Rev 2 Option

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 6 IVUS Rev 2 Option 6.1 Wiring Diagram The connection cables between the IVUS Control Room and Procedures Room components run in a dedicated under floor conduit (see Section 6.2).
  • Page 164: Run Interconnection Cables Through The Conduit

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 • Locate the position of the IVUS components in the Control Room, according to the workflow requirement of the operators. • Establish a 76 mm (3”) conduit below floor from the IVUS CPU location in the Control Room, to the foot end area of the Patient Table in the Procedures Room: ○...
  • Page 165: Electrical Requirements

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: You should consider running the ECG cable through the conduit in case, the customer requires interconnecting to a NON Mac-Lab ECG at the table base.
  • Page 166: Ethernet Network Requirement

    SOP class as an SCU (service class user). NOTE: All this applies to the IVUS mode of the s5i/GE, not for the FFR mode. For more information, see the Volcano s5 DICOM Conformance Statement located in the 806365-005 Volcano Service Manual.
  • Page 167: Ray On Lamp Distribution

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 7 X-Ray ON Lamp Distribution The fitter shall install an X-Ray On lamp in the Exam room, and in the control room if necessary, outside the patient area, so that they are connected to the equipment through connection terminals.
  • Page 168: Ready For X-Ray Lamp

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 8 Ready For X-Ray Lamp Innova 2121-IQ/3131-IQ provides contacts for driving a Ready for X-Ray lamp. These contacts are terminals 5 and 6 of TB2 bulkhead connector of the C2 cabinet.
  • Page 169: Room Light Distribution

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 9 Room Light Distribution Innova 2121-IQ/3131-IQ has the ability to control an external relay that applies power to the room light (dry contacts). The relay is to be provided by the hospital or contractor.
  • Page 170: Room Speaker

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 10 Room Speaker The remote speaker provides the operator in the exam room, the indication of record exposure sequence (same level information as provided by the control room control panel).
  • Page 171: Door Interlocks

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 11 Door Interlocks CAUTION IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of the field service to check that this requirement is not in contradiction with local regulation.
  • Page 172: Emergency

    UPS would provide to the customer about 10 minutes of autonomy. Optional Isolation The UPS output shall follow the same output specifications as shown in transformer specifications in Chapter 3, Contact your local GE Healthcare representative for more information and/or public power distribution analysis. 12 Emergency...
  • Page 173: Injectors

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 13 Injectors The injector is provided with an additional ground cable (P/N #2135737) to meet equipotentiality requirement at patient vicinity. Injector emergency power off: • if no emergency power off is required for the injector, the injector can be powered from a wall outlet socket.
  • Page 174: Advantage Windows

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 14 Advantage Windows A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main disconnect room device. Verify that the ground between AW, AW spliter and monitors ground are equipotential (0.1 ohm equipotentiallity between room ground point (busbar) and wall outlets).
  • Page 175: Mac-Lab System Ex

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 15 MAC-LAB System EX Electrical Ducts in For location of MAC–LAB Acquisition Unit Floor mount kit 408431–001, refer to Cable Channeling. For MAC–LAB System EX, refer to: - Marquette document P/N 2000465–001 Preinstallation Guide...
  • Page 176: Mis (Master Interconnect System)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 16 MIS (Master Interconnect System) Innova system interconnect cables are described in MIS (Master Interconnect System) documents. These documents specify all interconnections between components within the system.
  • Page 177: Cable Channeling

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 17 Cable Channeling 17.1 General High voltage and power cables must be separated from other cables. Use a separate trough in the duct system, or use a separate conduit. Minimize cable length between the line disconnect and the System Cabinet power unit to reduce voltage regulation problems and wiring costs.
  • Page 178 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 NOTE: MEDRAD Avanta Table mount: A 76.2 mm (3 in) and max 25 m (984 in) length conduit between technical room and patient room shall be prepared below the floor for the three injector cables.
  • Page 179 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-18: Floor ducts and outlets Chapter 6 Electrical Connections...
  • Page 180: Physical Runs

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 18 Physical Runs NOTICE All lengths of cable are: • in useable meter when you look at group level, or • in meters (connector to connector) when you look at the cable level.
  • Page 181 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-20: CABLE GROUP 1 – FROM TECHNICAL AREA TO EXAM ROOM Chapter 6 Electrical Connections...
  • Page 182 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-21: CABLE GROUP 2– FROM TECHNICAL AREA TO CONTROL ROOM 18 Physical Runs...
  • Page 183 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-22: CABLE GROUP 3 - FROM TECHNICAL AREA TO TECHNICAL AREA Chapter 6 Electrical Connections...
  • Page 184 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-23: CABLE GROUP 4 - FROM TECHNICAL AREA TO EXAM AREA 18 Physical Runs...
  • Page 185 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 6-24: CABLE GROUP 5 - FROM TECHNICAL AREA TO EXAM AREA Chapter 6 Electrical Connections...
  • Page 186 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 18 Physical Runs...
  • Page 187: Chapter 7 Additional Planning Aids

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 7 Additional Planning Aids 1 Shipping Information 1.1 Product Shipping Information Refer to Table 7-1. To obtain shipping information for components not specified in Table...
  • Page 188: Detail Of Innova Shipping Information

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2 Detail of Innova Shipping Information 1.2.1 Innova LC Positioner Gantry On Shipping Dolly Illustration 7-1: 1 Shipping Information...
  • Page 189: Innova Lc Positioner Air Shipment

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.2 Innova LC Positioner Air Shipment Illustration 7-2: Chapter 7 Additional Planning Aids...
  • Page 190: Innova Lp4 Positioner Gantry On Shipping Dolly (Longer And Shorter Dollies)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.3 Innova LP4 Positioner Gantry On Shipping Dolly (Longer and shorter dollies) Illustration 7-3: On Longer type dolly NOTE: The dimensions above are for shipping. When the LP4 and dolly are in rolling configuration for hospital access, the dimensions are Height : 2001 mm (78.8 in) and...
  • Page 191 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 7-4: On shorter type dolly Chapter 7 Additional Planning Aids...
  • Page 192: Innova Lp4 Positioner Stationary Rails Packaging

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.4 Innova LP4 Positioner Stationary Rails Packaging Illustration 7-5: 1 Shipping Information...
  • Page 193: Innova C1 Frontal, C1 Lateral, And C2 Lc/Lp Cabinets

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.5 Innova C1 Frontal, C1 Lateral, and C2 LC/LP Cabinets Illustration 7-6: C1 and C2 Cabinets on pallets The shipping weight is of about 36.7 kg (81 lbs) per dolly.
  • Page 194: Omega Shipment

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.6 Omega Shipment Illustration 7-7: 1 Shipping Information...
  • Page 195: Power Distribution Box (Pdb) Ce

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.7 Power Distribution Box (PDB) CE Illustration 7-8: 1.2.8 Power Distribution Box (PDB) UL Illustration 7-9: Chapter 7 Additional Planning Aids...
  • Page 196: Other Elements Package

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 1.2.9 Other Elements Package NOTE: All OEM parts are shipped inside there original boxes group as needed on pallets. Illustration 7-10: 1 Shipping Information...
  • Page 197: Tools And Test Equipment

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Tools and Test Equipment Refer to Table 7-2. To obtain a list of tools and test equipment for components not specified in Table 7-2, refer to the appropriate component Pre-Installation Manual listed in Chapter 2, System Compatibility.
  • Page 198 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 PRODUCT OR COMPO‐ TOOL OR TEST EQUIP‐ USED FOR SOURCE RECEIVED (DATE) NENT MENT Ethernet adaptation kit for General use (to be or‐ laptop 2128794 dered before delivery of...
  • Page 199: Route Survey

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 3 Route Survey 3.1 Step One — Sketch Start preparing Route Survey by sketching a floor plan of the hospital or clinic which will receive the equipment. Include all areas on the delivery route from outside the building to destination. See Illustration 7-11.
  • Page 200 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Table 7-3: Route Survey Ref. no. Area: loading dock, doorway, hallway, turn, elevator, ob‐ Limitations: loading capacity, height, width, depth flooring struction, etc. material, radius, etc.
  • Page 201: European Process Order Select

    Installation Specialist on Site. All cable and component selections must be sent at least 6 weeks before RSDD to your PMC. The selection tool is available on the website: http://gein.euro.med.ge.com/european_installations/ Chapter 7 Additional Planning Aids...
  • Page 202: Pre-Installation Checklist

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 5 Pre-Installation Checklist GE Healthcare Site Readiness Checklist GEHC Global Order # : Customer: GEHC On-site Representative : MI Supplier: Name of customer reviewed with :...
  • Page 203 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Access to a working phone at the facility for emergency use, including MR magnet delivery. All walls primed (final coat not needed on Day 1), and counter tops that will support equipment must be installed.
  • Page 204 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 5 Pre-Installation Checklist...
  • Page 205: Chapter 8 Ip Addressing Process

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 8 IP Addressing Process 1 IP Addressing Process To obtain an IP address, contact the following for your pole: • GEMSAM: Contact: OnLine Center–Americas, Network Products and Services (NP&S) Telephone: 1–800–321–7937...
  • Page 206 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 1 IP Addressing Process...
  • Page 207: Chapter 9 Seismic Calculations

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Chapter 9 Seismic Calculations 1 Overview Seismic requirements are determined and specified by the hospital architect of record and may require approval by the specific state or country agency.
  • Page 208: Calculations

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2 Calculations Seismic calculations are included for the following: • C1 Frontal Cabinet, Section 2.1 • C1 Lateral Cabinet, Section 2.2 • C2 LC/LP Cabinet, Section 2.3 •...
  • Page 209: C1 Frontal Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.1 C1 Frontal Cabinet Illustration 9-1: C1 frontal cabinet - Slab on grade / Upper floor Chapter 9 Seismic Calculations...
  • Page 210: C1 Lateral Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.2 C1 Lateral Cabinet Illustration 9-2: C1 Lateral cabinet - Slab on Grade / Upper Floor 2 Calculations...
  • Page 211: C2 Lc/Lp Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.3 C2 LC/LP Cabinet Illustration 9-3: C2 LP/LC cabinet - Slab on floor / Upper floor Chapter 9 Seismic Calculations...
  • Page 212: Coolix 4000 Chiller

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.4 Coolix 4000 Chiller Illustration 9-4: Coolix 4000 chiller - Slab on grade 2 Calculations...
  • Page 213 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-5: Coolix 4000 chiller - Upper floor Chapter 9 Seismic Calculations...
  • Page 214: Thermocon Detector Chiller

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.5 Thermocon Detector Chiller Illustration 9-6: Thermocon Detector Chiller - Slab on grade 2 Calculations...
  • Page 215 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-7: Thermocon Detector Chiller - Upper floor Chapter 9 Seismic Calculations...
  • Page 216: Fluoro Ups Ce Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.6 Fluoro UPS CE Cabinet Illustration 9-8: Fluoro UPS CE cabinet - Slab on grade 2 Calculations...
  • Page 217 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-9: Fluoro UPS CE cabinet - Upper floor Chapter 9 Seismic Calculations...
  • Page 218: Fluoro Ups Ul Cabinet

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.7 Fluoro UPS UL Cabinet Illustration 9-10: Fluoro UPS UL cabinet - Slab on grade 2 Calculations...
  • Page 219 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-11: Fluoro UPS UL cabinet - Upper floor Chapter 9 Seismic Calculations...
  • Page 220: Kva Cabinet Ups - Model 9125

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.8 3 kVA Cabinet UPS - model 9125 Illustration 9-12: 3 kVA Cabinet UPS (model 9125) - Slab on grade 2 Calculations...
  • Page 221 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-13: 3 kVA Cabinet UPS (model 9125) - Upper Floor Chapter 9 Seismic Calculations...
  • Page 222: Kva Cabinet Ups - Model 9130

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.9 3 kVA Cabinet UPS - model 9130 Illustration 9-14: 3 kVA Cabinet UPS (model 9130) - Slab on grade 2 Calculations...
  • Page 223 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-15: 3 kVA Cabinet UPS (model 9130) - Elevated Floor Chapter 9 Seismic Calculations...
  • Page 224: Lc Positioner

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.10 LC Positioner Illustration 9-16: LC Positioner - Slab on grade 1/2 2 Calculations...
  • Page 225 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-17: LC Positioner - Slab on grade 2/2 Chapter 9 Seismic Calculations...
  • Page 226 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-18: LC Positioner - Upper floor - 1/2 2 Calculations...
  • Page 227 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-19: LC Positioner - Upper floor 2/2 Chapter 9 Seismic Calculations...
  • Page 228: Lp4 Positioner

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.11 LP4 Positioner Illustration 9-20: LP4 Positioner 1/2 2 Calculations...
  • Page 229 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-21: LP4 Positioner 2/2 Chapter 9 Seismic Calculations...
  • Page 230: Main Disconnect Panel (Pdb - Ce)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.12 Main Disconnect Panel (PDB - CE) Illustration 9-22: Main Disconnect Panel (PDB - CE) - Slab on grade 2 Calculations...
  • Page 231 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-23: Main Disconnect Panel (PDB - CE) - Upper floor Chapter 9 Seismic Calculations...
  • Page 232: Main Disconnect Panel (Pdb - Ul)

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.13 Main Disconnect Panel (PDB - UL) Illustration 9-24: Main Disconnect Panel (PDB - UL) - Slab on grade 2 Calculations...
  • Page 233 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-25: Main Disconnect Panel (PDB - UL) - Upper floor Chapter 9 Seismic Calculations...
  • Page 234: Omega Iv Table

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.14 Omega IV Table Illustration 9-26: Omega IV Table - Slab on grade 1/2 2 Calculations...
  • Page 235 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-27: Omega IV Table - Slab on grade 2/2 Chapter 9 Seismic Calculations...
  • Page 236 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-28: Omega IV Table - Upper floor 1/2 2 Calculations...
  • Page 237 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-29: Omega IV Table - Upper floor 2/2 Chapter 9 Seismic Calculations...
  • Page 238: Omega V Long Table With Baseplate

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.15 Omega V Long Table with Baseplate Illustration 9-30: Omega V Long table w/baseplate - Slab on grade 1/2 2 Calculations...
  • Page 239 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-31: Omega V Long table w/baseplate - Slab on grade 2/2 Chapter 9 Seismic Calculations...
  • Page 240 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-32: Omega V Long table w/baseplate - Upper floor 1/2 2 Calculations...
  • Page 241 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-33: Omega V Long table w/baseplate - Upper floor 2/2 Chapter 9 Seismic Calculations...
  • Page 242: Omega V Long Table

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.16 Omega V Long Table Illustration 9-34: Omega V long table - Slab on grade 1/2 2 Calculations...
  • Page 243 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-35: Omega V Long table - Slab on grade 2/2 Chapter 9 Seismic Calculations...
  • Page 244 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-36: Omega V Long table - Upper floor 1/2 2 Calculations...
  • Page 245 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-37: Omega V Long table - Upper floor 2/2 Chapter 9 Seismic Calculations...
  • Page 246: Overhead Flat Panel Suspension - Ceiling Mounted

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.17 Overhead Flat Panel Suspension – Ceiling mounted Illustration 9-38: Overhead Flat Panel Suspension 1/2 2 Calculations...
  • Page 247 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-39: Overhead Flat panel suspension - 2/2 Chapter 9 Seismic Calculations...
  • Page 248: Monitor Flat Panel Suspension - Ceiling Mounted

    Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 2.18 8 Monitor Flat panel Suspension - Ceiling Mounted Illustration 9-40: 8 Monitor Flat panel Suspension - ceiling mounted 1/2 2 Calculations...
  • Page 249 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 Illustration 9-41: 8 Monitor Flat panel Suspension - ceiling mounted 2/2 Chapter 9 Seismic Calculations...
  • Page 250 Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual GE Healthcare Direction 5177951-4-1EN, Revision 1 This page left intentionally blank. 2 Calculations...
  • Page 252 © 2010 General Electric Company. GE Medical Systems, a General Electric Company, going to market as GE Healthcare. 283, rue de la Minière 78530, Buc FRANCE www.gehealthcare.com...