Checking the operating voltages
Low voltages
Check the operating voltages for the SIREMOBIL Compact against the values in Table 1.
from test point
M14.4 (0V)
M14.3 (0V)
M14.3 (0V)
D3.X8.1 (0V)
M13.S- (0V)
D3.X1.7 (0V)
Tab. 1
I.I. voltages
WARNING
The voltages E1 / E2 / E3 and A may be taken from the I.I. test protocol and checked or
adjusted against the control test points listed in Table 2.
Voltage
E1
E2
E3
30 kV, Anode
Tab. 2
Siemens AG
Medical Solutions
to test point
M14.5/6 (+5 V)
M14.2 (+15 V)
M14.1 (-15 V)
D3.X8.3 (+26,75 V)
M13.S+ (+13 V)
D3.X1.9 (+24 V)
Electrical voltage!
See Chapter 1, Safety Information.
Do not touch any part of the I.I. mini voltage supply!
Test point
Ground
point
(0V)
UE1
UE2
UI 15
UI 30
SPR2-130.061.02
Rev. 11
Voltage
Tolerance
+ 5 V
+ 0.2 V
+ 15 V
+ 0.1 V
- 15 V
+ 0.5 V
+26.75 V
+ 0.25 V
+13 V
+ 0.2 V
+ 26.5 V to
n.a.
+ 29.5 V
Potentiome-
ter for full
format
P10
P6
P2
P1
Page 1 of 2
09.04
CS PS 24
3 - 1
3
Potentiometer
M14. +5 V/Adj.
M14. +15 V/Adj.
M14. -15 V/Adj.
D3.R3
M13.TR1
not adjustable
Potentio-
Voltage
meter
divider
for zoom
ratio
format
P11
1:1
P7
1:1
P3
1:10000
P1
1:10000
SIREMOBIL Compact
System Manual
3
3