Probe Usage; Care And Maintenance; Inspecting Probes; Environmental Requirements - GE VOLUSON 730Pro Basic User Manual

Hide thumbs Also See for VOLUSON 730Pro:
Table of Contents
Probes and Biopsy
20.8

Probe Usage

For details on connecting, activating, deactivating, disconnecting, transporting and storing the probes,
refer to:
Transducer Connection
20.9

Care and Maintenance

20.9.1

Inspecting Probes

After each use, inspect the probe's lens, cable, and casing. Look for any damage that would allow
liquid to enter the probe. If any damage is found, do not use the probe until it has been inspected and
repaired/replaced by a GE Medical Systems - Kretztechnik Service Representative.
NOTE:
Keep a log of all probe maintenance, along with a picture of any probe malfunction.
20.9.2

Environmental Requirements

Probes should be operated, stored, or transported within the parameters outlined below.
Ensure that the probe face temperature does not exceed the normal
operation temperature range.
Probe Environmental Requirements
Operational
+10° to +40°  C
Temperature
+50° to +104° F
Humidity
non-condensing
Pressure
700 - 1060hPa
20-8
(chapter 4.4)
and
Storage
-10° to +60°  C
+14° to +140° F
Max. 90%
Max. 90%
non-condensing
700 - 1060hPa
Probe/Program Selection
Transport
-40° to  60°  C
-40° to +140° F
Max 90%
non-condensing
700 - 1060hPa
®
Voluson
730Pro - Basic User Manual
(chapter 4.5).
105831 Rev. 0