Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)

Electromagnetic Compatibility (EMC)

Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
Revision B
Changes or modification to this system not expressly approved by
GE Medical Systems could cause EMC issues with this or other
equipment. This system is designed and tested to comply with applicable
regulation regarding EMC and needs to be installed and put into service
according to the EMC information stated as follows.
WARNING
Use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause
unexpected or adverse operation.
WARNING
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked
use is necessary, the equipment or system should be
tested to verify normal operation in the configuration in
which it is being used.
The MAC 5000 is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure
that the MAC 5000 is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR11
RF emissions
Class A
CISPR11
Harmonic Emissions
Class A
EN 61000-3-2
Voltage fluctuations/
Flicker emissions
Complies
EN 61000-3-3
MAC™ 5000 resting ECG analysis system
2024917-010
Electromagnetic Environment - Guidance
The equipment uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The equipment is suitable for use in all
establishments including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
C-3