Radar level transmitter with hart protocol (24 pages)
Summary of Contents for Emerson Rosemount 5300 Series
Page 1
Reference Manual 00809-1600-4530, Rev AA March 2015 Rosemount 5300 Series Commissioning and Validation Procedures for Rosemount Radar...
Page 3
Avoid contact with the leads and terminals. High voltage that may be present on leads can cause electrical shock. Make sure the main power to the Rosemount 5300 Series Transmitter is off and the lines to any other external power source are disconnected or not powered while wiring the transmitter.
Page 4
Reference Manual Commissioning and Validation Procedure for Rosemount Radar 00809-1600-4530, Rev AA March 2015 Commissioning and Validation Procedure for Rosemount Radar...
Manual overview This manual provides commissioning, verification, and validation information for the Rosemount 5300 Series Radar Transmitters. This manual is intended to be used with the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Tools The following tools and documents are recommended to use during the installation and commissioning procedures: ®...
Page 8
Reference Manual Section 1: Introduction 00809-1600-4530, Rev AA March 2015 Introduction...
Reference Manual Section 2: Installation & configuration 00809-1600-4530, Rev AA March 2015 Section 2 Installation and Configuration Complete the Quick Start Guide ..........page 3 Complete the basic configuration .
Page 10
D. Lower Reference Point E. Upper Reference Point F. Upper Blind Zone G. Reduced accuracy H. Reduced accuracy I. Lower Blind Zone For more information, see Section 5: Configuration in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Installation and Configuration...
Select Archive Device to make a complete backup of the device, including several logs, and echo curves. View live values from the device. For more information, see Section 5.6 in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Installation and Configuration...
2 Probe End Projection 3 Dynamic Vapor Compensation 4 Signal Quality Metrics 5 Echo tracking 1 Analog output For more information, see Section 5.5 Basic configuration using a Field Communicator in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Installation and Configuration...
Basic configuration using AMS Suite (HART) or DeltaV The Rosemount 5300 Series can also be configured using AMS Suite or DeltaV. For more information, see Section 5.7 Basic Configuration Using AMS Suite (HART) in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530)....
Reference Manual Section 2: Installation & configuration 00809-1600-4530, Rev AA March 2015 2.3.2 Other optional advanced configurations Other functions that may be required can be found in Appendix C of the reference manual. Some of these may only be needed in certain conditions. This includes settings for: Hold Off Distance/Upper Null Zone which defines how close to the upper reference ...
Figure 2-5. Presentation of Analog Output Value in Rosemount Radar Master For more information, see Section Viewing Measurement Data In Rosemount Radar Master in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). 2.4.2 Verify analog output (HART devices only) Ensure the loop is set to manual mode in the Distributed Control System (DCS) for the applicable transmitter.
Reference Manual Section 2: Installation & configuration 00809-1600-4530, Rev AA March 2015 Also, or alternatively, activate the transmitter’s loop test function. Output 4, 12, and 20 mA and verify that the readings in the DCS match up. In Rosemount Radar Master, select Setup > Output > Analog Out 1 and click Loop test. Figure 2-7.
Adjust Upper Null Zone (UNZ). See Appendix C.3.2 Changing the Hold Off Distance/Upper Null Zone in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530) for details. Exclude any echo “before” the UNZ. Adjust Surface Threshold (ATC). See Appendix C.4 Threshold Settings in the Rosemount ...
Series Reference Manual (document number 00809-0100-4530). This helps to keep the probe centered in the chamber/still-pipe. If flexible probe in contact with nozzle, consider using a long stud. For more information, see Section 3.2 Mounting Considerations in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Installation and Configuration...
Page 19
Figure 2-10. The Diagnostics Window in Rosemount Radar Master To view Diagnostics in Rosemount Radar Master, select Tools > Diagnostics. For more information, see Section 7.14 Diagnostics in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530). Installation and Configuration...
Reference Manual Section 2: Installation & configuration 00809-1600-4530, Rev AA March 2015 Archive files Save echo curve and configuration backup file Permanently store the echo curve and configuration backup files for future use, with re-occurring verification procedures. In Rosemount Radar Master this can done using the Archive Device step under Guided Setup.
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA Section 3 Verification and Validation of GWR without Fluid Verification ..............15 Validation of device .
Page 22
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA March 2015 Reference pulse amplitude and location with regard to the zero line (see Table 3-1) Reference pulse amplitude varies with the type of probe used and the size of the chamber.
Page 23
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA General noise floor amplitude (see Figure 3-2) The noise floor should be relatively flat with low amplitude. This can vary with the size and shape of the chamber internal structure. For example, some inlet connections may have a rough weld edge and this can result in a small disturbance on the echo noise floor.
Page 24
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA March 2015 End of probe peak (see Figure 3-3) The end of probe peak is affected by the probe type and the use of centering discs. A metal centering disk will give a positive peak slightly offset from the probe end.
Page 25
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA Figure 3-4. Sample Verification Reflector Peak While the chamber is empty, save backup file and echo plot. Use the archive function in Rosemount Radar Master. Select Setup > ...
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA March 2015 Validation of device To validate the functionality of a device after it has been in service, it is possible to compare the current readings to baseline readings. To obtain the current readings, follow the same steps that were outlined under verification.
Page 27
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA Table 3-2. Validation Checks Initial Validation Description Allowable Tolerance reading reading Reference peak amplitude Normal variation is ±10% mostly depends on the ambient temperature. If the initial measurement was performed in one of the extreme temperature and the validation measurement is performed at...
Page 28
Reference Manual Section 3: Verification and Validation of GWR without Fluid 00809-1600-4530, Rev AA March 2015 Tag#______________________ Serial#___________________________ Within Initial Current Parameter Tolerance tolerance? Value value (Yes or No) Date Temperature of chamber LRV (4 mA setting) URV (20 mA setting) ±...
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Section 4 Verification Procedure with Fluid Verification procedure with active level measurement ......page 23 Common problems and recommended actions .
Page 30
D. Interface peak E. Interface threshold F. Surface threshold Distance Table 4-1. Typical Peak Amplitudes for Rosemount 5300 Series with Single Lead High Pressure Probe in 4-in. Chambers Approximate signal strength, ideal Peak conditions for single lead probe in 4 in.
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Common problems and recommended actions Surface pulse not detected Figure 4-3. Surface Threshold too High The Surface Threshold is above the Surface peak Distance, m A. Surface Threshold= Amplitude Threshold Curve (ATC) B.
Page 32
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Recommended actions Any of these options may be used to block false echoes: Raise general surface threshold to about 300 mV above false echo, but not more than ...
Verify that no errors or unattended warnings are present. Figure 4-6. The Diagnostics Window in RRM To view Diagnostics in RRM, select Tools > Diagnostics. For more information, see Section Diagnostics in the Rosemount 5300 Series Reference Manual (document number 00809-0100-4530) Monitor level while emptying During Step 1, the current level/interface reading was verified correct.
Page 34
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Figure 4-7. Log Registers Start emptying the tank or chamber making sure not to stop until it is completely empty. With a chamber, do not forget to close the process valves before draining. Figure 4-8.
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Echo curve verification with empty tank When the tank or chamber is empty, download and review the echo curve according to the following steps: Compare the echo curve with previous plots, taken during commissioning. There should be no major differences.
Reference Manual Section 4: Verification Procedure with Fluid 00809-1600-4530, Rev AA March 2015 Figure 4-9. The Configuration Report Window in RRM 4.7.1 Other validation suggestions To validate a device, do one of the following: Check the end of probe peak and label when the vessel or chamber is completely ...
Reference Manual Section 5: Additional Options 00809-1600-4530, Rev AA March 2015 In Rosemount Radar Master: Connect to the device. Right click on the device and select Properties. If “Diagnostics Suite” is mentioned in the Device Software Configuration 2 list, the device supports Signal Quality Metrics.
Reference Manual Section 5: Additional Options 00809-1600-4530, Rev AA March 2015 5.3.1 High level supervision Additionally, the reflector’s unique echo characteristics aid the transmitter to locate a liquid surface above the reflector, thereby offering increased reliability to detect high level conditions at a user selectable limit.
Page 40
Reference Manual Section 5: Additional Options 00809-1600-4530, Rev AA March 2015 Additional Options...
15.5 Vdc installations Note The Rosemount 5300 Series Transmitter is not safety-rated during maintenance work, configuration changes, multidrop, loop test, or other activity that affects the safety function. Alternative means should be used to ensure process safety during such activities.
Use a HART compliant master, such as Rosemount Radar Master (RRM) or a Field Communicator, to communicate with and verify configuration of the Rosemount 5300 Series. These instructions are applicable to the Rosemount 5300 Series safety-certified options with any differences noted.
Reference Manual Section 6: SIS Installations 00809-1600-4530, Rev AA March 2015 Site acceptance After installation and/or configuration, proper operation of the transmitter (including verification of all configuration changes) must be verified. A site acceptance test is therefore required. The proof test outlined in this document can be used for this. SIS operation and maintenance 6.3.1 Proof test...
Page 44
Suggested comprehensive proof test The suggested proof test described below detects approximately 89% in of possible Dangerous Undetected (DU) failures in the Rosemount 5300 Series Transmitters. Bypass the safety function and take appropriate action to avoid a false trip. Disable write protection in device (if enabled).
Page 45
Reference Manual Section 6: SIS Installations 00809-1600-4530, Rev AA March 2015 Using Loop Test, enter current value (mA) representing low alarm current. Verify that analog output current and terminal voltage are correct using reference meters. This step tests for possible quiescent current related failures. RRM: AMS Device Manager and Field Communicator: a.
Page 46
Reference Manual Section 6: SIS Installations 00809-1600-4530, Rev AA March 2015 Figure 6-2. Range Values A. Upper Blind Zone B. Reduced Accuracy C. Maximum Measuring Range D. Reduced Accuracy E. Lower Blind Zone F. 20mA G. Range 0 -100% H. 4mA Restore the loop to full operation.
Appendix A: Commissioning Checklist Reference Manual March 2015 00809-1600-4530, Rev AA Appendix A Commissioning Checklist Note Not all fields in this form are applicable to all transmitters. Plant information Completed Not completed Plant name: System No.: Country: Issuing agent/ Issued by: Company name: Date: ®...
Reference Manual Appendix A: Commissioning Checklist 00809-1600-4530, Rev AA March 2015 A.1.4 Electrical installation Completed Not completed Power supply within limits Voltage measured at the terminal at the transmitter: Groundings according to manual and local ...
Appendix A: Commissioning Checklist Reference Manual March 2015 00809-1600-4530, Rev AA A.1.5 Is the product’s Ex classification in accordance with requirements? YES NO For more information see Section: “Complete the basic configuration” on page 3 and Section “Echo curve verification at operating conditions” on page Primary Output: ...
Reference Manual Appendix A: Commissioning Checklist 00809-1600-4530, Rev AA March 2015 Echo tuning Completed Not completed For more information see Section: “Complete the advanced configuration” on page 7 “Review echo curve” on page 10l. Configuration tool used: ...
Appendix A: Commissioning Checklist Reference Manual March 2015 00809-1600-4530, Rev AA GWR radar verification in chambers UNIT: Note: Verification procedure is performed under ambient conditions. For dynamic vapor DATE: compensations units, ensure that the DC is calculated at 1.0 before starting verification process. If static TIME: vapor compensation is used, set vapor DC to 1 for the test and reset to desired value when finished.
Page 54
The water level will be visible in this line. In need of support with a transmitter, send a commissioning report together with a backup file from the transmitter to your local Emerson Process Management representative. Commissioning Checklist...
Page 55
Reference Manual March 2015 00809-1600-4530, Rev AA...
Page 56
Standard Terms and Conditions of Sale can be found at: Emerson Process Management www.rosemount.com\terms_of_sale. The Emerson logo is a trademark and service mark of Emerson Electric Co. Emerson FZE P.O. Box 17033, Rosemount and Rosemount logotype are registered trademarks of Rosemount Inc.