Troubleshooting In The Gas Supply; Troubleshooting In The Electrical Supply And Insulation - Honeywell ENCAL 3000 Hardware Manual

Gas chromatograph
Table of Contents
8.1.1

Troubleshooting in the gas supply

The following action can be taken if a fault occurs in the gas supply, (rectification of
the faults under items 1 – 4), e.g. abnormal calibration gas or carrier gas shrinkage.
Step
Step
Step
Step
1.
2.
3.
4.
5.
8.1.2

Troubleshooting in the electrical supply and insulation

You can take the following action if an error occurs in the electrical power supply
(rectification of the faults under items 5 and 6). Use a voltmeter to localize the error.
If the device is not supplied with electricity or the supply voltage/current is too low, you
cannot establish a connection.
.
EnCal 3000 Hardware Manual
Action:
Action: Gas supply fault rectification
Gas supply fault rectification
Action:
Action:
Gas supply fault rectification
Gas supply fault rectification
Complete a tightness test by checking the entire gas route (from the
cylinder/pipeline to the device) using a gas detector.
Seal any leaks properly.
Set the regulators and valves to the required values.
Disconnect blocked lines from the switched-off device and flush the pipeline with
non-flammable gas to remove any blockages and deposits while observing the
maximum admissible pressure.
Discharge the flushing gas properly.
Use suitable gas detectors for the tightness test. These must reliably
display even very small quantities of the possibly leaking process,
ATTENTION
calibration and carrier gases.
The connection plugs on the edge of the terminal board must always be
inserted until they lock. Check that they are secure.
CAUTION
Tighten the screws on any unused terminals.
ATTENTION
24 V DC Fluctuations, including mains fluctuations ± 15%, for example
with Quint-PS-100-240AC / Quint-PS-24DC/24DC/10 / Siemens SITOP
Power
/ Siemens Logo or equivalent power supply units with safe electrical
supply
isolation
A rating of up to 5 A/120 W can be supplied.
73022344 o
MAINTENANCE AND REPAIR
88