Operating Instructions - Sanyo MLS-3020U Instruction Manual

Labo autoclave
Table of Contents

OPERATING INSTRUCTIONS

Starting Operation
1. Exhaust tank setup
(1)Remove the exhaust tank from the main unit and, from the
opening at the top, fill the tank with water up to the LOW
level mark.
(2) Insert the exhaust hose that is connected to the interior of
the main unit into the opening at the top of the tank. Push it
in securely until the gasket is firmly engaged.
Turning the gasket only one direction during insertion can cause
the hose to become twisted and possibly lead to malfunctions.
Turn the gasket alternately to the left and right when inserting it.
(3) Load the exhaust tank into the main unit with taking care the ex-
haust hose not to bend or twist.
The amount of water in the exhaust tank increases in every op-
eration of sterilizer. If the water reaches the m f l j | level mark,
discard this water and pour new water up to the iMMffi mark. Do-
not try to pull the tank out from the main unit, if the exhaust hose
is still attached.
2. If media spills are discovered in the bottom
of chamber, proceed as follows:
Chamber should be cleaned daily when processing media. See
Page 28 (Maintenance and care) for cleaning.
3. Pour water for evaporation into the cham-
ber.
(1)Turn the handle counterclockwise until it stops. Open the
door with turning the arm counterclockwise.
(2) Close the drain valve.
(3) Pour in water until the top edge of the water level fitting on
the heater cover is covered. The level of water will be lower
after each sterilizing cycle, therefore this water must be
replenished.
19
Exhaust hose
Insert, turning
alternately
left and right
Gasket
Heater cover
///////W///////
Sterilization water