GE Moisture Image 1 Series Service Manual
GE Moisture Image 1 Series Service Manual

GE Moisture Image 1 Series Service Manual

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Quick Links

artisantg.com
(217) 352-9330 |
|
Visit our website - Click HERE
Table of Contents
loading

Summary of Contents for GE Moisture Image 1 Series

  • Page 1 [email protected] artisantg.com (217) 352-9330 | Visit our website - Click HERE...
  • Page 2 Moisture Image Series 1 Service Manual...
  • Page 3 June 2003 Process Control Instruments Moisture Image Series 1 Hygrometer Service Manual 910-108SB !ATTENTION! This manual contains instructions for Series 1 units equipped with the latest controller card (p/n 703- 1250). This controller card supports the PanaCom/ PanaView user interface software.
  • Page 4 GE Panametrics. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE Panametrics determines that the equipment was defective, the warranty period is: •...
  • Page 5: Table Of Contents

    Installation Instructions for CE Mark Conformity ....... . .1-2 Precautions for Modified or Non-GE Panametrics Cables ......1-3 Connecting the Recorder Outputs .
  • Page 6 June 2003 Table of Contents (cont.) Chapter 2: Troubleshooting and Maintenance Introduction ..............2-1 Testing the Alarm Relays .
  • Page 7 June 2003 Table of Contents (cont.) Appendix A: Application of the Hygrometer (900-901D1) Introduction ..............A-1 Moisture Monitor Hints .
  • Page 8 Chapter 1...
  • Page 9 Installing Optional Features Making Electrical Connections ......1-1 Connecting the Recorder Outputs ......1-4 Connecting the Alarms .
  • Page 10: Making Electrical Connections

    June 2003 Making Electrical In addition to moisture probes and oxygen cells, the Moisture Image Series 1 can support a variety of optional equipment: recorders, Connections alarms, pressure sensor inputs, auxiliary inputs, PCs and printers. Make all connections to the back panel of the meter. The panel is separated into six sections, one for each channel.
  • Page 11: Installation Instructions For Ce Mark Conformity

    Be sure to use the correct cable type for each probe and make sure that cables are not damaged during installation. If you are not using a GE Panametrics-supplied cable or are using a modified cable, read the following section carefully.
  • Page 12: Precautions For Modified Or Non-Ge Panametrics Cables

    • Carry the shield through any splice. A common mistake is to not connect the shields over the splice. If you are modifying a GE Panametrics cable, the shield will not be accessible without cutting back the cable insulation. Also, do not ground the shield at both ends.
  • Page 13: Connecting The Recorder Outputs

    June 2003 Connecting the The Series 1 has two optically isolated recorder outputs. These outputs provide either a current or voltage signal, which you set using Recorder Outputs switch blocks on the channel cards. Although the meter is configured at the factory, you should check the switch block positions before making connections.
  • Page 14: Replacing The Channel Card

    June 2003 Replacing the Channel 1. Once the switches are set, replace the channel card. Card If you intend to connect pressure inputs or other input devices to the Series 1, do not replace the cover because you will need to set switches on the channel card for those inputs as well.
  • Page 15: Connecting The Alarms

    June 2003 Connecting the Alarms You can order the Series 1 with high and low alarm relays for each channel. Hermetically sealed alarm relays are also available. Each alarm relay is a single-pole double-throw contact set that contains the following contacts (see Figure 1-4): •...
  • Page 16 June 2003 Connecting the Alarms (cont.) ALM A and ALM B STD/TF OXYGEN STD/TF PROBE PROBE Terminal Block CHANNEL 1 CHAN ALM A ALM B ALM A A REC REC B RTN 1 2 +24V RTN 1 Figure 1-5: ALM A and ALM B Terminal Blocks Installing Optional Features...
  • Page 17: Connecting Pressure Sensor Inputs

    Sensor Inputs is connected to the meter differently; therefore it is important to know which type of pressure sensor you are using. IMPORTANT: The transducer must be supplied by GE Panametrics or approved by GE Panametrics for use in this circuit.
  • Page 18: Connecting A Pressure Transducer

    (refer to Figure 1-6). Refer to Table 1-4 for the proper pin connections for the pressure transducer. If you are not using a GE Panametrics-supplied cable, refer to Figure 1-7 to make the proper pin connections to the pressure transducer connector.
  • Page 19 June 2003 Connecting a Pressure Transducer (cont.) STD/TF PROBE Probe Terminal Block Excitation White Pressure Black Transducer Output Green Shield Figure 1-7: Pressure Transducer Cable Assembly 1-10 Installing Optional Features...
  • Page 20: Connecting Pressure Transmitters

    June 2003 Connecting Pressure The Series 1 accepts two types of pressure transmitters: Transmitters Note: Optional auxiliary inputs are required. • Two-wire or loop-powered transmitter (this is always a 4 to 20-mA system) • Four-wire or self-powered transmitter (this can be either a current or voltage output system) Connect the pressure transmitters on the desired channel(s) to the designated pins on the AUX terminal block located on the back of the...
  • Page 21 June 2003 Connecting Pressure A. Connecting the Two-Wire or Loop-Powered Transmitter Use a two-wire non-shielded cable to make connections to the Transmitters (cont.) terminal block labeled AUX on the back of the electronics unit (refer to Figure 1-9). Use Table 1-5 to make the proper pin connections to the desired channel.
  • Page 22 June 2003 Connecting Pressure B. Connecting the Four-Wire or Self-Powered Transmitter Use a four-wire non-shielded cable to make connections to the Transmitters (cont.) terminal block labeled AUX on the back of the electronics unit (refer to Figure 1-9). Use Table 1-6 below to make the proper pin connections to the desired channel.
  • Page 23: Accessing The Channel Cards

    June 2003 Connecting Pressure C. Setting Input Switches Set switch block S1 on each channel card using a pressure transmitter Transmitters (cont.) as described below: Accessing the Channel Cards 1. Remove the screws on the front panel and slide the electronics unit out of its enclosure.
  • Page 24: Replacing The Channel Card

    June 2003 Connecting Pressure Replacing the Channel Card Transmitters (cont.) 1. Once the switches are set, replace the channel card. If you intend to connect pressure inputs or other type of input devices to the Series 1, do not replace the cover because you will need to set switches on the channel card(s) for those inputs as well.
  • Page 25: Connecting Auxiliary Inputs (Optional)

    0/4-20 mA or 0-2 VDC output, Inputs (Optional) including a variety of process control instruments available from GE Panametrics. Inputs may be self- or loop-powered. Self-powered inputs are either current or voltage. Loop-powered inputs are usually current.
  • Page 26 June 2003 Connecting Auxiliary After you make auxiliary input connections, you must set switch block S1 on the Series 1 channel card(s) for the current or voltage Inputs (Optional) input. Use the following steps to make the proper switch settings: (cont.) Accessing the Channel 1.
  • Page 27: Connecting A Personal Computer Or Printer

    June 2003 Connecting a Personal You can connect the Series 1 to a personal computer or serial printer using the RS232 communications port. Refer to the instructions Computer or Printer below to set up and connect your PC or printer. For more information on serial communications, consult the EIA-RS Serial Communications User’s Guide.
  • Page 28: Connecting The Pc Or Printer

    Black Lead (Return) The special GE Panametrics cable has a “MIS/MMS” label on one end of the cable. Connect the “MIS/MMS” end of the cable to the 9- pin connector on the rear of the electronics unit (see Figure 1-15).
  • Page 29: Performing An Mh Calibration Test/Adjustment

    June 2003 Performing an MH If you modify the supplied cables or do not use standard GE Panametrics-supplied cables, you must perform a calibration test/ Calibration Test/ adjustment to test the cable and, if necessary, compensate for any Adjustment error or offset introduced by splicing or long cable lengths. This procedure is also recommended for testing the installation of GE Panametrics cables.
  • Page 30 June 2003 Performing an MH Calibration Procedure (cont.) Calibration Test/ 3. Reprogram the meter with the new (corrected) low reference value Adjustment (cont.) (if required) as described in Entering High and Low Reference Values in Chapter 2 of the Programming Manual. 4.
  • Page 31 Chapter 2...
  • Page 32 Troubleshooting and Maintenance Introduction ......... . .2-1 Testing the Alarm Relays.
  • Page 33: Introduction

    Some of the most common problems and procedures are discussed in this section. If you cannot find the information you need in this section, please consult GE Panametrics. Caution! Do not attempt to troubleshoot the Series 1 beyond the instructions in this section.
  • Page 34: Testing The Alarm Relays

    June 2003 Testing the Alarm The Relay Test Menu enables you to either trip or reset the alarm relays. To enter this menu use Table 2-1. Relays Table 2-1: Moving from Main Menu to Relay Test menu Press the following menu keys: To enter the: SETTINGS* Settings Menu...
  • Page 35: Testing The Recorder Outputs

    June 2003 Testing the Recorder The Recorder Test Menu enables you to test outputs to make sure they are operating properly.You can use this menu to test any Outputs percentage of the full scale recorder range. Use Table 2-2 to enter the Recorder Test Menu.
  • Page 36: Procedure For Testing The Recorder Output

    June 2003 Procedure for Testing Make sure the recorder(s) are connected as described in Connecting the Recorder Outputs on page 1-4, and use the following steps to the Recorder Output perform a test: 1. Use the arrow keys to move the pointer to the channel and recorder you want to test, then press [YES].
  • Page 37: Trimming Recorder Outputs

    June 2003 Trimming Recorder The measured value of the recorder outputs can vary from the programmed value due to load resistance tolerance (e.g. chart Outputs recorder, display, computer interface, etc.). The Recorder Test Menu provides a trimming feature you can use to compensate for any variation in the recorder outputs.
  • Page 38: Procedure For Trimming The Zero Value

    June 2003 Trimming Recorder Outputs (cont.) Series 1 MOISTURE IMAGE SERIES 2 POWER Red- Red+ Chart Recorder Figure 2-4: Checking Series 1 Voltage Outputs Procedure for Trimming 1. Use Table 2-3 below to enter the Recorder Test Menu. The screen appears similar to Figure 2-5.
  • Page 39 June 2003 Procedure for Trimming the Zero Value (cont.) RECORDER OUTPUT TEST Recorder A Recorder B % of Scale NOTE: Recorders SUSPENDED during test. Recorder Test Menu HELP TRIM DONE Figure 2-5: Recorder Test Menu 2. Use the arrow keys to move the pointer to the recorder you want to test and press [YES].
  • Page 40 June 2003 Procedure for Trimming 4. Observe the multimeter reading. Wait at least 5 seconds for the recorder output to settle. The multimeter should display one of the the Zero Value (cont.) following: For the Recorder Output Range Desired Multimeter Reading 0 to 20 mA (5%) 1 mA 4 to 20 mA (0%)
  • Page 41: Procedure For Trimming The Span Value

    June 2003 Procedure for Trimming 1. Use the arrow keys to move the pointer to the recorder you want to test and press [YES]. the Span Value 2. Enter 100 for % of scale and press [YES]. 3. Observe the multimeter reading. Wait at least 5 seconds for the recorder output to settle.
  • Page 42: Screen Messages

    June 2003 Screen Messages The Series 1 has several screen messages that may display during operation. Refer to Table 2-4 for a list of these messages and the possible causes. Table 2-4: Screen Messages and the Possible Causes Screen Message Possible Cause System Response Action...
  • Page 43 Refer to page 3-29 of the 28.) Programming Manual. Over Range The input signal is above the Alarms and recorders Contact GE Panametrics regarding a (See Range Error calibrated range of the respond as programmed. higher calibrated probe. Description on page probe.
  • Page 44: Common Problems

    June 2003 Common Problems If the Series 1 measurement readings seem strange, or they do not make sense, there may be a problem with the probe or a component of the process system. Table 2-5 contains some of the most common measurement problems.
  • Page 45 Clean the sensor and the sensor shield as with non-conductive par- described in Aluminum Oxide Probe ticles (refer to Appendix Maintenance on page A-7 in Appendix A. Then reinstall the sensor. Exception screen Unrecoverable software Contact GE Panametrics. error. Troubleshooting and Maintenance 2-13...
  • Page 46: Checking/ Replenishing The Delta F Oxygen Cell Electrolyte

    Electrolyte check and replenish the electrolyte in the oxygen cell. Note: Some applications require that the electrolyte be changed periodically. Consult GE Panametrics. To check the electrolyte level: Using the min/max window on the oxygen cell, check the electrolyte level. The electrolyte should cover about 60% of the window. The electrolyte level should appear as shown in Figure 2-6.
  • Page 47 June 2003 Checking and To replenish the electrolyte: Once the oxygen cell receives the initial charge of electrolyte, you Replenishing the should monitor the level regularly. DO NOT let the fluid level drop Electrolyte in the below the “MIN” level mark on the window. Delta F Oxygen Cell (cont.) !WARNING!
  • Page 48: Adding Or Removing A Pcmcia Card

    Caution! The Series 1 is not compatible with a flash or ATA card. Please contact GE Panametrics for a list of compatible devices and formatting. Procedure for Adding or 1. Make sure you have a record of the following data, described in...
  • Page 49 June 2003 Procedure for Adding or Removing a PCMCIA Card (cont.) Retainer Bar Controller Board Top View Figure 2-7: Location of the Controller Board 5. Use Figure 2-7 to locate the controller board inside the electronics unit. 6. The controller board will appear similar to Figure 2-8. Insert the PCMCIA card into the brackets along the side of the cutout area.
  • Page 50 June 2003 Procedure for Adding or Removing a PCMCIA Card (cont.) PCMCIA Card Figure 2-8: PCMCIA Card Insertion 2-18 Troubleshooting and Maintenance...
  • Page 51: Installing A Channel Card

    If you decide to order additional channels Card later, or need to replace a channel, GE Panametrics will ship you a channel card. Use the steps below to install a channel card. 1. Turn the instrument off and unplug the main AC power cord.
  • Page 52: Replacing And Recalibrating The Moisture Probes

    Under severe conditions you should send the probes Moisture Probes back for recalibration more frequently; in milder applications you do not need to recalibrate probes as often. Contact a GE Panametrics applications engineer for the recommended calibration frequency for your application.
  • Page 53: Calibrating The Delta F Oxygen Cell

    June 2003 Calibrating the Delta F You should calibrate the Delta F Oxygen Cell when you initially receive it. After that, calibrate the oxygen cell once a month for the Oxygen Cell first three months, and then as needed. You should also calibrate the oxygen cell if you change the electrolyte.
  • Page 54 June 2003 Table 2-6: Measurement Modes and Units for the Series 1 Selected Displayed Measurement Measurement Mode Description of Units Mode Displayed Units Oxygen %= Percent Oxygen default Oxygen Oxygen /ppM = Parts Per Million Oxygen /ppB = Parts Per Billion Oxygen µA /µA = Microamps (Diagnostic Mode)
  • Page 55: Checking The Oxygen Cell Calibration

    To perform this part of calibration you must have a calibration gas with a known PPMv value and a calibration gas inlet valve. Note: GE Panametrics recommends a span calibration gas be 80- 100% of the span of the sensor’s overall range in a background of nitrogen (e.g., 80-100 PPM O...
  • Page 56: Entering The New Span Value

    June 2003 Checking the Oxygen Therefore, Cell Calibration (cont.) ) 290 0.4238 100 75 – – ------------------------------------------------------------- - 75 0.05 – ≅ The new span value (x) is 100 PPMv 387 µA. Enter the new value as described in the next sub-section. Entering the New Span 1.
  • Page 57: Delta F Oxygen Cell Background Gas Correction Factors

    June 2003 Delta F Oxygen Cell The factory calibration procedure for Delta F oxygen cells uses nitrogen as the reference background gas. The Series 1 will measure Background Gas oxygen incorrectly if the transport rate of oxygen through the cell Correction Factors diffusion barrier is different than the cell is calibrated for.
  • Page 58: Entering The Current Multiplier

    June 2003 Correcting for Different When the oxygen cell is used in a background gas other than nitrogen, users must enter the gas’s current multiplier, listed in Table 2-7. The Background Gases Series 1 will apply the appropriate correction to the oxygen signal. (cont.) The original calibration values for nitrogen are programmed into the “Oxygen Probe Calibration”...
  • Page 59 June 2003 Table 2-8: Background Gas Current Multipliers Background Gas Current Multipliers Up to 1000 PPM 5000-10,000 PPM 2.5% to 10% Argon (Ar) 0.97 0.96 0.95 0.98 Hydrogen (H 1.64 1.96 2.38 1.35 Helium (He) 1.72 2.13 2.70 1.39 Methane (CH 1.08 1.09 1.11...
  • Page 60: Range Error Description

    June 2003 Range Error Range Errors occur when an input signal that is within the capacity of the analyzer exceeds the calibration range of the probe. The Series 1 Description Over Range Under Range displays Range Errors with an message.The error condition extends to all displayed measurements Over Range, of that mode.
  • Page 61 Appendix A...
  • Page 62 Application of the Hygrometer (900-901D1) Introduction ......... . A-1 Moisture Monitor Hints .
  • Page 63: Introduction

    June 2003 Introduction This appendix contains general information about moisture monitoring techniques. System contaminants, moisture probe maintenance, process applications and other considerations for ensuring accurate moisture measurements are discussed. The following specific topics are covered: • Moisture Monitor Hints • Contaminants •...
  • Page 64: Moisture Monitor Hints

    One of the major advantages of the GE Panametrics hygrometer is that it can be used for in situ measurements (i.e. the sensor element is designed for installation directly within the region to be measured).
  • Page 65: Pressure

    Response Time The response time of the GE Panametrics standard M Series Aluminum Oxide Moisture Sensor is very rapid - a step change of 63% in moisture concentration will be observed in approximately 5 seconds.
  • Page 66: Temperature

    The converse will happen during the cooler evening hours. Flow Rate GE Panametrics hygrometers are unaffected by the fluid flow rate. The moisture probe is not a mass sensor but responds only to water vapor pressure. The moisture probe will operate accurately under both static and dynamic fluid flow conditions.
  • Page 67: Contaminants

    For convenience, the above particulates have been divided into three broad categories. Refer to the appropriate section for a discussion of their affect on the GE Panametrics moisture probe. Non-Conductive Note: Molecular sieve particles, organic liquid droplets and oil Particulates droplets are typical of this category.
  • Page 68: Conductive Particulates

    June 2003 Conductive Particulates Note: Metallic particles, carbon particles and conductive liquid droplets are typical of this category. Since the hygrometer reading is inversely proportional to the impedance of the sensor, a decrease in sensor impedance will cause an increase in the meter reading. Thus, trapped conductive particles across the sensor leads or on the sensor surface, which will decrease the sensor impedance, will cause an erroneously high dew point reading.
  • Page 69: Aluminum Oxide Probe Maintenance

    June 2003 Aluminum Oxide Probe Other than periodic calibration checks, little or no routine moisture probe maintenance is required. However, as discussed in the previous Maintenance section, any electrically conductive contaminant trapped on the aluminum oxide sensor will cause inaccurate moisture measurements. If such a situation develops, return of the moisture probe to the factory for analysis and recalibration is recommended.
  • Page 70 Probes which are not in proper calibration must be recalibrated. It is recommended that all moisture probes be recalibrated by GE Panametrics approximately once a year, regardless of the probe’s condition. Application of the Hygrometer (900-901D1)
  • Page 71: Corrosive Gases And Liquids

    June 2003 Corrosive Gases And GE Panametrics M Series Aluminum Oxide Moisture Sensors have been designed to minimize the affect of corrosive gases and liquids. Liquids As indicated in the Materials of Construction section of this appendix, no copper, solder or epoxy is used in the construction of these sensors.
  • Page 72: Materials Of Construction

    June 2003 Materials of Construction M1 and M2 Sensors: Sensor Element: 99.99% aluminum, aluminum oxide, gold, Nichrome, A6 Back Wire: 316 stainless steel Contact Wire: gold, 304 stainless steel Front Wire: 316 stainless steel Support: Glass (Corning 9010) Electrical Connector: Pins: Al 152 Alloy (52% Ni) Glass:...
  • Page 73: Calculations And Useful Formulas In Gas Applications

    0°C. GE Panametrics moisture probes are never calibrated with supercooled water. Caution is advised when comparing dew points measured with a GE Panametrics hygrometer to those measured with a mirror type hygrometer, since such instruments may provide the dew points of supercooled water.
  • Page 74: Parts Per Million By Volume

    To compute the moisture content for any ideal gas at a given pressure, refer to Figure A-1 on page A-20. Using a straightedge, connect the dew point (as measured with the GE Panametrics’ Hygrometer) with the known system pressure. Read the moisture content in PPM where the straightedge crosses the moisture content scale.
  • Page 75: Parts Per Million By Weight

    June 2003 Parts per Million by The water concentration in the gas phase of a system, in parts per million by weight, can be calculated directly from the PPM and the Weight ratio of the molecular weight of water to that of the carrier gas as follows: ×...
  • Page 76: Weight Of Water Per Unit Weight Of Carrier Gas

    June 2003 Weight of Water per Unit Occasionally, the moisture content of a gas is expressed in terms of the weight of water per unit weight of carrier gas. In such a case, the Weight of Carrier Gas unit of measure defined by the following equation is the most commonly used: ×...
  • Page 77 June 2003 Table A-1: Vapor Pressure of Water Note: If the dew/frost point is known, the table will yield the partial water vapor pressure ) in mm of Hg. If the ambient or actual gas temperature is known, the table will yield the saturated water vapor pressure (P ) in mm of Hg.
  • Page 78 June 2003 Table A-1: Vapor Pressure of Water (Continued) Aqueous Vapor Pressure Over Water Temp. (°C) 4.579 4.647 4.715 4.785 4.855 4.926 4.998 5.070 5.144 5.219 5.294 5.370 5.447 5.525 5.605 5.685 5.766 5.848 5.931 6.015 6.101 6.187 6.274 6.363 6.453 6.543 6.635...
  • Page 79 June 2003 Table A-1: Vapor Pressure of Water (Continued) Aqueous Vapor Pressure Over Water (cont.) Temp. (°C) 61.500 62.140 62.800 63.460 64.120 64.800 65.480 66.160 66.860 67.560 68.260 68.970 69.690 70.410 71.140 71.880 72.620 73.360 74.120 74.880 75.650 76.430 77.210 78.000 78.800 79.600...
  • Page 80 June 2003 Table A-1: Vapor Pressure of Water (Continued) Aqueous Vapor Pressure Over Water (cont.) Temp. (°C) 416.80 420.20 423.60 426.80 430.20 433.60 437.00 440.40 444.00 447.50 450.90 454.40 458.00 461.60 465.20 468.70 472.40 476.00 479.80 483.40 487.10 491.00 494.70 498.50 502.20 506.10...
  • Page 81 June 2003 Table A-2: Maximum Gas Flow Rates Based on the physical characteristics of air at a temperature of 77°F and a pressure of 1 atm, the following flow rates will produce the maximum allowable gas stream linear velocity of 10,000 cm/sec in the corresponding pipe sizes.
  • Page 82 June 2003 10,000 1,000 8,000 10,000 6,000 8,000 5,000 6,000 4,000 5,000 3,000 4,000 3,000 2,000 2,000 1,500 1,000 1,000 10.0 -100 -110 -120 -130 Figure A-1: Moisture Content Nomograph for Gases A-20 Application of the Hygrometer (900-901D1)
  • Page 83: Comparison Of Ppmv Calculations

    • the calculations described in this appendix • calculations performed with the slide rule device that is provided with each GE Panametrics hygrometer • values determined from tabulated vapor pressures For comparison purposes, examples of all three procedures are listed in Table A-4 below.
  • Page 84: Liquid Applications

    As discussed above, a GE Panametrics aluminum oxide sensor can be directly immersed in a hydrocarbon liquid to measure the equivalent dew point. Since the dew point is functionally related to the vapor...
  • Page 85 • the temperature of the liquid at the time of measurement • the saturation water concentration at the measurement temperature • the dew point, as measured with the GE Panametrics hygrometer Application of the Hygrometer (900-901D1) A-23...
  • Page 86 Note: The Karl Fischer analysis involves titrating the test sample against a special Karl Fischer reagent until an endpoint is reached. 2. Measure the dew point of the known sample with the GE Panametrics hygrometer. 3. Measure the temperature (°C) of the test solution.
  • Page 87 Typical Problems 1. Find the moisture content in benzene, at an ambient temperature of 30°C, if a dew point of 0°C is measured with the GE Panametrics hygrometer. a. From the literature, it is found that C for benzene at a temperature of 30°C is 870 PPM...
  • Page 88 K is constant over a large temperature range. 3. Find the moisture content in hexane, at an ambient temperature of 10°C, if a dew point of 0°C is measured with the GE Panametrics hygrometer. a. From the literature, it is found that C for hexane at a temperature of 20°C is 101 PPM...
  • Page 89 Either the dew point of the liquid or its percent saturation is the only value needed. For such applications, the saturation value for the liquid need not be known. The GE Panametrics hygrometer can be used directly to determine the dew...
  • Page 90: Empirical Calibrations

    June 2003 Empirical Calibrations For those liquids in which a Henry’s Law type analysis is not applicable, the absolute moisture content is best determined by empirical calibration. A Henry’s Law type analysis is generally not applicable for the following classes of liquids: •...
  • Page 91 June 2003 A. Instructions for Karl To perform a Karl Fisher analysis, use the apparatus in Figure A-3 on page A-33 and complete the following steps: Fischer Analysis 1. Fill the glass bottle completely with the sample liquid. 2. Close both valves and turn on the magnetic stirrer. 3.
  • Page 92 June 2003 B. Instructions for Note: This procedure is only for liquids that are highly miscible with water. Excessive equilibrium times would be required with Preparing Known less miscible liquids. Samples To prepare samples of known moisture content, use the apparatus in Figure A-3 on page A-33 and complete the following steps: 1.
  • Page 93 June 2003 C. Additional Notes for In addition to the topics already discussed, the following general application notes pertain to the use of GE Panametrics moisture Liquid Applications probes in liquid applications: 1. All M Series Aluminum Oxide Moisture Sensors can be used in either the gas phase or the liquid phase.
  • Page 94 June 2003 Figure A-2: Moisture Content Nomograph for Liquids A-32 Application of the Hygrometer (900-901D1)
  • Page 95 June 2003 Stainless Steel Tubing (soft soldered to cover) 3/4-26 THD Female (soft soldered to cover) M2 Probe Rubber Septum Exhaust Soft Solder Metal Cover with Teflon Washer Liquid Glass Bottle Magnetic Stirrer Bar Magnetic Stirrer Figure A-3: Moisture Content Test Apparatus Application of the Hygrometer (900-901D1) A-33...
  • Page 96: Solids Applications

    June 2003 Solids Applications A. In-Line GE Panametrics moisture probes may be installed in-line to continuously monitor the drying process of a solid. Install one sensor Measurements at the process system inlet to monitor the moisture content of the drying gas and install a second sensor at the process system outlet to monitor the moisture content of the discharged gas.
  • Page 97 If in-line measurements are not practical, then there are two possible laboratory procedures: Procedures 1. The unique ability of the GE Panametrics sensor to determine the moisture content of a liquid can be used as follows: a. Using the apparatus shown in Figure A-3 on page A-33, dissolve a known amount of the solids sample in a suitable hydrocarbon liquid.
  • Page 98 June 2003 Index Alarms Electrical Connections ....1-1, 1-2 Connecting ......1-6 Alarms.
  • Page 99 June 2003 Index (cont.) M Series Probes PCMCIA Card Precautions for Modified Cables ..1-3 Replacing ......2-16 Maintenance Personal Computer Channel Card, Installing .
  • Page 100 June 2003 Index (cont.) Screen Messages ..... 2-10 Settings Menu Entering Recorder Test Menu ..2-3 Entering Relay Test Menu .
  • Page 101 ATEX COMPLIANCE GE Panametrics 221 Crescent Street, Suite 1 Waltham, MA 02453 U.S.A. as the manufacturer, declare under our sole responsibility that the product Moisture Image Series 1 Analyzer to which this document relates, in accordance with the provisions of ATEX Directive 94/9/EC Annex II, meets the following specifications: II (1) G [EEx ia] IIC (-20°C to +50°C)
  • Page 102 DECLARATION CONFORMITY GE Panametrics Shannon Industrial Estate Shannon, Co. Clare Ireland declare under our sole responsibility that the Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer Moisture Monitor Series 35 Analyzer to which this declaration relates, are in conformity with the following standards: •...
  • Page 103 DECLARATION CONFORMITE GE Panametrics Nous, Shannon Industrial Estate Shannon, Co. Clare Ireland déclarons sous notre propre responsabilité que les Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer Moisture Monitor Series 35 Analyzer rélatif á cette déclaration, sont en conformité avec les documents suivants: •...
  • Page 104 KONFORMITÄTS- ERKLÄRUNG GE Panametrics Wir, Shannon Industrial Estate Shannon, Co. Clare Ireland erklären, in alleiniger Verantwortung, daß die Produkte Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer Moisture Monitor Series 35 Analyzer folgende Normen erfüllen: •...
  • Page 105 WORLDWIDE OFFICES MAIN OFFICES: GE PANAMETRICS INTERNATIONAL OFFICES: Australia Japan GE Panametrics P.O. Box 234 2F, Sumitomo Bldg. 221 Crescent St., Suite 1 Gymea N.S.W. 2227 5-41-10, Koishikawa, Bunkyo-Ku Waltham, MA 02453-3497 Australia Tokyo 112-0002 Telephone 61 (02) 9525 4055...
  • Page 106 GE Panametrics 221 Crescent Street, Suite 1 Waltham, MA 02453-3497 Telephone: (781) 899-2719 Toll-free: (800) 833-9438 Fax: (781) 894-8582 E-Mail: [email protected] http://www.panametrics.com Ireland GE Panametrics Shannon Industrial Estate Shannon, County Clare Ireland Telephone: 353-61-470200 Fax: 353-61-471359 E-Mail: [email protected]...

Table of Contents