4.1 Startup Sequence
By default, the Giraffe OmniBed starts up in user mode. The next section describes how to start up the
OmniBed in service mode. The following table describes the Giraffe OmniBed startup sequence for each mode:
User Mode
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL screen
is yellow.
3. All LEDs turn off. The EL screen displays the
Giraffe logo.
4. The alarm light turns off. Software revisions
appear on the EL screen with OKs.
5. The user screen appears.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 4: Troubleshooting
Service Mode
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL screen
is yellow.
3. All LEDs turn off. The EL screen displays the
Giraffe logo.
4. The alarm light turns off. Software revisions
appear on the EL screen without OKs.
5. The service screen appears.