Maintenance; Fan Filter; Other Maintenance Items - Hitachi EUB-5500 Technical Manual

Ultrasound diagnostic scanner
Table of Contents

9.1 Fan Filter

Clean the dust-proof air filter at the air inlet of power unit located at the bottom front of
equipment.
If the filter is choked, the breaker may be cut off or the power unit may possibly
be damaged.
NOTE: • Be sure to turn OFF the power before cleaning the filter.
• Never turn ON the power unless mounting the filter. Dust around the power unit
may enter into it causing a fire.

9.2 Other Maintenance Items

1. Probe
(1) After using the equipment, use any tissue or soft cloth to completely wipe off jelly stuck
to the probe.
(2) When wiping off the probe, do not use hard paper or cloth.
(3) Put the probe in the probe holder or probe case, and keep it carefully.
(4) Care should be taken not to give strong shock to the probe or drop it.
2. Control Panel
(1) After using the equipment, turn OFF the power and completely wipe off jelly stuck to the
surface of control panel.
3. System Cabinet
(1)
Before cleaning the system cabinet, turn OFF the power and extract the power cable
from the power outlet.
(2)
Use cloth moistened with a neutral detergent to remove stains, and then wipe off it with
soft dry cloth.
NOTE: • Never use organic solvent such as thinner.
• Care should be taken for any liquid such as a detergent not to directly drop on the
equipment.
• When cleaning the equipment, care should be taken no to drop any liquid on the
inside of equipment, probe, probe connector or control panel.