Siemens SIREMOBIL Compact L Service Instructions Manual page 73

Table of Contents
Replacing boards / replacing components
Monitor text
Maximum image number
Disable dose rate selection
in user setup
Unit serial number
Default LUT
Default Edge
Software backward compatible Configure No.
Product name
Default Program changeable
Create External disk
Enable Bridge
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Parameters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
Siemens AG
Medical Solutions
Configure Yes.
Configure 700.
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on Disk" and
menu "Dose on Hardcopy output" in User Setup are vis-
ible. Otherwise, set to "Yes".
Program the serial number of the SIREMOBIL Compact,
Do not enter the first zero.
Configure 2. Only upon customers request program a
different LUT curve.
Configure 1. Only upon customers request program a
different. Edge enhancement value
SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default: Yes.
Depending on country specific regulations, configure
Yes or No.
If a MOD-drive is present or the option "DICOM bridge"
is installed, configure Yes. Without MOD drive or
DICOM bridge, configure No.
If the option "DICOM bridge" is installed, configure Yes.
Otherwise configure No.
Change the programmed parameters only upon cus-
tomer request.
Configure to monitor A.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Configure horizontally. Configure vertically only upon
customer request.
Configure Yes.
Set the current date and time.
SPR2-130.061.02
Rev. 11
09.04
Page 21 of 28
SIREMOBIL Compact
CS PS 24
6 - 21
System Manual