Interfaces To Peripherals - GE marquette MAC 1100 Servicing Instructions

Table of Contents
Marquette Hellige GmbH
MO_SU/
Pin
Signal
Number
1
+UVERS
2
EN_MOSU
3
GND24V
Connector denotation:
Type:
The function of the individual pins is given in the following table . The definition as an
input/output is seen with reference to PCB Control CS_CI.
RS232_E/
Pin
Signal
Number
1
+5V
2
+5V
3
GND
4
GND
5
SMRXD1
6
SMTXD1
7
GND
8
V MODEM
9
EKGTRIG E
10
FERNST E
2.6

Interfaces to peripherals

The MAC1200 has only 3 interfaces for peripherals:
Mains input
Patient input
RS-232 interface
The mains input interface on the device is a 3-pin standard cold appliance socket
connection, which is integrated into the mains input module. Connection to the mains
is effected via a 3-pin power cord with a non-fused earth conductor.
The mains input is a wide range input from 95Vac ... 240Vac. The mains input
module contains the two primary fuses.
MAC 1100/1200 V 1.1
227 492 20 D - 0002
input/output
Function
Output
Supply Voltage + 26,5V
Output
Enable Modem Supply
Output
Ground
RS232_E/
male connector, 2x 5-pin, 180 °
input/output
Function
Output
Logic Supply
Output
Logic Supply
Ground
Ground
Input
Receive (RXD)
Output
Transmit (TXD)
Ground
Input
Modem Supply +8V
Output
ECG Trigger
Input
Remote Start Input
Page 25
Definition
Open, high: enable
Low:
disable
Definition
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Table of Contents