Appendix G. Communicating with the OPTICA Using Ethernet
G.2.8 4X40 Optica Configuration
Using an RS-232 communication link, the user can set the network settings. To see the set of supported commands,
type "$Help<>"
1. $SetIPAddress <>
(i.e., $SetIPAddress 3 112 60 36)
2. $SetIPSource <> (Assuming STATIC IP)
3. $SaveNetworkSettings<>
4. $GetNetworkSettings<> (Displays the current network settings)
It takes time to execute the "SaveNetworkSettings" command. Wait until the display updates to determine when to reset
the unit. In order for the settings to take effect, power to the unit must be cycled.
114