Quick Start Guide
7
Troubleshoot
7.1
Off-scale reading
Potential cause
Wiring is incorrect.
Recommended action
Verify and correct wiring.
Potential cause
Temperature element is open or shorted.
Recommended action
Check temperature element for open or short circuits.
See
Figure 7-1: Checking the Temperature Element
Potential cause
Sensor is not in process stream.
30
Figure
7-1.
A. Resistance temperature device (RTD)
B. Terminal strip in sensor junction box
C. Orange
D. Red
E. Gray
May 2020
Emerson.com/Rosemount