Aerosol-tight Applications ....................5-1 Basic Principles ......................5-2 Fill Level ........................5-2 Checking the Aerosol-Tightness ................5-2 Chapter 6 Maintenance and Care ......................6-1 Intervals ........................6-2 Cleaning ........................6-2 Disinfection ......................6-2 Handling and Care of Rotors and Accessories made of Aluminum ......6-3 Decontamination .....................6-4 Thermo Scientific 75003663 F15-8x50c...
Page 5
Table of Contents Autoclaving ......................6-4 Service of Thermo Fisher Scientific ................6-5 RCF-Values .......................A-1 Chemical Compatibility Chart ................B-1 Warranty and recommended Service Life ............C-1 Contact Information ....................D-1 75003663 F15-8x50c Thermo Scientific...
• Do not remove the magnet at the rotor bottom • Do not use rotors which show any signs of corrosion and/or cracks. • Use only with rotors that have been loaded properly. • Never overload the rotor. Thermo Scientific 75003663 F15-8x50c...
Page 7
ATTENTION means that material damage could occur. WARNING means that injuries or material damage or contamination could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. 75003663 F15-8x50c Thermo Scientific...
WARNUNG If the rotor cannot be properly locked in place after several attempts, then the AutoLock is defective and you are not permitted to operate the rotor. Check for any damage to the rotor, damaged rotors must not be used. Keep the hub area clear of objects. 75003663 F15-8x50c Thermo Scientific...
2. Grab the rotor handle with both hands and press against the green AutoLock button. At the same time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle. Make sure not to jam the rotor while doing this. Thermo Scientific 75003663 F15-8x50c...
Page 19
AutoLock Removing the Rotor 75003663 F15-8x50c Thermo Scientific...
4. Check the rotor’s suitability using the chemical compatibility chart on page B-1. 5. Make sure the tubes or bottles do not touch the lid. Proper Loading To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times. 75003663 F15-8x50c Thermo Scientific...
There are two options available for centrifuging samples whose weight, including adaptor, exceeds the maximum permissible load: • Reduce the fill level. • Reduce the speed. Use this formular: Maximum permissible load -------------------------------------------------------------- - Effective load = admissible speed = maximum speed 75003663 F15-8x50c Thermo Scientific...
This releases the carbonic acid gas which is bound in the water, resulting in excess pressure. Do not apply pressure to the lid when doing so. Leaks can be detected by escaping water or the sound of escaping gas. 75003663 F15-8x50c Thermo Scientific...
Page 26
Damaged seals are to be replaced immediately. When loading the rotor, ensure that the rotor lid closes securely. Damaged or clouded rotor covers are to be replaced immediately. Thermo Scientific 75003663 F15-8x50c...
Keep in mind the risk of infection when touching articles and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be. 75003663 F15-8x50c Thermo Scientific...
• Place the rotors on a plastic grate with their cavities pointing down. • If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures could damage the material and shorten the lifetime of the parts. Thermo Scientific 75003663 F15-8x50c...
Clean the rotor before autoclaving and rinse it with distilled water. Remove all accessories (tubes, adaptors) from the rotor. No chemical additives are permitted in the steam. CAUTION Never exceed the permitted temperature and duration when autoclaving. If the rotor shows signs of corrosion or wear, it must be replaced. 75003663 F15-8x50c Thermo Scientific...
Defective material is replaced. Thermo Fisher Scientific offers inspection and service contracts for this work. This presupposes, however, that all prior service to the centrifuge has exclusively been done by a Thermo Fisher Scientific authorized service technician. Thermo Scientific 75003663 F15-8x50c...
Page 33
Maintenance and Care Service of Thermo Fisher Scientific 75003663 F15-8x50c Thermo Scientific...
S S S S S S M S S S Benzene S S U U S U M U S U U S U U U M U M U U U S U U S U S Thermo Scientific 75003663 F15-8x50c...
Page 41
U U S U U U U U U U - U S U - Ethylene Dichloride Ethylene Glycol S S S S S S S S S S S S - S U S S S S S S S S M S M S 75003663 F15-8x50c Thermo Scientific...
Page 42
S S S S S S S S S S S S S S S S S S S M S S S Magnesium Chloride Mercaptoacetic Acid U S U - S M S - S M S U U U U - S U U S M S U S S S S Thermo Scientific 75003663 F15-8x50c...
Page 43
S S S S S S S S S U M U S S Phosphoric Acid (10%) Phosphoric Acid (conc.) U U M M - U S - M S U U M M S S S M S M S U M U - 75003663 F15-8x50c Thermo Scientific...
Page 44
S S U S S S M U S U U U U U U M U M U U U S U M S U S Xylene Zinc Chloride U U S S S S U S S S S S S S S S S S S S S S S U S S S Thermo Scientific 75003663 F15-8x50c...
Page 45
Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. Because no organized chemical resistance data exists for materials under the stress of centrifugation, when in doubt we recommend pretesting sample lots. 75003663 F15-8x50c Thermo Scientific...
Goods, performed by any person or entity other than Thermo Fisher Scien- tific without Thermo Fisher Scientific’s prior written approval, or any use of replacement parts not sup- plied by Thermo Fisher Scientific, shall immediately void and cancel all warranties with respect to the affected Goods. Thermo Scientific 75003663 F15-8x50c...
Page 47
Warranty and recommended Service Life 75003663 F15-8x50c Thermo Scientific...
Page 49
+86 10 8419 3588 India +91 22 6716 2200 Japan +81 45 453 9220 Other Asian Countries +852 2885 4613 Latin America +1 866 984 3766 Other Countries +49 6184 90 6940 +33 2 2803 2180 75003663 F15-8x50c Thermo Scientific...