F15-8x50c
Instruction manual
50120015-1
June 2009
Table of Contents
loading

Summary of Contents for Thermo Scientific FIBERLite F15-8x50c

  • Page 1 Thermo Fisher Scientific FIBERLite F15-8x50c Instruction manual 50120015-1 June 2009...
  • Page 2 IMPORTANT! Please reference this page for the most up-to-date information on the following* • web site addresses • contact information • copyright and trademark information *All subsequent pages in this manual may have incorrect web site addresses and contact information. ©2013 Thermo Fisher Scientific Inc.
  • Page 3 Thermo Fisher Scientific FIBERLite F15-8x50c Instruction Manual 50120015-1 June 2009...
  • Page 4: Table Of Contents

    Aerosol-tight Applications ....................5-1 Basic Principles ......................5-2 Fill Level ........................5-2 Checking the Aerosol-Tightness ................5-2 Chapter 6 Maintenance and Care ......................6-1 Intervals ........................6-2 Cleaning ........................6-2 Disinfection ......................6-2 Handling and Care of Rotors and Accessories made of Aluminum ......6-3 Decontamination .....................6-4 Thermo Scientific 75003663 F15-8x50c...
  • Page 5 Table of Contents Autoclaving ......................6-4 Service of Thermo Fisher Scientific ................6-5 RCF-Values .......................A-1 Chemical Compatibility Chart ................B-1 Warranty and recommended Service Life ............C-1 Contact Information ....................D-1 75003663 F15-8x50c Thermo Scientific...
  • Page 6: Preface

    • Do not remove the magnet at the rotor bottom • Do not use rotors which show any signs of corrosion and/or cracks. • Use only with rotors that have been loaded properly. • Never overload the rotor. Thermo Scientific 75003663 F15-8x50c...
  • Page 7 ATTENTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. 75003663 F15-8x50c Thermo Scientific...
  • Page 8: Rotor Specification

    Rotor Specification Contents “Heraeus” page 1-2 • “Sorvall” page 1-4 • Thermo Scientific 75003663 F15-8x50c...
  • Page 9: Heraeus

    / min. radius [cm] 4.3 / 10.4 4.3 / 10.4 4.3 / 10.4 Pitch angle [°] Accel. / braking time [s] 75 / 65 75 / 65 80 / 80 aerosol-tight max temperature for autoclaving °C 75003663 F15-8x50c Thermo Scientific...
  • Page 10 / min. radius [cm] 4.3 / 10.4 4.3 / 10.4 4.3 / 10.4 Pitch angle [°] Accel. / braking time [s] 85 / 85 85 / 85 100 / 80 aerosol-tight max temperature for autoclaving °C Thermo Scientific 75003663 F15-8x50c...
  • Page 11: Sorvall

    / min. radius [cm] 4.3 / 10.4 4.3 / 10.4 4.3 / 10.4 Pitch angle [°] Accel. / braking time [s] 75 / 65 75 / 65 80 /80 aerosol-tight max temperature for autoclaving °C 75003663 F15-8x50c Thermo Scientific...
  • Page 12 / min. radius [cm] 4.3 / 10.4 4.3 / 10.4 4.3 / 10.4 Pitch angle [°] Accel. / braking time [s] 85 / 65 85 /65 100 / 80 aerosol-tight max temperature for autoclaving °C Thermo Scientific 75003663 F15-8x50c...
  • Page 13 / min. radius [cm] 4.3 / 10.4 4.3 / 10.4 4.3 / 10.4 Pitch angle [°] Accel. / braking time [s] 85 / 65 85 / 65 100 / 80 aerosol-tight max temperature for autoclaving °C 75003663 F15-8x50c Thermo Scientific...
  • Page 14 Accessories Contents „Rotor data“ on page 2-2 • • „Accessories“ on page 2-2 Thermo Scientific 75003663 F15-8x50c...
  • Page 15: Chapter 2 Accessories

    Accessories F15-8x50c Carbon Fiber Rotor Thermo Scientific General Purpose Rotors F15-8x50c Fixed Angle Carbon Fiber Rotor Rotor Specifications Capacity (ml) 8 x 50 conical Tube Dimensions (mm) 30 x 121 Angle Maximum Speed (rpm) 14,500 K Factor 1,062 Net Weight (kg)
  • Page 16 AutoLock Contents “Rotor Installation” page 3-2 • “Removing the Rotor” page 3-3 • Thermo Scientific 75003663 F15-8x50c...
  • Page 17: Chapter 3 Autolock Tm

    WARNUNG If the rotor cannot be properly locked in place after several attempts, then the AutoLock is defective and you are not permitted to operate the rotor. Check for any damage to the rotor, damaged rotors must not be used. Keep the hub area clear of objects. 75003663 F15-8x50c Thermo Scientific...
  • Page 18: Removing The Rotor

    2. Grab the rotor handle with both hands and press against the green AutoLock button. At the same time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle. Make sure not to jam the rotor while doing this. Thermo Scientific 75003663 F15-8x50c...
  • Page 19 AutoLock Removing the Rotor 75003663 F15-8x50c Thermo Scientific...
  • Page 20 Rotor Loading Rotor Loading Contents “Before Run” page 4-2 • • “Proper Loading” page 4-2 “Improper Loading” page 4-3 • “Maximum Loading” page 4-4 • Thermo Scientific 75003663 F15-8x50c...
  • Page 21: Chapter 4 Rotor Loading

    4. Check the rotor’s suitability using the chemical compatibility chart on page B-1. 5. Make sure the tubes or bottles do not touch the lid. Proper Loading To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times. 75003663 F15-8x50c Thermo Scientific...
  • Page 22: Improper Loading

    Rotor Loading Improper Loading Improper Loading Thermo Scientific 75003663 F15-8x50c...
  • Page 23: Maximum Loading

    There are two options available for centrifuging samples whose weight, including adaptor, exceeds the maximum permissible load: • Reduce the fill level. • Reduce the speed. Use this formular: Maximum permissible load -------------------------------------------------------------- - Effective load = admissible speed = maximum speed 75003663 F15-8x50c Thermo Scientific...
  • Page 24 Aerosol-tight Applications Contents “Basic Principles” page 5-2 • “Fill Level” page 5-2 • “Checking the Aerosol-Tightness” page 5-2 • Thermo Scientific 75003663 F15-8x50c...
  • Page 25: Chapter 5 Aerosol-Tight Applications

    This releases the carbonic acid gas which is bound in the water, resulting in excess pressure. Do not apply pressure to the lid when doing so. Leaks can be detected by escaping water or the sound of escaping gas. 75003663 F15-8x50c Thermo Scientific...
  • Page 26 Damaged seals are to be replaced immediately. When loading the rotor, ensure that the rotor lid closes securely. Damaged or clouded rotor covers are to be replaced immediately. Thermo Scientific 75003663 F15-8x50c...
  • Page 27 Aerosol-tight Applications Checking the Aerosol-Tightness 75003663 F15-8x50c Thermo Scientific...
  • Page 28: Maintenance And Care

    6-2 • “Handling and Care of Rotors and Accessories made of Aluminum” page 6-3 • “Decontamination” page 6-4 • “Decontamination” page 6-4 • • “Autoclaving” page 6-4 “Service of Thermo Fisher Scientific” page 6-5 • Thermo Scientific 75003663 F15-8x50c...
  • Page 29: Intervals

    Keep in mind the risk of infection when touching articles and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be. 75003663 F15-8x50c Thermo Scientific...
  • Page 30: Handling And Care Of Rotors And Accessories Made Of Aluminum

    • Place the rotors on a plastic grate with their cavities pointing down. • If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures could damage the material and shorten the lifetime of the parts. Thermo Scientific 75003663 F15-8x50c...
  • Page 31: Decontamination

    Clean the rotor before autoclaving and rinse it with distilled water. Remove all accessories (tubes, adaptors) from the rotor. No chemical additives are permitted in the steam. CAUTION Never exceed the permitted temperature and duration when autoclaving. If the rotor shows signs of corrosion or wear, it must be replaced. 75003663 F15-8x50c Thermo Scientific...
  • Page 32: Service Of Thermo Fisher Scientific

    Defective material is replaced. Thermo Fisher Scientific offers inspection and service contracts for this work. This presupposes, however, that all prior service to the centrifuge has exclusively been done by a Thermo Fisher Scientific authorized service technician. Thermo Scientific 75003663 F15-8x50c...
  • Page 33 Maintenance and Care Service of Thermo Fisher Scientific 75003663 F15-8x50c Thermo Scientific...
  • Page 34: Rcf-Values

    10.4 1400 10.4 1500 10.4 1600 10.4 1700 10.4 1800 10.4 1900 10.4 2000 10.4 2100 10.4 2200 10.4 2300 10.4 2400 10.4 2500 10.4 2600 10.4 2700 10.4 2800 10.4 2900 10.4 3000 10.4 1046 Thermo Scientific 75003663 F15-8x50c...
  • Page 35 5900 10.4 1673 4047 6000 10.4 1731 4186 6100 10.4 1789 4326 6200 10.4 1848 4469 6300 10.4 1908 4615 6400 10.4 1969 4763 6500 10.4 2031 4912 6600 10.4 2094 5065 6700 10.4 2158 5219 75003663 F15-8x50c Thermo Scientific...
  • Page 36 9600 10.4 4430 10716 9700 10.4 4523 10940 9800 10.4 4617 11167 9900 10.4 4712 11396 10000 10.4 4807 11627 10100 10.4 4904 11861 10200 10.4 5002 12097 10300 10.4 5100 12335 10400 10.4 5200 12576 Thermo Scientific 75003663 F15-8x50c...
  • Page 37 13300 10.4 8504 20567 13400 10.4 8632 20878 13500 10.4 8761 21191 13600 10.4 8892 21506 13700 10.4 9023 21823 13800 10.4 9155 22143 13900 10.4 9288 22465 14000 10.4 9423 22789 14100 10.4 9558 23116 75003663 F15-8x50c Thermo Scientific...
  • Page 38 RCF-Values Speed (rpm) RCF R RCF R 14200 10.4 9694 23445 14300 10.4 9831 23776 14400 10.4 9969 24110 14500 10.4 10108 24446 Thermo Scientific 75003663 F15-8x50c...
  • Page 39 RCF-Values 75003663 F15-8x50c Thermo Scientific...
  • Page 40: Chemical Compatibility Chart

    S S S S S S M S S S Benzene S S U U S U M U S U U S U U U M U M U U U S U U S U S Thermo Scientific 75003663 F15-8x50c...
  • Page 41 U U S U U U U U U U - U S U - Ethylene Dichloride Ethylene Glycol S S S S S S S S S S S S - S U S S S S S S S S M S M S 75003663 F15-8x50c Thermo Scientific...
  • Page 42 S S S S S S S S S S S S S S S S S S S M S S S Magnesium Chloride Mercaptoacetic Acid U S U - S M S - S M S U U U U - S U U S M S U S S S S Thermo Scientific 75003663 F15-8x50c...
  • Page 43 S S S S S S S S S U M U S S Phosphoric Acid (10%) Phosphoric Acid (conc.) U U M M - U S - M S U U M M S S S M S M S U M U - 75003663 F15-8x50c Thermo Scientific...
  • Page 44 S S U S S S M U S U U U U U U M U M U U U S U M S U S Xylene Zinc Chloride U U S S S S U S S S S S S S S S S S S S S S S U S S S Thermo Scientific 75003663 F15-8x50c...
  • Page 45 Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. Because no organized chemical resistance data exists for materials under the stress of centrifugation, when in doubt we recommend pretesting sample lots. 75003663 F15-8x50c Thermo Scientific...
  • Page 46: Warranty And Recommended Service Life

    Goods, performed by any person or entity other than Thermo Fisher Scien- tific without Thermo Fisher Scientific’s prior written approval, or any use of replacement parts not sup- plied by Thermo Fisher Scientific, shall immediately void and cancel all warranties with respect to the affected Goods. Thermo Scientific 75003663 F15-8x50c...
  • Page 47 Warranty and recommended Service Life 75003663 F15-8x50c Thermo Scientific...
  • Page 48: Contact Information

    +49 6184 90 6940 France +33 2 2803 2180 +33 2 2803 2000 Italy +39 02 02 95059 341 Netherlands +31 76 571 4440 Nordic / Baltic Countries +35 89 329 100 Russia +7 (812) 703 42 15 Thermo Scientific 75003663 F15-8x50c...
  • Page 49 +86 10 8419 3588 India +91 22 6716 2200 Japan +81 45 453 9220 Other Asian Countries +852 2885 4613 Latin America +1 866 984 3766 Other Countries +49 6184 90 6940 +33 2 2803 2180 75003663 F15-8x50c Thermo Scientific...
  • Page 50 Contact Information ........D-1 Sorvall .............1-4 Decontamination ..........6-4 Warranty ............C-1 Disinfection ............ 6-2 Fill Level ............5-2 Heraeus ............1-2 Improper Loading ........... 4-3 Intervals ............6-2 Maintenance ........... 6-1 Precautions ............iii Preface .............. iii Thermo Scientific 75003663 F15-8x50c...