Reference Manual
00809-0200-4648, Rev BB
Figure 2-7. Connection Diagram for Rosemount 648 Wireless and Field Communicator
1
2
P/N 00753-9200-0020
3
4
COMM
2.4.1

Field Communicator

When performing device configuration directly, connect the bench equipment as shown in
Figure 2-7
using a Field Communicator, any configuration changes must be sent to the transmitter by
using the Send key (F2).
The Field Communicator will search for a HART-compatible device and indicate when the
connection is made. If the Field Communicator fails to connect, it will indicate that no
device was found. If this occurs. refer to
Note
For HART Wireless transmitter communication via a Field Communicator, a Rosemount 648
Wireless Device Dashboard (DD) is required. Rosemount 648 Wireless Transmitters
equipped with Rosemount X-well Technology requires DD revision 648 Dev. 4 Rev. 1 or
higher to view Rosemount X-well functionality. To obtain the latest DD, visit the 475 Field
Communicator System Software and Device Description site at:
EmersonProcess.com/Field-Communicator
2.4.2

AMS Device Manager and AMS Wireless Configurator

When configuring the Rosemount 648 Wireless using AMS Device Manager or AMS Wireless
Configurator, double click the Rosemount 648 Wireless device icon (or right click and select
Configure/Setup), then select the Configure/Setup tab. AMS Device Manager
configuration changes are implemented when the Apply button is selected.
Note
For HART Wireless transmitter communication via AMS, a Rosemount 648 Wireless Device
Dashboard (DD) is required. Rosemount 648 Wireless Transmitters equipped with
Rosemount X-well Technology requires DD revision 648 Dev. 4 Rev. 1 or higher to view
Rosemount X-well functionality. To obtain the latest DD, visit the Emerson Process
Management. Easy Upgrade site at:
EmersonProcess.com/Device-Install-Kits
Configuration
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
above, and turn on the field communicator by pressing the ON/OFF key. When
Section 6: