Routine Zero - Siemens Milltronics BW100 Instruction Manual

Integrator for belt scales
Table of Contents

Routine Zero

Run the conveyor empty for several minutes to warm up the belt and insure that it is
empty. Test weights are not used during a zero calibration.
From the
run mode
Note:
E3 is an indication that the mechanical system is errant. The use of P377,
initial zero
, should be used judiciously and only after a thorough mechanical
investigation has been exercised.
The cause of the increased deviation must be found and rectified. A zero recalibration as
previously described can then be retried.
If the operator deems this deviation to be acceptable, set P377 to 1 to invoke an initial
zero calibration. Further deviation limits are now based on this new initial zero.