5.9.5 Chassis Replacement

Refer to "Figure 5-50 Bed Disassembly", "Figure 5-51 Heat Sink and Fan", "Figure 5-52 Fan Motor", "Figure 5-54
Chassis Bottom Cover", and "Figure 6-13 Upper Chassis".
1. Remove the bedsides, rotating bed, translation deck, tilt platform and pan.
2. Disconnect the compartment probe jack from the probe panel.
3. Remove the bottom chassis cover. (Refer to section 5.9.1.) Remove the two side chassis covers (3 mm hex
key).
4. Remove the front wall.
5. Disconnect the wiring harnesses.
6. Remove the two screws at the bottom corners that secure the probe panel to the enclosure. (Refer to
section 5.13.2.)
WARNING:
Be sure the front of the chassis is supported before removing the screws from the bracket to
prevent the chassis from falling.
7. While supporting the chassis, remove the 5 bolts on either side of the bracket casting that secure the
chassis to the bracket and remove the chassis.