Replacing boards / replacing components
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Parameters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
Hospital Name
Storage of Dose on Disk
Dose on hardcopy output
Perform the IQ quick test.
Siemens AG
Medical Solutions
Change the programmed parameters only upon cus-
tomer request.
Configure to monitor A.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Configure horizontally.
Configure vertically only upon customer request.
Configure Yes.
Set the current date and time.
Program the hospital.
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.