10.1 Closure commands for ramp segments
NEXT
SPWT (T)
SET-POINT WAIT
LOOP
HOLD
END
The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.
closure
command
ramp
°C
segment 1
spwt (t)
delayed
programme start
Follow-on with next programme segment
Wait until the setpoint temperature is
reached.
The oven only starts the next programme
segment when the programmed setpoint
temperature has been reached, even if the
programmed heating time has already
elapsed.
Ramp repeat function
The set programme is repeated after passing
through all programmed segments.
1-99 = repeats
cont = continuous repeat function
End of programme without switching off the
heating; temperature and all other settings
(e.g. air valve) are maintained.
End of programme, heating is switched off,
all other settings (e.g. air valve) are reset to
base status.
closure
command
ramp
closure
command
segment 3
ramp
spwt (t)
segment 2
next
segment 1
Ramp
closure
command
ramp
segment 4
next
segment 4
page 19
closure
command
ramp
segment 5
end
t=time
PERFECT