ABB Endura AZ30 series Programming Manual page 63

Integral probe and remote transmitter combustion oxygen monitor
Show thumbs Also See for Endura AZ30 series:
Table of Contents
Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
Parameter
...Input/Output
...Digital I/O 1 (2)
Type
One point
Two point
Sequence
Calibration
Accuracy Check
Cal. On Error Limit
Warn On Error Limit
Table 6.12 Input/Output Menu (Continued)
COI/AZ30E–EN Rev. D
Comment / [Range]
Selects the type of automatic calibrations started by
a digital input – see Section 7.1, page 69.
Selects the type of calibration / check performed
when activated by a digital input.
Performs a calibration that updates the calibration
offset and factor automatically if successful.
Performs an accuracy check (the normal calibration
sequence) but does not update the calibration offset
and factor in use by the transmitter.
Performs an accuracy check but, if the error is
greater than the value set in the Cal. Error Limit field
(see page 45), the calibration offset and factor are
updated automatically.
Performs an accuracy check, but if the error is
greater than the value set in the Cal. Error Limit field
(see page 45) a calibration warning is generated.
6 Programming
Default
Two point
Calibration
61