Single Ramp & Dwell

Functions

Functions
This type of controller has two step ramp and dwell programming capabilities. The Ramp and Dwell
can be configured to three different modes.
1. Mode 1 (DWELL) The dwell time begins once the setpoint reaches the set threshold. The END
TYPE action is executed when the dwell timer reaches the end.
2. Mode 2 (DELY) The timer starts immediately upon instrument power-up or, when run is
selected. The instrument remains in standby until the time has elapsed. After the time has elapsed,
the instrument controls up to setpoint.
These instructions are used with the Single Setpoint models with OTP only 
(See models listed on front page).