Operating Modes - GE Corometrics 170 Series Operator's Manual

Fetal monitor
Hide thumbs Also See for Corometrics 170 Series:
Table of Contents

Operating Modes

FECG Mode
Technique:
Heart Rate Counting Range:
Heart Rate Resolution:
Artifact Elimination:
Countable Input Signal Range:
Offset Voltage Tolerance (Differential):
Maximum Common Mode Voltage:
Common Mode Rejection:
Balanced:
Unbalanced 5kΩ RA or LA:
Input Impedance:
Differential:
Common Mode:
Mains Frequency Rejection:
Leakage Current:
Isolation, Mains-to-Patient:
Ultrasound Mode
Technique:
Transducer Type:
Pulse Repetition Frequency:
Pulse Duration:
Transmitter Frequency:
Spatial-Average Temporal Average
Intensity:
Focal 20 dB Beam Area:
Peak Instantaneous Intensity:
Heart Rate Counting Range:
Leakage Current:
a
Isolation
, Mains-to-Patient:
Uterine Activity Mode
Range:
Resolution:
Bandwidth:
Excitation Voltage:
Zero Set Temperature Drift:
Leakage Current:
b
Isolation
, Mains-to-Patient:
a
The ultrasound connector is isolated for the primary ultrasound channel (combi-connector) on Model 174 Monitors only
b
The uterine activity connector is isolated on Model 173 and 174 Monitors only (units equipped with IUP).
Revision D
Technical Specifications: Operating Modes
Table 11-2. Operating Mode Specifications
Peak detecting, beat-to-beat cardiotachometer
30–240 BPM
1 BPM
Service selectable, ±25 BPM artifact rejection
15 µV to 2 mV peak-to-peak
±300 mVdc maximum
20 V peak-to-peak
> 120 dB at mains frequency, with patient cable
> 110 dB at mains frequency
> 10 MΩ
> 20 MΩ
> 40 dB
Complies with EN60601.1 and/or EN60601.1.1 harmonized national standard
> 5656 Vdc
Pulsed Doppler with autocorrelation processing
9-crystal
2 kHz (all modes)
92 µs
1.151 MHz
2
Isata< 5 mW/cm
2
16.6 cm
, at a range = 7 cm
2
1.8 mW/cm
50–210 BPM
Complies with EN60601.1 and/or EN60601.1.1 harmonized national standard
a
> 5656 Vdc
Strain Gauge
0–100 mmHg
1 mmHg
dc to 3 Hz
+4.0 Vdc
< 0.1 mmHg/°C (0.013 kPa/°C), excluding transducer
Complies with EN60601.1 and/or EN60601.1.1 harmonized national standard
b
> 5656 Vdc
170 Series Monitor
2003023-001
Tocotransducer
0–100 relative units
1 relative unit
dc to 0.5 Hz
+4.0 Vdc
.
11-3

Hide quick links:

Table of Contents
loading

Table of Contents