Fields That Begin with the Letter I

Fields That Begin with the Letter I
Ideal Mobile Power
202
This field displays the power level that a mobile station on a call should be
transmitting to the Test Set. The value displayed for Ideal Mobile Power should
closely match the value for Average or Channel Power, providing the following
conditions are met:
The Average Power measurement is zeroed, if you are making an Average Power
measurement.
The Channel Power measurement is calibrated, if you are making a Channel Power
measurement.
The Power Control mode is Closed Loop.
Ideal Mobile Power is a calculated value based on an equation found in the EIA/
TIA IS-95 standard under "Estimated Open Loop Output Power". To calculate
Ideal Mobile Power, the Test Set sums the following values:
-RF Power
-73
+Nom Pwr
+Init Pwr
Example:
If Sector A Power is set at -73 dBM/BW, and Sector B Power and AWGN are
OFF, RF Power will be -73 dBm. If Nom Pwr and Init Pwr are both set to 0, the
Ideal Mobile Power will be 0 dBm/BW.
-(-73 dBm/BW)
-73
+0 db
+0 dB
=0 dBm/BW
Screen(s) Where Field is Present
CDMA CALL CONTROL
CDMA TRANSMITTER POWER RANGE TEST