GE Druck DPI 610 User Manual page 89

Portable pressure calibrator series
Hide thumbs Also See for Druck DPI 610:
Table of Contents
Typical Uploaded Data log File (DPI 610)
The following gives details of a typical data file upload.
FILENAME: TEST 5
DATE:
05/11/1999
TIME:
15:58
TRIGGER:
KEYPRESS
AMBIENT TEMP : 24.1 C
NO. OF POINTS 11
RECORD TYPE
0
CURRENT mA PRESSURE INT bar
3.902
0.008
6.076
2.311
7.598
4.404
8.085
5.023
9.949
7.249
13.002
11.300
17.005
16.102
17.766
17.106
17.812
17.105
18.532
17.965
20.007
20.212
The Record Type number contained in the header is a 16-bit field, with the individual bits
representing result formats as detailed below.
Record Type 1
Record Type 2
Record Type 4
Record Type 8
Record Type 120
Record Type 896
Record Type 1024
Record Type 2048
Record Type 4096
Record Type 8192
Record Type 16384
-0.65 %span
1.42 %span
0.47 %span
0.41 %span
0.94 %span
-0.23 %span
0.77 %span
0.51 %span
0.80 %span
1.00 %span
-1.02 %span
P-Display
Switch Test
Leak Test
Input channel: Min/Max ON
Combination of all input process functions
Combination of all output process functions
mA step mode
mA value mode
mA ramp mode
no input selected
no output selected
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
05/11/1999
79
15:58:55
16:00:03
16:00:57
16:01:45
16:02:42
16:03:30
16:05:00
16:06:07
16:07:49
16:16:35
16:18:23
K415 Issue No. 1