Service Platform (cont'd)
7.) Verify that the following screen (Service Platform) is displayed on the monitor.
8.) Verify that Service Platform Version is 2.2.0Q.
9.) Click on x located at the upper right corner of the service platform screen to close the Service
Platform and return to the scan panel.
When the service platform is NOT displayed, check if CAPS lock is selected. The CAPS should not be
NOTICE
selected.
8-2-11-2
Functional Checks for Probe Recognition
1.) Touch Scan button on the Touch Panel to return to the scan screen.
2.) Connect each probe to ensure that they are recognized.
3.) Check every probe in the following modes and ensure that no artifacts or no problems are found in:
-
-
-
-
-
4.) Install the removed parts in the reverse order of removal.
8-2-11-3
Calibration and adjustment
No calibration or adjustments are needed after this part replacement.
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
14
Figure 8-125 Service Platform
B-mode
Color FLow
Pulsed Doppler
M-mode
CW (option)
LOGIQ™ 7 S
M
ERVICE
ANUAL
8-59