Verify Operation - Emerson Rosemount 648 Reference Manual

Wireless temperature transmitter with rosemount x-well technology
Hide thumbs Also See for Rosemount 648:
Table of Contents
Reference Manual
00809-0200-4648
4.3

Verify operation

The transmitter can be commissioned before or after installation. It may be useful to
commission it on the bench, before installation, to ensure proper operation and to
become familiar with its functionality. When applicable, make sure the instruments are
installed in accordance with intrinsically safe or non-incendive field wiring practices. The
device will be powered whenever the power module is installed. To avoid depleting the
power module, make sure it is removed when the device is not in use.
Operation can be verified in four locations: at the device via the LCD display, using a Field
Communicator, the Wireless Gateway's integrated web interface, or using AMS Suite
Wireless Communicator or AMS Device Manager.
4.3.1
LCD display
During normal operation, the LCD display will display the PV value at the wireless transmit
rate up to as fast as one minute intervals. Refer to
codes and other LCD display messages. Press the Diagnostic button to display the TAG,
Device ID, Network ID, Network Join Status and Device Status screens. For Device Status
screens, see .
Figure 4-1: Diagnostic Screen Sequence
Figure 4-2: Network Status Screens
Rosemount 648 Wireless
Tag
Device ID
A b c d e
i d - 1 2
f g h
3 4 5 6 7 8
Searching for
Joining network
network
N E T w K
S R C H N G
LCD display screen messages
Network ID
n e t w k
13 0 5
I D
Connected with
1 parent
n e t w k
n e t w k
N E G O T
L I M - O P
Commissioning
for error
Network status
Device variables
n e t w k
s n s r
1 0. 0 2
O K
d e g c
Connected with
2 parents
n e t w k
O K
May 2020
45

Hide quick links:

Table of Contents
loading

Table of Contents