Am Accuracy Performance Test - HP 8647A Operation And Service Manual

Signal generator
Table of Contents
AM Accuracy Performance Test
AM Accuracy
Performance Tkst
Connect the Test
Equipment
HP
8647A under test
0
0000
@
Measuring Receiver
Figure 8-3. AM Accuracy Equipment Setup
Configure the
Measuring Receiver
Configure the H P
8647A
Measure Depths
1. Reset
2. AM mode
3. Peak+ detector
4.
300 Hz high-pass filter
5. 15 kHz low-pass filter
1.
Turn AM on; press
2. Set the rate; press
( - 1 .
3. Turn the RF output on; press
I - ) .
[MOD ON/OFF].
1.
Enter the amplitudes, frequencies and depths shown in the test
2. Record the test results and compare the results to the limits in the
record.
test record.