Call Configure (Call Control)
Figure 12
The CALL CONFIGURE Screen
NOTE:
The number of fields displayed on the CALL CONFIGURE screen depends on the selected
system type.
NOTE:
In order to ensure that the Test Set makes the correct decisions regarding the presence of the
mobile stations's RF carrier, the Test Set's RF power meter should be zeroed before using the
Call Processing Subsystem. Failure to zero the power meter can result in erroneous RF power
measurements.
This screen is used to set some of the less commonly used Test Set configuration
parameters.
When the CALL CONFIGURE screen is displayed and the Call Processing
Subsystem is in the connect state, the Test Set constantly monitors the mobile
station's transmitted carrier power. If the power falls below 0.0005Watts, the
error message
RF Power Loss indicates loss of Voice Channel
be displayed and the Test Set will terminate the call and return to the active state.
Chapter 6, Description of Screens
Call Configure (Call Control)
will
113