Thermo Scientific  Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List
Thermo Scientific  Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Quick Links

_________________________________________________________________________________
Part of Thermo Fisher Scientific
Therm Scientific Barnstead
Nanopure
Life Science (UV/UF)
Ultrapure Water System - Series 2114
Operating Manual and Parts List LT2114X1 Rev. 1
Table of Contents
loading

Summary of Contents for Thermo Scientific Thermo Scientific Barnstead Nanopure Life Science 2114 Series

  • Page 1 Therm Scientific Barnstead Nanopure ™ Life Science (UV/UF) Ultrapure Water System - Series 2114 Operating Manual and Parts List LT2114X1 Rev. 1 _________________________________________________________________________________ Part of Thermo Fisher Scientific...
  • Page 2 Models covered in this manual Models Voltage D11931 (7146) 90-240V MANUAL NUMBER LT2114X1 (7007146) 26925 3/10/11 Decreased water pressure rating from 100 to 80 psi 7/20/10 Transfer to Marietta (was LT2114X1 3/4/10) ECR/ECN DATE DESCRIPTION Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 3 Thermo Fisher Scientific makes no representations or warranties with respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to the use of this manual. ©2010 Thermo Fisher Scientific. All rights reserved. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 4 Always use the proper protective equipment (clothing, gloves, goggles, etc.) 4 Always dissipate extreme cold or heat and wear protective clothing. 4 Always follow good hygiene practices. 4 Each individual is responsible for his or her own safety. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 5 We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
  • Page 6: Table Of Contents

    Resetting the Cleaning Timer .......9-5 Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 7 Wiring Diagram ........14-5 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 8: Safety Information

    Safety Information Section 1 Your Thermo Scientific Barnstead Nanopure Life Science (UV/UF) ultrapure water system has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. This manual contains important safety information.
  • Page 9 8. This unit is equipped with an ultraviolet lamp. Ultraviolet radiation is harmful to the eyes and skin. Do not attempt to observe the lamp directly. 9. Refer servicing to qualified personnel. 10. A full cartridge pack may weigh about 20 lbs. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 10: Introduction

    Introduction Section 2 Congratulations on your purchase of a Thermo Scientific Nanopure Life Science (UV/UF) ultrapure water system. This water purification system is designed to provide nuclease-free (RNase and DNase), pyrogen-free, low TOC, high resistivity, reagent grade water that exceeds ASTM Type I, ISO 3696 and CLSI-CLRW Type I standards.
  • Page 12: General Specifications

    Accuracy: ......± 5% Repeatability: ......± 3% Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 13 Measurement, Control and Laboratory Use; Part I: General Requirements per the provisions of the Electromagnetic Compatibility Directive 2004/108/EC, and per the provisions of the Low Voltage Directive 2006/95/EC. Copies of the Declaration of Conformity are available upon request. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 14: Unpacking

    Plug Adapter ......CEX42 Power Cords Also required for installation: * Customer Supplied Fasteners to mount unit and remote display. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 16: Installation

    Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located. Do not use in the presence of flammable materials; fire or explosion may result. This device contains components which may ignite such materials. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 17: Ultrafilter Installation

    Top connector making sure the connection is tight. Flow 7. Secure the ultrafilter Velcro strap with the velcro strap. Metal bracket Bottom connector Figure 5-2. Ultrafilter Installation Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 18: Uv Lamp Installation

    6. Make sure the UV lamp cable is tucked behind the chamber before closing and relatching the door. Figure 5-4. UV Lamp Installation - Left Side View Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 19: Wall Mounting

    A minimum of four (customer-supplied) fasteners will be required - two on the top and two on the bottom. 3. Attach the wall bracket to the wall using the customer-supplied fasteners. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 20: Install Control Panel In Remote Location

    AY1367X2. Wall Mount Note The blank Bracket protective cover on the rear side panel display connector should remain in place when the display is not remotely located. Figure 5-5. Wall Mounting Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 21 7. Tighten the cable screws on each end to secure the cable to the unit and control panel. 8. The control panel may now be mounted on the wall. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 22: Water Connections

    Note The cartridge pack and ultrafilter must be installed before connecting the water line. 3. If a pressure switch is to be used, see the “Installing a Float or Pressure Switch” section for more information. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 23: Atmospheric Drain Connection

    4. To remove tubing, push collet toward fitting body and pull on tubing. To reuse Figure 5-9. Tubing Removal fitting, begin assembly at Step 1. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 24: Electrical Connections

    Before connecting the power cord to the power module, install the Power switch proper fuses in the fuse holder. Refer to Figure 10. Power cord Figure 5-10. Electrical Connections Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 26: Controls

    LED is not illuminated: System is OFF or in (Idle) mode. • Solid Green LED: System is recirculating and operating normally. • Blinking Green LED: System is in Standby mode. • Red LED: System is alerting user to an operational error or maintenance issue. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 27: Switches

    Push to the left to deliver water manually until button is released. Note The unit MUST be in the normal recirculating mode (when the display is showing purity) in order to dispense water. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 28: Initial Operation

    Low Organics, Type 1 Water, Reverse Osmosis or Distilled Water Feed DIamond Kit Type 1 Deionized Feed D50283 Low Organics, Type 1 Water, DI Water Feed Each cartridge pack includes one 0.2 micron absolute final filter. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 29: Initial Rinse

    2. Attach an atmospheric drain line (1/4” [.64 cm]) tubing by pushing into the quick connect fitting on the lower left back of unit. See layout connection under the “Water Connections” section. 3. Check to ensure the dispense knob is in the OFF (middle) position. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 30 10. Unit will display, “Air Purge?” Press ENTER. 11. Upon completion of the air purge, complete the “System Cleaning Procedure.” Caution Never dispense water without first installing hose barb provided in parts bag or final filter supplied with cartridge pack. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 31: System Cleaning Procedure

    4. Disconnect feed water supply at the quick-disconnect inlet valve on left side of unit. 5. Press the UP or DOWN ARROW to select (YES) and then press ENTER. Display will now read, “Injected Cleaner (NO).” Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 32 13. Press the UP or DOWN arrow to select (YES) and press ENTER. Display will now read, “Unit Cleaning.” 14. Unit may now be left unattended until display reads, “Cleaning Complete: Press ENTER.” 15. After cleaning is complete, the unit will return to the (Idle) mode. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 33 15-20 liters through the cartridges and 0.2 micron filter may be necessary. Caution Never dispense water without first installing hose barb provided in parts bag or final filter supplied with cartridge pack. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 34: Normal Operation

    6. Allow the water’s resistivity to rise to the desired purity before drawing off water. 7. The system should be left On or in Standby during the work day. See “Use of Standby Mode.” Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 35: Automatic Dispensing

    Note Press any switch except ENTER to stop dispensing. Display will read, “Auto Dispense Interrupted by User.” Note If the Nanopure “Auto Dispense” feature is set to (TIME), a connected Accudispense accessory (if one exists) will not automatically dispense. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 36: User Settings

    BACK switch until the display reads Language “LANGUAGE.” 2. The currently selected language is displayed on line 2 of the display. 3. Press the UP or DOWN arrow to select the language desired. 4. Press ENTER. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 37: Setting The Date And Time

    10. Press the UP or DOWN arrow until you reach the desired hour value. Press ENTER. 11. Press the UP or DOWN arrow until you reach the desired minute value. Press ENTER. 12. Press BACK twice to return to the (Idle) mode. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 38: Setting The Cell Constant

    14. If the cell constants displayed on line 1 of the display are correct, press the UP or DOWN arrow to select “Yes.” 15. Press ENTER. 16. Press BACK twice to return to the (Idle) mode. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 39: Selecting The Set Point

    UP or DOWN arrow until display reads, “Advanced Menu?” 2. Press ENTER. 3. Proceed through the options by pressing ENTER until “Adjust Set Point” is displayed. 4. Press the UP or DOWN arrow to choose “Yes.” Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 40: Resetting The Uv Timer

    With the Nanopure in an (Idle) position, the UV timer can be reset by performing the following steps. Timer Note When the user begins receiving “Change UV Lamp” reminders, it will be time to change UV lamp and reset UV timer as described below. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 41: Setting Volumetric Dispensing

    UP arrow: +0.25 up to 1 L, then 5 L, then incremental by 5 L. DOWN arrow: Decrement by 1 L until 1 L, then by 0.25 L. Maximum: 60 L Minimum: 0.25 L Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 42: Performing A System Flush

    Note An automatic 1-minute flush will occur once every 12 hours. If a unit is (Idle) or in STANDBY for over 12 hours, an automatic system flush will occur immediately after the START/STOP switch is pressed. Note Press the STOP/STOP switch to cancel the flush. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 43: Auto Standby

    1. From the “Nanopure (Idle) xx:xx:xx xx/xx/xx”, press the UP or DOWN arrow until display reads, “Advanced Menu.” 2. Press ENTER. 3. Press ENTER to proceed through the options until “AUTO STANDBY START” is displayed. Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 44: Unit Under Counter

    4. Press the UP or DOWN arrow to choose “YES.” Note When “UNIT UNDER COUNTER” is set to “YES,” Auto Dispense will not be available from the main unit dispenser. 5. Press ENTER. 6. Press BACK until you return to the (Idle) display. Thermo Scientific Nanopure Life Science (UV/UF)
  • Page 46: Accessories

    Nanopure into the remaining opening of the PVC tee. 4. Route the cable from the float or low pressure switch to the top left of the Nanopure. Thermo Scientific Nanopure Life Science (UV/UF) 10-1...
  • Page 47: Dispense Overflow Cutoff Float

    Note If an auto dispense is stopped by the AY1367X1, the Nanopure will change modes to normal recirculation. Figure 10-2. AY1367X1 Dispense Overflow Cutoff Float 10-2 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 48: Optional N.i.s.t. Calibration Module

    7. From the (Idle) mode, Press the UP or DOWN arrow until the display reads, “N.I.S.T. Calibrate?” Press ENTER. 8. Display will show “Calibration Module Installed?” Press the UP or Connection DOWN arrow to select (YES). Figure 10-3. PC Board Thermo Scientific Nanopure Life Science (UV/UF) 10-3...
  • Page 49: Remote Dispensers

    14. Reattach the cover plate and close and latch the door. 15. Reconnect the power cord. The calibration procedure is complete. Remote Dispensers Refer to the Operating Instructions for your Manual or Accudispense volumetric remote dispenser (LT1366X11). 10-4 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 50: Computer/Printer Setup

    12. When you are finished storing data from the Nanopure, save your file. You can exit Hyperterminal or set up a new experiment. 13. Hyperterminal data can be manipulated and graphed in Excel®. Thermo Scientific Nanopure Life Science (UV/UF) 11-1...
  • Page 51: Procomm

    The optional printer (part no. AY1137X1) provides a paper tape record for future reference. During normal Recirculation, the printer prints date, time, purity and temperature readings once every minute. 11-2 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 52: Printer Setup

    Located on the front of the printer (AY1137X1), the power switch turns power to the printer ON and OFF. The POWER light will illuminate when the printer is ON. For additional setup and operation, refer to the instruction manual included with the printer. Thermo Scientific Nanopure Life Science (UV/UF) 11-3...
  • Page 54: Maintenance And Servicing

    8. Reinstall the cell into the cell well and hand-tighten. 9. Reroute the cable and reconnect to J2 on circuit board. 10. Replace PC board cover and close door. 11. Reconnect water and turn unit ON. Thermo Scientific Nanopure Life Science (UV/UF) 12-1...
  • Page 55: System Cleaning

    4. Disconnect feed water supply at the quick-disconnect inlet valve. 5. Press the UP or DOWN ARROW to select (YES) and then press ENTER. Display will now prompt, “Injected Cleaner (NO)?” 6. Remove solution syringe from wrapper. 12-2 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 56: System Depressurization

    2. From “Air Purge?” press the UP arrow. 3. From “Depressurize” press the UP arrow to select (YES). 4. Press ENTER. Display will show “System Flushing” and the time remaining as the system depressurizes. Thermo Scientific Nanopure Life Science (UV/UF) 12-3...
  • Page 57: Cartridge Pack Replacement

    7. Install new cartridge pack, rinse and clean according to the instructions in the “Initial Operation” section. Figure 12-1. Cartridge Pack Replacement 12-4 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 58: Micron Filter Replacement

    3. Remove the protective cap from the filter bell. 4. Rinse 1-2 liters of water through the filter to drain prior to using the product water. 0.2 Micron Filter Figure 12-2. Replacing the 0.2 Micron Filter Thermo Scientific Nanopure Life Science (UV/UF) 12-5...
  • Page 59: Ultrafilter Replacement

    96 hours without use, it will become contaminated and may affect product water quality. 7. Pull the ultrafilter out of the unit and turn it to the left side. (This will make it easier to disconnect the bottom fitting.) 12-6 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 60 Figure 12-3. Tool Used for Ultrafilter Removal 10. Clean the Nanopure according to the “System Cleaning” section of this manual. Top connector Flow Velcro strap Metal bracket Bottom connector Figure 12-5. Ultrafilter Installation Thermo Scientific Nanopure Life Science (UV/UF) 12-7...
  • Page 61: Uv Lamp Replacement

    Even a light coating of perspiration will reduce the effectiveness of the lamp. If the glass portion of the lamp is touched, clean it with a damp, lint-free cloth: use isopropyl alcohol as required. 12-8 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 62: Fuse Replacement

    2. Pull out the fuse drawer located in the power entry module. 3. Remove old fuses and replace with fuses of the same type and rating. See “Replacement Parts.” 4. Replace fuse drawer. 5. Reattach the power cord. 6. Operate normally. Thermo Scientific Nanopure Life Science (UV/UF) 12-9...
  • Page 63: Shutdown

    If the system has remained inactive and full of water for more than 96 hours, the unit should be drained, cleaned and new cartridge pack installed prior to use. 12-10 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 64: Troubleshooting

    Check the expiration date. Cartridge packs begin to lose capacity after being stored for 2 years from the Short cartridge pack life. Cartridge pack is beyond expiration date. date of manufacture. Replace the cartridge pack with an unexpired one. Thermo Scientific Nanopure Life Science (UV/UF) 13-1...
  • Page 65 Turn the Nanopure OFF (I/O Switch) and Display reads, “Meter Board Problem.” PC board communication error. restart the unit. Replace the resistivity meter board mounted System electronics failure. on the inside of the right door. 13-2 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 66 Debris on o-ring or bottom of unit manifold. Remove any objects with a lint-free cloth. Water is leaking at atmospheric drain connec- Insert tubing into connection until it bottoms Tubing not fully inserted. tion out. Thermo Scientific Nanopure Life Science (UV/UF) 13-3...
  • Page 68 Main Fuse, 230 Volt - Type T Time Lag, 0.63 Amp, 250V FZX54 PTFE Tape, Roll 06078 UV Lamp LMX13 Transformer Fuse, Type F Fast-Acting, 3.15 Amp, 250 V AZ9027 Cleaning Solution CMX25 Ultrafilter FL1192X1 Thermo Scientific Nanopure Life Science (UV/UF) 14-1...
  • Page 69: Replacement Parts

    Resistivity Meter PC Board PC896X2 Check Valve PMX214 Wire Harness, UV Bulb WH1191X1 Flow Sensor ME1190X1 Flush Solenoid RY1190X3 Cartridge Pack Manifold BK1190X3 Dispenser Manifold W/Solenoid BK1367X1 Dispense Switch SW1190X1 Cell O-ring GSX29 14-2 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 70: Ordering Procedures

    All parts listed herein may be ordered from the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory.
  • Page 71: Flow Chart - Operational Overview

    Remote Gun System flush? Date Format Auto Dispense Set Date Menu? Set Time Arrow Legend: Set Cell Constant Keypress Options Set Display Automatic Entry Brightness Menu Wrap-around Adjust Set Point Diagnostics Operational Overview menu? 14-4 Nanopure Life Science (UV/UF) Thermo Scientific...
  • Page 72: Wiring Diagram

    Section 14 Replacement Parts Wiring Diagram Thermo Scientific Nanopure Life Science (UV/UF) 14-5...
  • Page 74 Thermo Fisher Scientific 401 Millcreek Road Marietta, Ohio 45750 United States www.thermofisher.com...