Incubator Fan, Fan Motor, And Optical Fan Sensor; Figure 5-50 Bed Disassembly - GE Giraffe OmniBed Service Manual

Table of Contents
Chapter 5: Repair Procedures

5.9.2 Incubator Fan, Fan Motor, and Optical Fan Sensor

Refer to "Figure 5-50 Bed Disassembly", "Figure 5-51 Heat Sink and Fan", and "Figure 5-52 Fan Motor".
Mattress
Clear Plate
X-ray tray
Rotating bed
Translation deck
Tilt platform
Tilt release
Pan

Figure 5-50 Bed Disassembly

1. Remove the rotating bed, translation deck, tilt platform, and pan.
2. Remove the fan, fan hub, and fan seal from the top of the fan motor shaft.
3. Remove the chassis cover. (Refer to section 5.9.1)
4. Disconnect the motor leads.
5. Remove the 4 screws in the motor bracket and remove the fan motor assembly.
6. The motor isolators pop out of the holes in the bracket.
7. To replace the optical sensor, disconnect its connector, remove the screw from the boss in the chassis that
holds it in place, and remove the assembly.