Field replaceable units: Replace main unit parts
b. Remove the wireless card from the PC card bracket by pressing
the CF card eject button. To press the CF card eject button,
carefully insert the flat end of a pen approximately 1- inch deep
and push the CF flash card eject button as shown.
c.
Carefully disconnect the antennas using a pen tip, small
screwdriver, or needle nose pliers.
d. Place the wireless card in a safe place.
4. Remove the DAS assembly according to steps starting on page 7-37.