GE TONOPORT V Service Manual
Hide thumbs Also See for TONOPORT V:
TONOPORT V
Service Manual
Firmware Versions FW1.4 and FW2.0
Hardware Versions HW1.2 and HW2.0
2001589-023
Revision H
Table of Contents
loading

Summary of Contents for GE TONOPORT V

  • Page 1 TONOPORT V Service Manual Firmware Versions FW1.4 and FW2.0 Hardware Versions HW1.2 and HW2.0 2001589-023 Revision H...
  • Page 2 NOTE: The information in this manual only applies to the product TONOPORT V, Firmware Versions FW1.4 and FW2.0 and Hardware Versions HW1.2 and HW2.0. Due to continuing product innovation, specifications in this manual are subject to change without notice. © 2006 General Electric Company. All rights reserved...
  • Page 3: Table Of Contents

    Firmware Loading Procedure for FW2.0 ....8-1 Settings for TONOPORT V Update Program V2.1 for Windows ..8-1 Loading Procedure ..........8-2...
  • Page 4 Exchange Units ..........13-3 Appendix: Drawings ......14-1 TONOPORT V Revision H...
  • Page 5: Introduction

    See the operator manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use. 1.1.3 Intended Audience This manual is intended for the person who uses, maintains, or troubleshoots this equipment. Revision H TONOPORT V 2001589-023...
  • Page 6: Safety Information

    Introduction: Safety Information Safety Information 1.2.1 Responsibility GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if: Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE Medical Systems Information Technologies.
  • Page 7: Service Information

    Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents. Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
  • Page 8 Introduction: Service Information For your notes TONOPORT V Revision H 2001589-023...
  • Page 9: Block Diagram

    Block Diagram Tubing system Revision H TONOPORT V 2001589-023...
  • Page 10 For your notes TONOPORT V Revision H 2001589-023...
  • Page 11: Functional Description

    AAAA accu Info bbbb battery Info Start 1111 signal tone ON (TONOPORT V, V2.0 only) Info 0000 signal tone OFF (TONOPORT V, V2.0 only) Measurement limits - Max. cuff inflation time = 100 seconds - Max. measurement time after start of measurement = 180 seconds...
  • Page 12 For your notes TONOPORT V Revision H 2001589-023...
  • Page 13: Pin Connections

    Parity Cable USB TONOPORT V - PC (2001589-040) - not shown This cable connects the USB port of TONOPORT V, V2.0 (USB B-mini, male, 5-pin) to the USB port (USB A, male) of the PC. Cable TONOPORT V - Printer...
  • Page 14 For your notes TONOPORT V Revision H 2001589-023...
  • Page 15: Instrument Versions

    TONOPORT V, V2.0 Displaying the firmware version and the language for the direct printout Switch TONOPORT V on and wait until the time appears on the screen Press INFO five times: Display shows "H 5". Press START: the firmware version and the language for the direct printout are now displayed.
  • Page 16 14 : 02 English 14 : 03 French 14 : 04 Italian 14 : 05 Spanish 20 : 01 German 20 : 02 English 20 : 03 French 20 : 04 Italian 20 : 05 Spanish TONOPORT V Revision H 2001589-023...
  • Page 17: Pictures

    Pictures Pictures of TONOPORT V up to version HW1.4 Revision H TONOPORT V 2001589-023...
  • Page 18 Pictures: Pictures of TONOPORT V for version HW2.0 TONOPORT V Revision H 2001589-023...
  • Page 19: Firmware Loading Procedure For Fw1.4

    Firmware Loading Procedure for FW1.4 NOTE To update older TONOPORT V units (HW1.2 and before), use firmware version FW1.4. Preparation 7.1.1 Settings for Hyperterminal under Windows 95/98SE/NT4.0/ 2000 (in this example with Windows 95SE) Start Hypertrm.exe. Type in the name: TONOPORT V.
  • Page 20 Firmware Loading Procedure for FW1.4: Preparation TONOPORT V Revision H 2001589-023...
  • Page 21 7.1.2 Settings for Telix for Windows (Hyperterminal for NT does not work) Extract all TONOPORT V, FW1.4 Firmware files into one directory (e.g. C:\tonoport\V1.4) by starting: FIRMWARE TONOPORT V FW14.exe Load all files from the Telix program into the same directory on the Hard Disk Start the Telix program (\telix.exe)
  • Page 22: Loading Procedure

    TONOPORT V-FW1.4/DEUTSCH HHLMM (time HAUPT14.PR 1. Steps 2 and 3 are not applicable if V 1.3 or 1.4 is installed on TONOPORT V. 2. Note down the "LANGUAGE" (DEUTSCH, ENGLISH, FRANCAIS, ITALIANO, ESPAÑOL) which is installed on TONOPORT V (Step1)
  • Page 23 Firmware Loading Procedure for FW1.4: Loading Procedure Switch off TONOPORT V. Disconnect TONOPORT V from PC. Check the correct Firmware version on TONOPORT V and set the correct language: Step Action Diplay Switch on TONOPORT HH:MM (time) Press “INFO” 5 times Press “START”...
  • Page 24 Firmware Loading Procedure for FW1.4: Loading Procedure For your notes TONOPORT V Revision H 2001589-023...
  • Page 25: Firmware Loading Procedure For Fw2.0

    Firmware Loading Procedure for FW2.0 NOTE To update TONOPORT V HW2.0, use only firmware version 2.0 or higher. Settings for TONOPORT V Update Program V2.1 for Windows Extract all TONOPORT V, FW2.0 Firmware files into one directory (e.g. C:\Tonoport\V2.0) by starting: TONOPORT_V_Update_FW20.ZIP.
  • Page 26: Loading Procedure

    NOTE After TONOPORT V has been switched on, updating can only be carried out within the first 10 minutes. After 10 minutes TONOPORT V enters a power down mode and can be reactivated by switching on and off. Click the “Start Update” button and follow the displayed prompts.
  • Page 27 Firmware Loading Procedure for FW2.0: Loading Procedure Screenshot of a successful loading procedure: Revision H TONOPORT V 2001589-023...
  • Page 28 Firmware Loading Procedure for FW2.0: Loading Procedure For your notes TONOPORT V Revision H 2001589-023...
  • Page 29: Diagnosis Codes / Error Codes

    Difference between systolic and diastolic pressure too small (10 mmHg or less) E 22 Motion artifact during systole detection. E 24 No systole detected in the provided time frame. E 26 Systolic pressure below 60 mmHg. Revision H TONOPORT V 2001589-023...
  • Page 30 At the same time the device switches to a deflation rate of 4 mmHg/s. When it detects more than 13 oscillations later on, the rate is changed to 6 mmHg/s. TONOPORT V Revision H 2001589-023...
  • Page 31: Adjustment Instructions

    Only the pressure amplifier needs to be calibrated. This is done by switching on the TONOPORT V and waiting until the time is displayed on the LCD. Measure the direct voltages at TP2 and TP3. This is done by turning the trimming potentiometer R59 with a screwdriver until the voltages at TP2 is 1.8...
  • Page 32: Interior View

    Adjustment Instructions: Interior View 10.2 Interior View TONOPORT V up to Version HW1.4 10-2 TONOPORT V Revision H 2001589-023...
  • Page 33 Adjustment Instructions: Interior View TONOPORT V, V2.0 Revision H TONOPORT V 10-3 2001589-023...
  • Page 34 Adjustment Instructions: Interior View For your notes 10-4 TONOPORT V Revision H 2001589-023...
  • Page 35: Technical Specifications

    (-400 m relative to sea level) Transport and storage Temperature -10...+ 70 °C Relative humidity 10...90 %, no condensation Atmospheric pressure 500 hPa (altitude +4500 m relative to sea level) to 1060 hPa (-400 m relative to sea level) Revision H TONOPORT V 11-1 2001589-023...
  • Page 36 Technical Specifications: Dimensions Height 27 mm Width 80 mm Depth 100 mm Weight 199 g with batteries 11-2 TONOPORT V Revision H 2001589-023...
  • Page 37: Technical Inspections

    Press INFO four times: display indicates "H 4". Press START: display indicates internal value, which must lie between 30 and 70. If the value lies outside this range, the TONOPORT V must be sent in for inspection. Press START again: display indicates "0" (the device is now in the calibration mode and the display indicates the pressure in mmHg).
  • Page 38: Technical Inspections Of The Measuring System

    Press INFO four times: display indicates "H 4". Press START: display indicates internal value, which must lie between 30 and 70. If the value lies outside this range, the TONOPORT V must be sent in for inspection. Press START again: display indicates "0" (the device is now in the calibration mode and the display indicates the pressure in mmHg).
  • Page 39: Parts Lists

    Operator’s Manual TONOPORT V, V2.0 RUS/POL/CZE/HUN/SLO/CHS 13.2 Accessories Part No. Description 2001589-011 Cable RS232 TONOPORT V ———> PC (with CardioSoft V4.14 or higher) 2001589-012 Cable RS232 TONOPORT V ———> Printer (LQ-Emulation) 2001589-040 Cable USB TONOPORT V ———> PC (approx. 1.5 m)
  • Page 40: Spare Parts

    PCB HW1.2. That means if you have a TONOPORT V with lower serial numbers than in the list, then upgrade to devices with the low noise pump and with the PCB HW1.2.
  • Page 41: Exchange Units

    Parts Lists: Exchange Units 13.4 Exchange Units Part No. Description 2003386-001 Exchange Recorder ABP TONOPORT V German 2003386-002 Exchange Recorder ABP TONOPORT V, V2.0 Revision H TONOPORT V 13-3 2001589-023...
  • Page 42 Parts Lists: Exchange Units For your notes 13-4 TONOPORT V Revision H 2001589-023...
  • Page 43: Appendix: Drawings

    Appendix: Drawings Revision H TONOPORT V 14-1 2001589-023...
  • Page 44 For your notes 14-2 TONOPORT V Revision H 2001589-023...
  • Page 52 Asia Headquarters European Representative GE Medical Systems GE Medical Systems GE Medical Systems Information Technologies, Inc. Information Technologies Asia; GE (China)Co., Ltd. Information Technologies GmbH 8200 West Tower Avenue Munzinger Straße 3 24th Floor Shanghai MAXDO Centre, Milwaukee, WI 53223 USA...

Table of Contents