Dionex ICS-5000, ICS-2100, ICS-1600, ICS-1100, and ICS-900 systems, as well as legacy IC systems. The Dionex AS-DV holds 50 vials (either 0.5 mL or 5.0 mL, or a combination of the two sizes). Vials can be sampled in any order and multiple samples can be taken from each vial.
® ® Adob Acrobat or Adobe Reader for assistance. Chapter 1, Introduces the Dionex AS-DV; explains the conventions Introduction used in this manual, including safety-related information. Chapter 2, Describes Dionex AS-DVoperating features and the Description software required for autosampler control.
Thermo Scientific does not recommend the use of flammable solvents in the Dionex AS-DV vials. Operation of a Dionex AS-DV in a manner not specified by Thermo Scientific may result in personal injury.
Page 12
Dionex AS-DV Operator’s Manual Messages d’avertissement en français Signale une situation de danger immédiat qui, si elle n'est pas évitée, entraînera des blessures graves à mortelles. Signale une situation de danger potentiel qui, si elle n'est pas évitée, pourrait entraîner des blessures graves à mortelles.
1 • Introduction to the Dionex AS-DV 1.3.2 Safety Symbols The symbols below appear on the Dionex AS-DV or on Dionex AS-DV labels. Alternating current Primary protective conductor terminal Secondary protective conductor terminal Power supply is on Power supply is off Indicates a potential hazard.
Dionex AS-DV Operator’s Manual 1.4.1 Declaration of Conformity The cETLus and CE marks on the Dionex AS-DV model/data label indicate that the autosampler is in compliance with the following standards. Doc. 065259-05 10/12...
2 • Description Operating Features 2.1.1 Front View Figure 2-1 illustrates the main operating features of the Dionex AS-DV. Vial Carousel Carousel Cover Carousel Release Button (Not visible in photo) Sampling Head Tower LED Display Front Cover Figure 2-1. Dionex AS-DV Front View...
Page 16
Figure 2-1). • The sampling head tower delivers sample from the sample vials to the sample out line, which exits the front of the Dionex AS-DV (see Section 2.1.2). For a description of how the sampling mechanism operates, see Section 2.2.1.
Page 17
PolyVials or a combination of the two sizes. Vial adapters (P/N 071901, package of 50) are required to load 0.5 mL vials into the carousel. Adapters are included in the Dionex AS-DV Ship Kit for 0.5 mL vials (P/N AAA-071115). PolyVials serve both rinse and sample functions.
Page 18
Vial Tray Vial trays (either for 0.5 mL or 5.0 mL vials) are included in the respective Dionex AS-DV Ship Kit (P/N AAA-071115, Ship Kit for 0.5 mL vials; P/N 068887, Ship Kit for 5.0 mL vials). Place vials into the tray and then fill them until the liquid in the vials reaches the top of the tray.
Figure 2-4. Vial Tray (0.5 mL Version Shown) 2.1.2 Front Interior Features Figure 2-5 illustrates the front interior of the Dionex AS-DV. To access the front interior, remove the front cover by gripping the indentations at each side and pulling the cover straight off toward you. Sample...
Page 20
The sample out line is routed from the sampling mechanism through an aperture in the front interior. From there, it is connected either to the high- pressure valve in the Dionex AS-DV (if installed) or to the injection valve in the IC system.
Power Receptacle Figure 2-6. Dionex AS-DV Rear Panel Drain Liquid from any spills exits the Dionex AS-DV enclosure through the port. Connect a drain hose (P/N 055075) to the port and place the DRAIN free end of the hose into either a waste container or an appropriate drain.
Page 22
Chromeleon is installed. Connect a USB cable between the USB connector on the Dionex AS-DV and a USB port on the PC. Alternatively, the Dionex AS-DV can be connected to a hub or an IC system with a USB port. For USB connection instructions, see Section B.6.
2 • Description Dionex AS-DV Operating Principles 2.2.1 Sampling Operating Principles The sampling head tower (see Figure 2-1) consists of a motor-driven needle, with a sampling tip on the end, that moves down to engage the top of the vial cap.
Section B.8.2). When performing a rinse, the Dionex AS-DV takes the rinse liquid from the first vial in the range. If the first vial is empty, the next vial in the range is used, and so on. To define rinse cycles in Chromeleon, see Section 3.8.3.
Page 25
2 • Description Figure 2-8 Figure 2-9 show the flow schematics for the 6-port and 10- port valves. B Position Figure 2-8. Optional 6-Port HIgh-Pressure Valve Flow Schematics B Position Figure 2-9. Optional 10-Port High-Pressure Valve Flow Schematics Operation as an Application Valve The optional valve can be configured for various application functions, including sample injection, concentrator column loading, or matrix elimination.
Page 26
Dionex AS-DV Operator’s Manual IC System AS-DV Injection Valve Optional Valve To Column (A Position) (Load Position) AS-DV Sample Out Line (5) S L (4) Sample Loop Concentrator (6) W C (3) Column (1) L P (2) PolyVial From Pump...
2 • Description Chromeleon Software The Dionex AS-DV is controlled with a PC configured with Chromeleon 7 Chromatography Data System or Chromeleon 6.8 Chromatography Data System. 2.3.1 Software Control Modes Two modes of software control are available: direct control and automated control.
Dionex AS-DV Operator’s Manual Figure 2-13. Chromeleon 6.8 Panel Tabset: Dionex AS-DV Control Panel • With automated control, you create a predefined list of commands and parameters (called an instrument method in Chromeleon 7 or a program in Chromeleon 6.8) for controlling the Dionex AS-DV.
Page 29
2 • Description to monitor the usage of replaceable parts. After replacing a part, click Reset to reset the Current counter to 0. Figure 2-14. Dionex AS-DV Wellness Panel • The Carousel Jam Counter increments each time the carousel jams.
3 • Operation and Maintenance This chapter describes routine operating and maintenance procedures for the Dionex AS-DV. Preparing the Sample Vials Thermo Fisher Scientific recommends thoroughly rinsing all PolyVials and caps with deionized water before filling to remove any traces of dust. Thermo Fisher Scientific DOES NOT recommend reusing vials or caps and does not guarantee their cleanness or proper operation if reused.
Dionex AS-DV Operator’s Manual 3.1.2 Cleaning the Caps Use the following procedure to clean caps without filters. For trace-level analyses, use caps without filters to avoid introducing contaminants from the filter. 1. Place caps into a clean 4-liter polyethylene container. Fill the container three-quarters full with deionized water and cap.
3 • Operation and Maintenance 3.1.4 Filling the Vials and Installing the Caps A cap must be installed in each vial in order for the autosampler to operate correctly. A vial without a cap will be sensed as an empty position.
Page 34
Cap Insertion Tool Figure 3-3. Caps Installed (0.5 mL Vials Shown) NOTE Do not use the concave end of the cap insertion tool for installing caps on Dionex AS-DV vials. It is for use with another autosampler. Doc. 065259-05 10/12...
3. To start the Chromeleon 7 client, click Start > All Programs > Chromeleon 7 > Chromeleon 7. 4. To display the Dionex AS-DV ePanel, click the Instruments category bar on the Console. Click your instrument in the Navigation Doc. 065259-05 10/12...
Page 37
3 • Operation and Maintenance Pane. Chromeleon 7 connects to the instrument and displays the ePanel Set. Click the AS-DV tab (see Figure 3-4). Figure 3-4. Example Chromeleon 7 ePanel Set: Dionex AS-DV ePanel Doc. 065259-05 10/12...
Dionex AS-DV Operator’s Manual 3.4.2 Connecting to Chromeleon 6.8 1. Turn on the PC. 2. Check the Chromeleon 6.8 Server Monitor icon on the Windows taskbar and start the server, if it is not already running: • If the icon is gray , the server is already running (idle).
Page 39
3 • Operation and Maintenance 5. To display the Dionex AS-DV Control panel, click the AS-DV tab (see Figure 3-5). Figure 3-5. Example Chromeleon 6.8 Panel Tabset: Dionex AS-DV Control Panel Doc. 065259-05 10/12...
A Chromeleon 7 instrument method or Chromeleon 6.8 program is a list of control commands to be executed at specified times, allowing automated operation of the Dionex AS-DV. This section provides an overview of how to create an instrument method or a program with a wizard. For detailed instructions, refer to the Chromeleon Help.
Page 41
AS-DV high-pressure valve configured as an injection valve, or the injection valve in the chromatography system module. The auxiliary valve is the optional Dionex AS-DV valve con- figured as an auxiliary valve. d. For Basic Edit Mode: Select the Volume of sample to deliver from the vial, select a Bleed option, and enable or disable a Rinse.
The range is 0 to 500 µL. The default value of 125 µL is appropriate for the 1.4 m (56 in) length of sample out line provided with the Dionex AS-DV. You can set the Delay Volume in the Chromeleon 7 Instrument Configuration Manager or the Chromeleon 6.8...
Instrument Method Wizard or Program Wizard, a default list of commands is displayed in the Command/Parameters table. The Dionex AS-DV executes the commands in the order in which they appear in the table. Default Basic Edit Mode Commands •...
Dionex AS-DV Operator’s Manual that will direct the bleed to waste. To disable the bleed, select None. For details about performing a bleed, see Section 3.8.2. • Rinse: Delivers rinse fluid through the sample out line. To enable the rinse, select the check box and then specify the volume. The volume is taken from the vials in the rinse vial range, specified in the Chromeleon 7 Instrument Method Manager or Chromeleon 6.8...
Page 45
3 • Operation and Maintenance The following two commands are available only if the optional valve is installed and configured as an auxiliary valve (see Section 2.2.3). • AuxValve.A: Sets the valve to the A position. • AuxValve.B: Sets the valve to the B position. Adding, Modifying, and Deleting Commands •...
1. Start the Sequence Wizard: • For Chromeleon 7: On the Console, on the menu bar, click Create. Select the instrument in which the Dionex AS-DV is configured and click Next >. • For Chromeleon 6.8: Select File > New. Select Sequence (using Wizard) and click OK.
Page 47
3 • Operation and Maintenance 3. On the last page of the wizard, click Finish (Chromeleon 7) or Done (Chromeleon 6.8). The sequence opens on the Chromeleon 7 Console (see Figure 3-6) or in the Chromeleon 6.8 Browser (see Figure 3-7).
During a run, do not press the Carousel Release button. Doing so clears all current vial information from Dionex AS-DV memory. (During a run, the Dionex AS-DV stores the vial size installed in each carousel position and the remaining volume in each vial.) After creating the sequence, you can start a run.
3 • Operation and Maintenance • To interrupt the autosampler when a sequence is not running, click Stop on the Dionex AS-DV ePanel in Chromeleon 7 or the Dionex AS-DV Control panel in Chromeleon 6.8. NOTE After the Dionex AS-DV stops its current operation, the carousel returns to the home position (position 1).
Page 50
Dionex AS-DV Operator’s Manual Figure 3-9). For suggested valve plumbing connections, see “Plumbing the Optional Valve” on page Figure 3-8. Sample Preparation Options (Chromeleon 6.8 Version Shown): Bleed Enabled Using the Injection Valve Doc. 065259-05 10/12...
Page 51
3 • Operation and Maintenance Figure 3-9. Sample Preparation Options (Chromeleon 6.8 Version Shown): Bleed Enabled Using the Auxiliary Valve Selecting Bleed adds two commands to the Command/Parameters table (see Figure 3-8 Figure 3-9): • The valve command switches the valve position to direct the bleed volume to waste.
Dionex AS-DV Operator’s Manual 3.8.3 Specifying a Rinse A rinse delivers rinse liquid from a vial to waste. To perform a rinse cycle, the Dionex AS-DV needs the following information: • The position of the vial containing the rinse liquid •...
Page 53
3 • Operation and Maintenance Server Configuration program (see “Specifying the Rinse Vial Position” on page 45). Figure 3-10. Sample Preparation Options (Chromeleon 6.8 Version Shown): Deliver Rinse in Basic Edit Mode Doc. 065259-05 10/12...
Page 54
Dionex AS-DV Operator’s Manual Selecting the Rinse check box and entering a Volume adds two commands to the beginning of the Command/ Parameters table (see Figure 3-10): • The InjectPosition command switches the valve position, directing the rinse volume to waste.
Page 55
Figure 3-10. When the Dionex AS-DV performs the rinse, it takes the rinse liquid from the first vial in the rinse vial range. If the first vial is empty, the next vial in the range is used, and so on. If the specified rinse volume is greater than a single vial volume, the Dionex AS-DV takes the Doc.
Page 56
Dionex AS-DV Operator’s Manual volume from more than one vial. Verify that the specified volume is available in the vials in the rinse vial range. Method 2: Enter a specific vial position in the Chromeleon instrument method or program. 1. On the Sample Preparation page of the Instrument Method Wizard/Editor or Program Wizard/Editor, select Advanced Edit Mode.
3-14, Sample 1 is taken from the vial in position 1. Samples 2 and 3 are taken from the vials in positions 2 and 3, respectively. For Sample 4, the Dionex AS-DV returns to the vial in position 1. Figure 3-14. Chromeleon Sequence Example (Chromeleon 6.8 Version Shown): Returning to a Vial...
A message informs you that a queue or batch is currently running. Click Yes to continue. 4. Wait until the Dionex AS-DV carousel and sampling tip are not moving. Then lift the top cover, remove the old vials, and replace them with the new vials.
3 • Operation and Maintenance 3.8.7 Controlling the Delivery Volume The Dionex AS-DV can deliver a full vial volume in one injection or deliver a partial vial volume in multiple injections. Sample delivery volume features vary, depending on the Edit Mode.
Page 60
• Full: The Dionex AS-DV delivers the full vial volume. • 1666 L: The Dionex AS-DV delivers 1666 L from the vial. Select this option if you want to take three injections from a 5000 L vial. • 2500 L: The Dionex AS-DV delivers 2500 L from the vial.
Page 61
3 • Operation and Maintenance For example, to take five injections from a filled 5.0 mL vial, select 1000 L. In the instrument method Script or program Commands view, a DeliverSample Volume=1000 command appears (see Figure 3-16). Figure 3-16. Chromeleon 6.8 Commands View Example: Deliver Multiple Injections Per Vial by Specifying the Volume Notes: •...
Page 62
Example: Using the From Sequence Volume Option When From Sequence is selected for the sample delivery volume, the Dionex AS-DV calculates the delivery volume using the formula shown in the Total Volume Delivered Estimator. In this case, the DeliverSample command is included without a specified volume.
Dionex AS-DV always injects a full loop volume. 3.8.8 Using Sample Overlap Mode The Dionex AS-DV can be configured to overlap sample preparation functions. This means that while data collection is occurring for one injection, the Dionex AS-DV performs the sample preparation steps for the next injection in the sequence.
Page 64
Dionex AS-DV Operator’s Manual Program Wizard, Chromeleon automatically adds the EndSamplePrep and BeginOverlap commands. See the example in Figure 3-18. Figure 3-18. Chromeleon 6.8 Commands View Example: Sample Overlap Enabled Doc. 065259-05 10/12...
Page 65
3 • Operation and Maintenance Enabling Sample Overlap Mode In the Instrument Configuration Manager or Server Configuration program, open the Dionex AS-DV Properties dialog box and click the Options tab. Select the Sample Overlap check box (see Figure 3-19). Figure 3-19. Dionex AS-DV Configuration Properties: Options Tab Page with Sample Overlap Enabled Doc.
Routine Maintenance This section describes routine maintenance procedures that the user can perform for the Dionex AS-DV. Any maintenance procedures for the Dionex AS-DV that are not described here must be performed by Thermo Fisher Scientific personnel. For assistance, contact Technical Support for Dionex products. In the U.S. and Canada, call 1-800-346-6390.
4 • Troubleshooting This chapter is a guide to troubleshooting problems that may occur while operating the Dionex AS-DV. • Section 4.1 describes the error messages displayed in the Chromeleon audit trail. • Section 4.2 through Section 4.8 describe other operating problems and how to resolve them.
Page 68
Dionex AS-DV Operator’s Manual The table below lists the Dionex AS-DV error messages. For troubleshooting assistance, refer to the page indicated in the table Alarms and Error Conditions Default Severity Level Attempt to deliver an invalid vial. Error page 59 Attempt to reset vials while needle is not home.
Page 69
This error occurs if you attempt to reset vial memory, either by pressing the button or clicking a Reset Vial Memory button on the Carousel Release Dionex AS-DV Control panel (see Figure 3-5), when the needle is down. To troubleshoot: 1.
Page 70
This error occurs when the needle is jammed or there is excessive backpressure. The motor speed is reduced and the Dionex AS-DV attempts to continue delivery. If the problem persists, the run is stopped, and the “Needle is jammed”...
Page 71
4 • Troubleshooting Needle is jammed. This error occurs when the needle is jammed or there is excessive backpressure. To troubleshoot: 1. Press the button to release the carousel and raise the Carousel Release needle. If the needle does not raise automatically, manually raise the needle (see Section 5.4).
Page 72
TTL triggered while another command is in progress. This error occurs if a controlling device sends a signal to the TTL input while the Dionex AS-DV is already performing a command. To troubleshoot: 1. After the other command is completed, send the signal to the TTL input again.
Page 73
NOTE Turning off the Dionex AS-DV power does not reset vial memory. c. On the Dionex AS-DV ePanel or Control panel, try to toggle the valve position by clicking first the Load Position (or Position A) button and then the Inject Position (or Position B) button.
Dionex AS-DV memory. (For example, a 5.0 mL vial is installed in position 10, but the Dionex AS-DV expects a 0.5 mL vial to be installed in that position.) Vial memory must be reset before starting a new run.
4 • Troubleshooting • Loose or damaged fitting connection(s) to optional high-pressure valve in the Dionex AS-DV or injection valve in the IC system 1. Make sure the liquid lines to the valve are tight. Replace any damaged fittings. See Section 5.1...
Dionex AS-DV Operator’s Manual • Liquid leak 1. Do not reuse caps or sample vials. 2. Inspect the lip and socket of each new cap for irregularities, nicks, scratches, or other damage that might interfere with sealing as the cap is pushed into the vial.
Excessive Backpressure • Injection valve in wrong position On the Dionex AS-DV ePanel (in Chromeleon 7) or Control panel on the panel tabset (in Chromeleon 6.8), toggle the valve position. • Restriction in the liquid flow path 1. If the high backpressure persists, follow the liquid flow path backward from the injection valve, through the high-pressure valve (if installed), loosening fittings and reconnecting them until you locate the restriction.
Sampling tip jams as it is being withdrawn 1. Check for excessive backpressure (see Section 4.7). 2. Check the sampling tip for possible obstruction: a. Turn off the power to the Dionex AS-DV. b. Manually raise the sampling head (see Section 5.4). c. Check for blockages.
Refer to the warranty statement in the Dionex Terms and Conditions for more information. Replacing Tubing and Fittings The Dionex AS-DV is plumbed with the needle and tubing assembly listed below: Tubing Assembly Used for Dionex AS-DV Sampling Tip Kit...
2. Turn off the Dionex AS-DVpower and disconnect the power cord. Raise the top cover. 3. Remove the Dionex AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you. 4. Use a #1 Phillips screwdriver to remove the three flathead screws from the...
Page 81
5 • Service 5. Use a 3-mm Allen wrench to remove the set screw on the sampling head (see Sampling Figure 5-2). Head Remove the set screw Figure 5-2. Removing the Set Screw 6. Pull slightly on the sample out line to gain a small amount of slack in the line.
Page 82
Dionex AS-DV Operator’s Manual 8. Lift the needle out of the carousel and set it to the right (see Figure 5-4). Lift the needle Set the needle to the right Figure 5-4. Lifting the Needle Out of the Carousel 9. Use a 2.5-mm Allen...
Page 83
5 • Service 10. Unscrew the thumbscrew on the carousel, and lift the carousel out of the autosampler. Figure 5-6. Removing the Carousel 11. Clear any blockages and clean the basin. 12. Align the holes on the bottom of the carousel Alignment Pins on Carousel with the pins on the top of the carousel shaft and reinstall the carousel.
Carousel Release position. 3. Turn off the Dionex AS-DV power and disconnect the power cord. Raise the top cover. 4. Remove the Dionex AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you.
Page 85
5 • Service 6. Use a 3-mm Allen wrench to remove the set screw on the sampling head (see Sampling Figure 5-8). Head Remove the set screw Figure 5-8. Removing the Set Screw 7. Pull slightly on the sample out line to gain a small amount of slack in the line.
Page 86
Dionex AS-DV Operator’s Manual 11. Thread the free end of the tubing up through the needle and screw the tip into place on the needle. 12. Continue threading the free end of the tubing up through the sampling head and out the aperture on the front of the Dionex AS-DV.
1. Turn off the Dionex AS-DVpower and disconnect the power cord. Raise the top cover. 2. Remove the Dionex AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you. 3. Use a #1 Phillips screwdriver to remove the three flathead screws from the...
Page 88
Dionex AS-DV Operator’s Manual 4. Insert a 2.5-mm Allen wrench into the screw on top of the sampling mechanism (see Figure 5-11) and turn the wrench clockwise. The needle begins moving up. 5. Continue turning the wrench until the top of the sampling head is even with the top of the sampling mechanism.
The following tool is needed to complete this procedure: 3-mm Allen wrench. 1. Turn off the Dionex AS-DVpower. The Dionex AS-DV power must be turned off before you rebuild a valve. If the power is on when you disconnect or connect the valve cable, the Dionex AS-DV will be damaged.
Page 90
Dionex AS-DV Operator’s Manual 5. Remove the 6-port or 10-port high-pressure valve: a. Use a 3-mm Allen wrench to remove the two screws holding the valve assembly in place (see Figure 5-13). Remove these 2 screws Figure 5-13. Removing the Valve Assembly b.
Dionex AS-DV may be damaged. 2. Raise the top cover. 3. Remove the Dionex AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you. 4. Disconnect each liquid line connected to the valve.
Page 92
Dionex AS-DV Operator’s Manual 5. Remove the 6-port or 10-port high-pressure valve: a. Use a 3-mm Allen wrench to remove the two screws holding the valve assembly in place (see Figure 5-15). Remove these 2 screws Black Locking Ring Figure 5-15. Removing the Valve Assembly b.
Page 93
11. Reinstall the valve assembly and reconnect all liquid lines. 12. Turn on the power to the Dionex AS-DV. 13. Start a test run to check for leaks from the valve. Tighten fittings as required. 14. Reinstall the front cover and close the top cover.
Dionex AS-DV Operator’s Manual Replacing the Main Power Fuses 1. Turn off the power switch on the Dionex AS-DV rear panel (see Figure 5-18). 2. Disconnect the main power cord from both its source and from the Dionex AS-DV rear panel.
A • Specifications A.1 Electrical Main Power 100 to 240 Vac, 50 to 60 Hz, 45 W Typical input power: 15 W Maximum line draw: 0.22 A at 110 Vac Fuses Two fast-blow IEC 127 fuses rated at 3.15 A (P/N 954745) A.2 Physical Dimensions Width: 44.45 cm (17.5 in)
For optimum operation, place the Dionex AS-DV close as possible to the left side of the IC system. • Allow at least 38 cm (15 in) of free space behind the Dionex AS-DV for connections and ventilation. •...
1. Raise the Dionex AS-DV top cover. 2. Remove the Dionex AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you. The sample out line is threaded through an aperture in the chassis (see Figure B-1).
Section B.3.2). B.3.1 Connecting the Sample Out Line Directly to an IC System 1. Route the sample out line through the side slot in the Dionex AS-DV chassis. Leave a little slack in the tubing. Route sample out line through...
Page 100
Installing the Optional Valve 1. Verify that the power to the Dionex AS-DV is off. The Dionex AS-DV power must be turned off before you install a valve. If the power is on when you connect or disconnect the valve cable, the Dionex AS-DV will be damaged.
Page 101
4. Remove the 6-port or 10-port valve from the Dionex AS-DV Valve Kit (6-port, P/N 068920; 10-port, P/N 068921). 5. Align the electrical cable with the connector on the back of...
Page 102
• Plumbing the 10-port valve as a bleed valve (see Figure B-7). For information about plumbing the valve for other functions, refer to the appropriate Thermo Scientific Dionex Application Note. To Column AS-DV Sample Out From Pump Sample Loop To Waste Figure B-5.
Page 103
Dionex AS-DV and is not elevated at any point. 2. Connect the Dionex AS-DV sample out line: a. Install a fitting and ferrule onto the end of the Dionex AS-DV sample out line. b. Connect the sample out line to the appropriate port on the valve:...
5. Replace the front cover and close the top cover on the Dionex AS- Go on to Section B.4. B.4 Connecting the Drain Line 1. Locate the corrugated drain hose (P/N 055075) in the Dionex AS-DV Ship Kit. 2. Push the hose onto the port on the DRAIN Dionex AS-DV rear panel.
USB ports. • Use a hub if the Dionex AS-DV is more than 5 meters (16 ft) from the PC. The USB standard limits the USB cable length to 5 meters (16 ft). Each USB device can be separated from the PC by no more than five hubs.
Dionex AS-DV Operator’s Manual B.7 Connecting the Power Cord 1. Verify that the power switch on the rear panel of the Dionex AS-DV is turned off. (The power switch may be turned on accidentally when the module is unpacked.) 2. Connect the power cord (IEC 320 C13) (ordered separately) from the main power receptacle on the rear panel to a grounded power source.
Chromeleon 6.8 Server is starting. 5. Press the power switch on the Dionex AS-DV rear panel to turn on the power. 6. Windows automatically detects the new USB devices. A message flashes on the screen to inform you that new hardware was found.
• For Chromeleon 6.8, click Start on the Windows taskbar and click All Programs > Chromeleon > Server Configuration. 2. Add the Dionex AS-DV to an existing instrument or timebase or create a new instrument or timebase. Doc. 065259-05 10/12...
Page 109
B • Installation 3. In the Dionex AS-DV Properties dialog box, click the Options tab (see Figure B-8). Figure B-8. Dionex AS-DV Properties Dialog Box: Options Tab Page 4. If a high-pressure valve is installed, select the function for which the valve will be used: •...
Console. Click your instrument in the Navigation Pane. Chromeleon 7 connects to the instrument and displays the ePanel Set. Click the AS-DV tab (see Figure B-9). Figure B-9. Example Chromeleon 7 ePanel Set: Dionex AS-DV ePanel Doc. 065259-05 10/12...
Page 111
View > Default Panel Tabset or click the Default Panel Tabset toolbar button. 3. To display the Dionex AS- DV Control panel, click the AS-DV tab (see Figure B-10). Figure B-10. Example Chromeleon 6.8 Panel Tabset: Dionex AS-DV Control Panel Doc. 065259-05 10/12...
Verifying Chromeleon Communication To verify that Chromeleon is communicating with the Dionex AS-DV: 1. On the Dionex AS-DV ePanel (in Chromeleon 7) or Control panel (in Chromeleon 6.8), under Settings, enter a Vial Position. 2. Under Commands, click Move to Vial. The carousel should move to the entered position.
C-1) on the rear panel of the Dionex AS-DV provides one relay output and one TTL input. Figure C-1. Relay Output and TTL Input Connector (on Dionex AS-DV Rear Panel) The relay output can be programmed in Chromeleon to switch any low-voltage device.
When the input is triggered for the first time, the Dionex AS-DV performs a full vial delivery on vial position 1. When triggered again, the Dionex AS-DV performs a full vial delivery on vial position 2, and so on.
If the Dionex AS-DV has been connected to Chromeleon, the delivery speed, flush factor, and TTL homing delay last entered are used. To change one of the options, you can connect the Dionex AS-DV to Chromeleon, change the option’s value, and then disconnect the Dionex AS-DV from Chromeleon.
Page 116
• For Chromeleon 6.8, click Start on the Windows taskbar and click All Programs > Chromeleon > Server Configuration. 2. Under the instrument or timebase, double-click the Dionex AS-DV icon. The Dionex AS-DV Properties dialog box appears. 3. Click the TTL Inputs tab and then press the F2 key.
Page 117
C • Relay Control The Device Configuration dialog box appears (see Figure C-2). Figure C-2. Device Configuration Dialog Box: Selecting the TTL Input Control Mode 4. Select the TTL input control mode required by the controlling device from the Mode list. Doc.
When attaching wires to the connector plug, be careful not to allow stray strands of wire to short to an adjoining position on the connector. 3. Plug the connector into the 4-pin connector on the Dionex AS-DV rear panel (see Figure C-1).
TTL-2 TTL-3 TTL-4 Figure C-4. Example TTL Input Connection: Dionex AS-DV Connected to Dionex IC20 To control the Dionex AS-DV, set up a Dionex IC20 method that includes timed events for triggering on the autosampler. Figure C-5 illustrates the timed TTL IN events in an example method.
Figure C-5. Dionex IC20 Timed Events Example C.4 Controlling the Relay Output The Dionex AS-DV provides relay contacts for control of functions in external devices. The relay output can be programmed in Chromeleon to switch any low- voltage device. Switched current must be no more than 200 mA at 5 VDC. The relay-contact closures are normally open.
Page 121
C • Relay Control Controlling the Relay Output Directly Using Chromeleon 7 Open the ePanel Set (see Section 3.4.1) and press the F8 key. The Command window appears. 2. In the list of devices in the left pane, click Sampler_Relay. 3.
Page 122
Dionex AS-DV Operator’s Manual 3. Execute the required command (see the example in Figure C-6). Figure C-7. Chromeleon 6.8 Commands Dialog Box: Example Relay Control Command Doc. 065259-05 10/12...
Page 123
C • Relay Control Controlling the Relay Output with an Instrument Method or Program To add relay output control commands to a Chromeleon 7 instrument method, use the Script Editor (see Figure C-1). Figure C-1. Chromeleon 7 Script Editor Doc. 065259-05 10/12...
Page 124
Dionex AS-DV Operator’s Manual In Chromeleon 6.8, you can enter the commands on the Relay and State Devices Options page in the Program Editor or Program Wizard (see Figure C-8). Figure C-8. Chromeleon 6.8 Program Wizard: Example Relay Output Control Command...