Additional Conduit Runs; Outlet Legend For Ge Equipment - GE SIGNA PIONEER Installation Manual

Table of Contents
9'-4"
MAGNET ROOM
9'-4"
5'-1"
0'-3"
5
4
4
3
B
CONTROL ROOM
Typical
SIGNA PIONEER
2
2
1
A
6
5
EQUIPMENT ROOM
8
7
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG
ITEM QTY
1
2
3
4
5
6
7
8
9
ITEM QTY
6 9
C
Main Disconnect Panel
Main Disconnect Panel
System emergency off
System emergency off
Magnet Rundown Unit
Injector control unit
Integrated Battery Charging
LOCATED ELSEWHERE
Rev
DESCRIPTION
(CONTRACTOR SUPPLIED & INSTALLED)
Cable ladder 450mm x 150mm [18" x 6"]
Non-ferrous cable ladder 450mm x 150mm [18" x 6"]
Box above ceiling size per local code
150mm x 100mm [6" x 3 1/2"] Surface wall duct
100mm x 100mm x 50mm [4" x 4" x 2"] J-Box
Main disconnect panel
One 50mm [2"] cnd above ceiling
One 75mm [3"] cnd above ceiling
Breaker Boxes

Outlet Legend for GE Equipment

System emergency off (SEO), (recommended height 1.2m [48"] above floor)
Door interlock switch (needed only if required by state/local codes)
Emergency exhaust fan switch 1.2m [48"] height recommended)
Duplex hospital grade, dedicated wall outlet 120-v, single phase power
Network outlet
Dedicated telephone lines/network connection
Duplex hospital grade, dedicated outlet 120-v emergency, single phase power, 15a
Duplex hospital grade, dedicated outlet 120-v, single phase outlet routed through RF
filter

Additional Conduit Runs

(Contractor Supplied and Installed)
From
To
Facility power
Integrated Systems Cabinet
Integrated Cooling Cabinet
System emergency off
Integrated Cooling Cabinet
Integrated Systems Cabinet
Door Switch
Integrated Cooling Cabinet
Magnet
RF filter
RF filter
Room Light
RF filter
RF filter
Facility emergency power
Remote graphic display
Chiller
Facility power
TV Camera
Waveguide or RF filter
TV Monitor
Injector head
Waveguide or RF filter
Unit
A
Date
23/Jul/2018
E2 - Electrical Layout
Qty
Size (in)
Size (mm)
as Req'd
1
1
as Req'd
1
as Req'd
1/2
1
16
1/2
1
16
1
3/4
20
1
3/4
20
1
1
25
as Req'd
1
as Req'd
1
1
as Req'd
1
as Req'd
3/4
1
20
as Req'd
1
1
1
25
1
1
25
1
2 1/2
70
as Req'd
1
as Req'd
1