Figure 8-8. Harmonics Equipment Setup
1. Frequency span: 500 kHz
2. Res BW: 10 kHz
3. Video BW: 30 kHz
1. Turn the RF output on; press
I - ] .
2. Turn the modulation
off
press
[MOD ONIOFF].
I
1.
Measure the level of the 2nd, 3rd, and 1/2 harmonics shown in the
test record.
2. Convert the harmonic level to decibels below the fundamental
(dBc) and compare the results to the corresponding limits.