GE Vivid T8 Basic Service Manual page 258

Hide thumbs Also See for Vivid T8:
Table of Contents
Replacement Procedures
8-4-8-1 Loading the System Software with USB memory stick(continued)
NOTE:
NOTE:
NOTE:
WARNING
WARNING
8-30
To select [c], only upgrade Application Software. This
option will only upgrade the application software and will
not backup the system.
To load the system software with USB memory stick, it
will take about twenty minutes to complete the
procedure.
To load the system software by software download, it
takes about ten hours to complete the procedure. The
software downloaded is about 6 GB. The actual time
required for software download varies with the network
condition.
To load the system software with software reload, it will
take about twenty minutes to complete the procedure.
To select [d], erase the complete disk. This option will
destroy all patient data.
Selection of [d] DOES NOT meet the GE Global Secure Wipe
Standard. Back up the data before executing.
While the software install procedure is designed to
preserve data, you should select choice [b] to format disk
C only.
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10