Read Calibration Data; Calibration Interval Setting; Calibration Sequence Setting - Bosch BEA 150 Repair Instructions

Emissions analysis
Table of Contents
4.12.6

Read calibration data

4.12.7

Calibration interval setting

4.12.8

Calibration sequence setting

30
The following data are read out in this
menu:
Calibration date
Date of next calibration.
Calibration interval
Interval before recalibration in days.
Calibration sequence
Reaction after the calibration interval has
elapsed (Sec. 4.12.8).
i To access the Set calibration interval
menu, set the Service switch on the
test computer PCB to On.
In this menu, enter the number of days
after which the NO measuring sensor
must be recalibrated.
Enter the number of days using the key-
board.
Confirm your entry with E or by clicking
OK.
The OK window is displayed for 1 s.
This menu enables you to define how the
BEA should react once the calibration
interval has elapsed.
You can choose between the following
options:
Refuse access when recalibration
due:
Once the calibration interval has elapsed,
the BEA is disabled and displays a
message.
The BEA is only ready for measurement
once more after the NO measuring sensor
has been successfully recalibrated.
Notify when recalibration due:
Once the calibration interval has elapsed,
the BEA displays a message but remains
ready for measurement.
Ignore recalibration due
No message appears and the BEA
remains ready for measurement.