Fault finding guidelines

Advice!
All faults that may be corrected by the operator are indicated by the HMI. In the event of
a fault occurring that cannot be cleared by the operator, please refer to GEW or your local agent
for service.
Warning!
There are no user serviceable parts inside the e-Brick; do not remove covers under
any circumstances. Please note that if factory security seals are broken, the warranty is void.
With the exception of a communication problem, should any fault occur, the alarm bell flashes and
the fault is displayed on the left hand side of the screen. In the event of a communication failure, a
text message "Communication failure with lamp x, please deselect" will be displayed. The system
will still run with a communication failure, however the operator must deselect the faulty lamp on
the HMI. The 'fault' screen (Figure ), shows the faults that have occurred for each lamp.

6.5.1. General faults

Water cooling unit (chiller) failure
The cooling unit has failed while the printing machine is running.
reported
Emergency stop failure
The operator may have triggered an emergency stop of the printing machine
linking master-slave / slave/slave carry the e-Brick e-stop links)
Fan failure
The fan fails to operate when a lamp is started.
GEW UV e-Brick installation and operating manual
Figure 9: 'Fault History' screen
Check chiller MCB and reset if necessary
Ensure that the chiller is switched on
The chiller may have developed a fault; check the control panel to see if a fault is
Refer to the chiller manual for further fault finding suggestions
Check external emergency stop (e-stop) circuitry and switches
Check panel circuit breakers in external equipment
Check e-stop links between e-Brick units are intact (communications cables
Check the electrical supply to the fan
Check the MCB supplying power to the fan, reset if necessary
Check the thermal overload device on the fan. If the switch has been tripped,
reset it by pressing "1" button on the switch housing, then reset the fault on the
touch screen