Maintenance - GE Marquette MAC 1200 Operator's Manual

Hide thumbs Also See for Marquette MAC 1200:
Table of Contents

11.4 Maintenance

Checks before each use
Before each use, visually inspect the device, the
leads and electrodes for signs of mechanical damage.
If you detect damages or impaired functions that
may adversely affect the safety of the patient or user,
do not use the device before it has been repaired.
52
Cleaning, Disinfection and Maintenance
MAC® 1200
Technical Inspections
For safety, the devices require regular maintenance.
To ensure functional and operational safety of the
MAC 1200 units, Technical Inspections should be
carried out on an annual basis.
These checks should be performed by persons with
adequate training and experience.
The checks can be carried out by Marquette within
the framework of a service contract. The inspections
include the following checks:
• Visually inspect the device and the accessories
for signs of mechanical damage that may impair
the device functions.
• Check that the device labelling relevant for safety
is legible.
• Run a performance test as described in the
operator's manual.
• Measure the resistance of the protective earth
conductor and the equivalent leakage current per
your national regulations.
The device does not require any other maintenance.