10.2 Calibration and factor setting

All inputs are factory calibrated and recalibration should only be done by qualified personnel. If you
are not familiar with these procedures do not attempt to calibrate this instrument.
The calibration steps are:
a) Configure the input type to be calibrated.
b) Configure the lower and upper indication limits for the maximum span of the selected input type.
c) Connect to the input terminals a signal corresponding to a known indication value a little above the
lower display limit.
d) Access the parameter
applied signal. Then press the
e) Inject a signal that corresponds to a value a little lower than the upper limit of indication.
f) Access the parameter
applied signal. Then press the
Note: When checking the controller calibration with a Pt100 simulator, pay attention to the simulator minimum
excitation current requirement, which may not be compatible with the 0.170 mA excitation current provided by
the controller.

11. Copyright

The contents of this document are the property of the company LABOM Mess- und Regeltechnik
GmbH. Copying or reproduction is prohibited without prior approval. Contents of this document
refer to the device described therein.

12. Disclaimer

All technical content within this document can be modified without notice. The content of this
document is content of a recurring audit.
In losses due fire, earthquake, interventions by third or other accidents or intentional or accidental
misuse or incorrect using or using under unnormal conditions, repairs will be invoiced.
LABOM Mess- und Regeltechnik is not liable for inadvertently loss due using or not using this
product, such as loss of business income. LABOM Mess- und Regeltechnik is not liable for cases of
an unnormal usage.

13. Other regulations and standards

WEEE Informationen
Disposal of old electrical and electronic equipment (as in the European Union and other
European countries with separate collection systems).This symbol on the product or on its
packaging indicates that this product may not be treated as household waste. Instead, it
should be taken to an applicable collection point for the recycling of electrical and electronic
equipment. If the product is disposed of correctly, you will help prevent potential negative
environmental factors and human health, which could otherwise be caused by inappropriate
disposal. The recycling of materials will help to conserve natural resources. For more information
about recycling this product, please contact your local city office, your household waste disposal
service or the shop where you purchased the product.

14. Techncal Support

If you have any problems or questions please contact us at:
LABOM Mess- und Regeltechnik GmbH
Im Gewerbepark 13 • 27798 Hude • Germany
Phone: +49 4408 804-0 • Fax -100
www.labom.com
Operating instruction BA_074_2020-01_10.00 Universal panel indicator PH5100
. With the keys
   



key to store.
P
. With the keys
   



key to store.
P
and
adjust the display reading to match the
and
adjust the display reading to match the
- 15 -
BA_074e, Rev.1C5
Page 15/15
Table of Contents
loading

Table of Contents